Ablation catheter–induced mechanical deformation in myocardium: computer modeling and ex vivo experiments
Medical & Biological Engineering & Computing · 2024
DOI: 10.1007/s11517-024-03135-7 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical & Biological Engineering & Computing, 2024. Verfasst von Yukako Ijima und 5 weiteren Autor:innen. Beteiligte Einrichtung: Biofluid Science and Engineering Laboratory, Global Course of Engineering and Science, Graduate School of Engineering and Science, Shibaura Institute…. Hauptthemen laut MEDLINE: Computer Simulation, Myocardium, Catheter Ablation. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 29 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Ministerio de Ciencia e Innovación, Shibaura Institute of Technology, Japan Student Services Organization.
Abstract (Originalsprache)
Cardiac catheter ablation requires an adequate contact between myocardium and catheter tip. Our aim was to quantify the relationship between the contact force (CF) and the resulting mechanical deformation induced by the catheter tip using an ex vivo model and computational modeling. The catheter tip was inserted perpendicularly into porcine heart samples. CF values ranged from 10 to 80 g. The computer model was built to simulate the same experimental conditions, and it considered a 3-parameter Mooney-Rivlin model based on hyper-elastic material. We found a strong correlation between the CF and insertion depth (ID) ( R 2 = 0.96, P < 0.001), from 0.7 ± 0.3 mm at 10 g to 6.9 ± 0.1 mm at 80 g. Since the surface deformation was asymmetrical, two transversal diameters (minor and major) were identified. Both diameters were strongly correlated with CF ( R 2 ≥ 0.95), from 4.0 ± 0.4 mm at 20 g to 10.3 ± 0.0 mm at 80 g (minor), and from 6.4 ± 0.7 mm at 20 g to 16.7 ± 0.1 mm at 80 g (major). An optimal fit between computer and experimental results was achieved, with a prediction error of 0.74 and 0.86 mm for insertion depth and mean surface diameter, respectively. Graphical Abstract
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Biofluid Science and Engineering Laboratory, Global Course of Engineering and Science, Graduate School of Engineering and Science, Shibaura Institute of Technology, 307 Fukasaku, Minuma-Ku, Saitama-City, Saitama, 337-85…
- Department of Industrial Engineering, Faculty of Engineering, Srinakharinwirot University, Ongkharak, Nakhon Nayok, Thailand.
- BioMIT, Department of Electronic Engineering, Universitat Politecnica de Valencia, Camino de Vera, 46022, Valencia, Spain.
- Translational Medical Device Lab, School of Medicine, University of Galway, Galway, Ireland.
- Universidad Internacional de Valencia, Valencia, Spain.
- BioMIT, Department of Electronic Engineering, Universitat Politecnica de Valencia, Camino de Vera, 46022, Valencia, Spain. eberjano@eln.upv.es.
Förderung
- Ministerio de Ciencia e Innovación
- Shibaura Institute of Technology
- Japan Student Services Organization
Zitieren
APA
Ijima, Y., Masnok, K., Perez, J. J., González-Suárez, A., Berjano, E., & Watanabe, N. (2024). Ablation catheter–induced mechanical deformation in myocardium: computer modeling and ex vivo experiments. Medical & Biological Engineering & Computing. https://doi.org/10.1007/s11517-024-03135-7
Vancouver
Ijima Y, Masnok K, Perez JJ, González-Suárez A, Berjano E, Watanabe N. Ablation catheter–induced mechanical deformation in myocardium: computer modeling and ex vivo experiments. Medical & Biological Engineering & Computing. 2024. doi:10.1007/s11517-024-03135-7
BibTeX
@article{Ijima2024,
author = {Ijima, Yukako and Masnok, Kriengsak and Perez, Juan J and González-Suárez, Ana and Berjano, Enrique and Watanabe, Nobuo},
title = {Ablation catheter–induced mechanical deformation in myocardium: computer modeling and ex vivo experiments},
journal = {Medical & Biological Engineering & Computing},
year = {2024},
doi = {10.1007/s11517-024-03135-7}
}
EUVIMED