EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Accuracy and time efficiency of a novel deep learning algorithm for Intracranial Hemorrhage detection in CT Scans

Tommaso D'Angelo, Giuseppe M Bucolo, Tarek Kamareddine, Ibrahim Yel, Vitali Koch, Leon D Gruenewald, Simon Martin, Leona S Alizadeh, Silvio Mazziotti, Alfredo Blandino, Thomas J Vogl, Christian Booz

La radiologia medica · 09.08.2024 · 129 (10) · 1499-1506

DOI: 10.1007/s11547-024-01867-y ↗

Kurz gefasst

Fachartikel in der Zeitschrift La radiologia medica, online erschienen am 09.08.2024. Verfasst von Tommaso D'Angelo und 11 weiteren Autor:innen. Beteiligte Einrichtung: Diagnostic and Interventional Radiology Unit, BIOMORF Department, University of Messina, Messina, Italy. tommasodang@gmail.com. Hauptthemen laut MEDLINE: Deep Learning, Tomography, X-Ray Computed, Intracranial Hemorrhages. Die Arbeit wurde bisher 25-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To evaluate a deep learning-based pipeline using a Dense-UNet architecture for the assessment of acute intracranial hemorrhage (ICH) on non-contrast computed tomography (NCCT) head scans after traumatic brain injury (TBI). Materials and methods This retrospective study was conducted using a prototype algorithm that evaluated 502 NCCT head scans with ICH in context of TBI. Four board-certified radiologists evaluated in consensus the CT scans to establish the standard of reference for hemorrhage presence and type of ICH. Consequently, all CT scans were independently analyzed by the algorithm and a board-certified radiologist to assess the presence and type of ICH. Additionally, the time to diagnosis was measured for both methods.

Results

A total of 405/502 patients presented ICH classified in the following types: intraparenchymal ( n = 172); intraventricular ( n = 26); subarachnoid ( n = 163); subdural ( n = 178); and epidural ( n = 15). The algorithm showed high diagnostic accuracy (91.24%) for the assessment of ICH with a sensitivity of 90.37% and specificity of 94.85%. To distinguish the different ICH types, the algorithm had a sensitivity of 93.47% and a specificity of 99.79%, with an accuracy of 98.54%. To detect midline shift, the algorithm had a sensitivity of 100%. In terms of processing time, the algorithm was significantly faster compared to the radiologist’s time to first diagnosis (15.37 ± 1.85 vs 277 ± 14 s, p < 0.001).

Conclusion

A novel deep learning algorithm can provide high diagnostic accuracy for the identification and classification of ICH from unenhanced CT scans, combined with short processing times. This has the potential to assist and improve radiologists’ ICH assessment in NCCT scans, especially in emergency scenarios, when time efficiency is needed.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Deep LearningTomography, X-Ray ComputedIntracranial HemorrhagesHumansRetrospective StudiesMaleFemaleMiddle AgedAdultAgedAlgorithmsBrain Injuries, TraumaticSensitivity and SpecificityTime FactorsAged, 80 and over

Stichwörter der Autor:innen

Artificial intelligenceBrain injuriesComputer-assistedDiagnosisIntracranial hemorrhageTomographyTraumaticX-ray computed

Autor:innen und Institutionen

  • Diagnostic and Interventional Radiology Unit, BIOMORF Department, University of Messina, Messina, Italy. tommasodang@gmail.com.
  • Department of Radiology and Nuclear Medicine, Erasmus MC, 3015 GD, Rotterdam, The Netherlands. tommasodang@gmail.com.
  • Diagnostic and Interventional Radiology Unit, BIOMORF Department, University of Messina, Messina, Italy.
  • Division of Experimental Imaging, Department of Diagnostic and Interventional Radiology, University Hospital Frankfurt, Frankfurt Am Main, Germany.
  • Department of Diagnostic and Interventional Radiology, University Hospital Frankfurt, Frankfurt Am Main, Germany.
  • Department of Diagnostic and Interventional Radiology and Neuroradiology, Bundeswehr Central Hospital Koblenz, Koblenz, Germany.

Zitieren

APA

D'Angelo, T., Bucolo, G. M., Kamareddine, T., Yel, I., Koch, V., Gruenewald, L. D., Martin, S., Alizadeh, L. S., Mazziotti, S., Blandino, A., Vogl, T. J., & Booz, C. (2024). Accuracy and time efficiency of a novel deep learning algorithm for Intracranial Hemorrhage detection in CT Scans. La radiologia medica, 129(10), 1499-1506. https://doi.org/10.1007/s11547-024-01867-y

Vancouver

D'Angelo T, Bucolo GM, Kamareddine T, Yel I, Koch V, Gruenewald LD, et al. Accuracy and time efficiency of a novel deep learning algorithm for Intracranial Hemorrhage detection in CT Scans. La radiologia medica. 2024;129(10):1499-1506. doi:10.1007/s11547-024-01867-y

BibTeX
@article{DAngelo2024,
  author = {D'Angelo, Tommaso and Bucolo, Giuseppe M and Kamareddine, Tarek and Yel, Ibrahim and Koch, Vitali and Gruenewald, Leon D and Martin, Simon and Alizadeh, Leona S and Mazziotti, Silvio and Blandino, Alfredo and Vogl, Thomas J and Booz, Christian},
  title = {Accuracy and time efficiency of a novel deep learning algorithm for Intracranial Hemorrhage detection in CT Scans},
  journal = {La radiologia medica},
  year = {2024},
  volume = {129},
  number = {10},
  pages = {1499--1506},
  doi = {10.1007/s11547-024-01867-y}
}
Kostenlos starten →