Diffusion correction of Beer-Lambert law in visible light optical coherence tomography for retinal vessels

Yan Shi, Peng Miao, Shanbao Tong

Medical & Biological Engineering & Computing · 2026-06 · 64 (6) · 2177-2187

DOI: 10.1007/s11517-026-03563-7 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical & Biological Engineering & Computing, online erschienen am 20.04.2026. Verfasst von Yan Shi und 2 weiteren Autor:innen. Beteiligte Einrichtung: School of Biomedical Engineering, Shanghai Jiao Tong University, Shanghai, 200240, China. Hauptthemen laut MEDLINE: Retinal Vessels, Tomography, Optical Coherence. Die Arbeit nennt 31 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Vis-OCT is an emerging technique in ophthalmology, providing metabolic information of the retina, specifically oxygen saturation ([Formula: see text]), based on the modified Beer-Lambert law (mBLL). However, the conventional mBLL-based model simplifies the scattering attenuation with an exponential term accounting for the tissue absorption. This simplification reduces accuracy because multiple scattering is non-negligible. Additionally, the OCT spectral signal for a blood vessel is usually extracted from a region of interest (ROI) that is either located on the posterior wall or encompasses the entire vessel, thereby compromising spatial resolution. To overcome such limitations, we proposed a diffusion correction of the Beer-Lambert law (dcBLL) from the perspective of the radiative transfer equation (RTE), solved by the diffusion approximation, which takes both single and multiple scatterings into account. Pixel-wise [Formula: see text] estimation was validated through Monte Carlo simulations (MCS) and in vivo human retinal imaging. The dcBLL model demonstrated clear superiority in MCS, with RMSE values less than 10% of those obtained with the mBLL model. In human retinal oximetry, dcBLL further enables pixel-wise [Formula: see text] mapping, in contrast to the conventional approach that averages values over large vessels.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Retinal VesselsTomography, Optical CoherenceHumansMonte Carlo MethodDiffusionLightOxygenOximetryComputer Simulation

Stichwörter der Autor:innen

Beer-Lambert lawoxygen saturationretinal vesselsvisible-light optical coherence tomography

Substanzen

Oxygen

Autor:innen und Institutionen

  • School of Biomedical Engineering, Shanghai Jiao Tong University, Shanghai, 200240, China.
  • School of Biomedical Engineering, Shanghai Jiao Tong University, Shanghai, 200240, China. stong@sjtu.edu.cn.

Zitieren

APA

Shi, Y., Miao, P., & Tong, S. (2026). Diffusion correction of Beer-Lambert law in visible light optical coherence tomography for retinal vessels. Medical & Biological Engineering & Computing, 64(6), 2177-2187. https://doi.org/10.1007/s11517-026-03563-7

Vancouver

Shi Y, Miao P, Tong S. Diffusion correction of Beer-Lambert law in visible light optical coherence tomography for retinal vessels. Medical & Biological Engineering & Computing. 2026;64(6):2177-2187. doi:10.1007/s11517-026-03563-7

BibTeX
@article{Shi2026,
  author = {Shi, Yan and Miao, Peng and Tong, Shanbao},
  title = {Diffusion correction of Beer-Lambert law in visible light optical coherence tomography for retinal vessels},
  journal = {Medical & Biological Engineering & Computing},
  year = {2026},
  volume = {64},
  number = {6},
  pages = {2177--2187},
  doi = {10.1007/s11517-026-03563-7}
}
Kostenlos starten →