EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Spatial temperature monitoring of preterm infants using a multi-modal camera setup

Florian Voss, Simon Lyra, Celine Noehl, Milian Brasche, Konrad Heimann, Luisa Hensel, Thorsten Orlikowsky, Steffen Leonhardt, Markus Lueken

Medical & Biological Engineering & Computing · 2026-06 · 64 (6) · 2189-2208

DOI: 10.1007/s11517-026-03575-3 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical & Biological Engineering & Computing, online erschienen am 20.04.2026. Verfasst von Florian Voss und 8 weiteren Autor:innen. Beteiligte Einrichtung: Medical Information Technology, Helmholtz Institute for Biomedical Engineering, RWTH Aachen University, 52074, Aachen, Germany. voss@hia.rwth-aachen.…. Hauptthemen laut MEDLINE: Thermography, Infant, Premature. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Bundesministerium für Bildung, Wissenschaft, Forschung und Technologie, RWTH Aachen University.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Accurate and reliable skin temperature monitoring is critical for the thermoregulation of premature infants, but current methods using wired sensors are invasive and prone to error. Infrared thermography offers a non-invasive, wireless alternative to current wired temperature monitoring methods. This work presents a novel non-contact system for monitoring skin temperature of premature infants in closed incubators using infrared thermography. To enable temperature monitoring of specific body parts, an automated body part segmentation model was developed. The multimodal U-Net incorporated color and long-wave infrared image data. Transfer learning and data augmentation techniques improved the performance of the model, resulting in an average Intersection over Union of 0.77 across all body parts (head, torso, arms, and legs). To ensure the accuracy of the thermal data itself, a novel temperature correction algorithm was developed. This compensated for systematic errors caused by factors such as an infrared window, reflections from incubator walls, and camera drift. The algorithm achieved a mean absolute error of 0.17 [Formula: see text]C and a maximum error of 0.83 [Formula: see text]C when validated against a blackbody reference. By combining these steps, our system extracts spatial temperature using a percentile-based method, resulting in a mean absolute error of 0.41 [Formula: see text]C compared to a reference adhesive temperature sensor on the torso. The analysis revealed that the torso and arms provided more robust central and peripheral temperature measurements than the head and legs. These results demonstrate the potential of this non-contact system for accurate and reliable clinical temperature monitoring in premature infants, offering significant benefits in terms of patient comfort, reduced risk of infection and improved workflow for medical staff.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ThermographyInfant, PrematureHumansInfant, NewbornMonitoring, PhysiologicAlgorithmsSkin TemperatureInfrared RaysIncubators, InfantBody Temperature

Stichwörter der Autor:innen

Deep learningInfrared thermographyNeonatal intensive careTemperature

Autor:innen und Institutionen

  • Medical Information Technology, Helmholtz Institute for Biomedical Engineering, RWTH Aachen University, 52074, Aachen, Germany. voss@hia.rwth-aachen.de.
  • Medical Information Technology, Helmholtz Institute for Biomedical Engineering, RWTH Aachen University, 52074, Aachen, Germany.
  • Section of Neonatology, University Hospital RWTH Aachen, 52074, Aachen, Germany.

Förderung

  • Bundesministerium für Bildung, Wissenschaft, Forschung und Technologie
  • RWTH Aachen University

Zitieren

APA

Voss, F., Lyra, S., Noehl, C., Brasche, M., Heimann, K., Hensel, L., Orlikowsky, T., Leonhardt, S., & Lueken, M. (2026). Spatial temperature monitoring of preterm infants using a multi-modal camera setup. Medical & Biological Engineering & Computing, 64(6), 2189-2208. https://doi.org/10.1007/s11517-026-03575-3

Vancouver

Voss F, Lyra S, Noehl C, Brasche M, Heimann K, Hensel L, et al. Spatial temperature monitoring of preterm infants using a multi-modal camera setup. Medical & Biological Engineering & Computing. 2026;64(6):2189-2208. doi:10.1007/s11517-026-03575-3

BibTeX
@article{Voss2026,
  author = {Voss, Florian and Lyra, Simon and Noehl, Celine and Brasche, Milian and Heimann, Konrad and Hensel, Luisa and Orlikowsky, Thorsten and Leonhardt, Steffen and Lueken, Markus},
  title = {Spatial temperature monitoring of preterm infants using a multi-modal camera setup},
  journal = {Medical & Biological Engineering & Computing},
  year = {2026},
  volume = {64},
  number = {6},
  pages = {2189--2208},
  doi = {10.1007/s11517-026-03575-3}
}