Understanding the mismatch between in-vivo and in-silico rhinomanometry
Medical & Biological Engineering & Computing · 2026
DOI: 10.1007/s11517-025-03450-7 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical & Biological Engineering & Computing, 2026. Verfasst von Marco Atzori und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Aerospace Science and Technology, Politecnico di Milano, Campus Bovisa, Milano, 20156, Italy. marco.atzori@polimi.it. Hauptthemen laut MEDLINE: Rhinomanometry, Computer Simulation. Die Arbeit nennt 28 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Ministero dell'Istruzione, dell'Università e della Ricerca, Ministero dell’Istruzione, dell’Università e della Ricerca.
Abstract (Originalsprache)
Numerical simulations and clinical measurements of nasal resistance are in quantitative disagreement. The order of magnitude of this mismatch, that sometimes exceeds 100%, is such that known sources of uncertainty cannot explain it. The goal of the present work is to examine a source of bias introduced by the design of medical devices, which has not been considered until now as a possible explanation. We study the effect of the location of the probe on the rhinomanometer that is meant to measure the ambient pressure. Rhinomanometry is carried out on a 3D silicone model of a patient-specific anatomy; a clinical device and dedicated sensors are employed side-by-side for mutual validation. The same anatomy is also employed for numerical simulations, with approaches spanning a wide range of fidelity levels. We find that the intrinsic uncertainty of the numerical simulations is of minor importance. To the contrary, the position of the pressure tap intended to acquire the external pressure in the clinical device is crucial, and can cause a mismatch comparable to that generally observed between in-silico and in-vivo rhinomanometry data. A source of systematic bias may therefore exist in rhinomanometers, designed under the assumption that measurements of the nasal resistance are unaffected by the flow development within the instruments. Graphical abstract
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Aerospace Science and Technology, Politecnico di Milano, Campus Bovisa, Milano, 20156, Italy. marco.atzori@polimi.it.
- Department of Aerospace Science and Technology, Politecnico di Milano, Campus Bovisa, Milano, 20156, Italy.
- Department of Aerospace Science and Technology, Politecnico di Milano, Campus Bovisa, Milano, 20156, Italy. maurizio.quadrio@polimi.it.
Förderung
- Ministero dell'Istruzione, dell'Università e della Ricerca
- Ministero dell’Istruzione, dell’Università e della Ricerca
Zitieren
APA
Atzori, M., Ciacci, G. D., & Quadrio, M. (2026). Understanding the mismatch between in-vivo and in-silico rhinomanometry. Medical & Biological Engineering & Computing. https://doi.org/10.1007/s11517-025-03450-7
Vancouver
Atzori M, Ciacci GD, Quadrio M. Understanding the mismatch between in-vivo and in-silico rhinomanometry. Medical & Biological Engineering & Computing. 2026. doi:10.1007/s11517-025-03450-7
BibTeX
@article{Atzori2026,
author = {Atzori, Marco and Ciacci, Gabriele Dini and Quadrio, Maurizio},
title = {Understanding the mismatch between in-vivo and in-silico rhinomanometry},
journal = {Medical & Biological Engineering & Computing},
year = {2026},
doi = {10.1007/s11517-025-03450-7}
}
EUVIMED