Belun Sleep Platform versus in-lab polysomnography for obstructive sleep apnea diagnosis
Sleep and Breathing · 2025
DOI: 10.1007/s11325-025-03433-w ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Sleep and Breathing, 2025. Verfasst von Vipada Tirachaimongkol und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Otorhinolaryngology, Faculty of Medicine Siriraj Hospital, Mahidol University, 2 Wanglang Road, Bangkok Noi, Bangkok, 10700, Thailand. Hauptthemen laut MEDLINE: Sleep Apnea, Obstructive, Polysomnography, Monitoring, Ambulatory. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 17 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Mahidol University.
Abstract (Originalsprache)
Objective
We aimed to compare the Belun Sleep Platform (BSP), an artificial intelligence-driven home sleep testing device, with polysomnography (PSG) for diagnosing obstructive sleep apnea. The BSP analyzes oxygen saturation, heart rate, and accelerometry patterns.
Methods
Participants scheduled for PSG and with no significant cardiovascular or neuromuscular comorbidities were recruited. They underwent simultaneous in-laboratory, full-night PSG with the BSP. We assessed diagnostic properties, including sensitivity, specificity, positive predictive value, negative predictive value, and accuracy.
Results
A total of 40 participants (54.3% male) with a mean age of 49.9 years were enrolled. For an apnea-hypopnea index (AHI) cutoff of ≥ 15 events/h, BSP showed an accuracy of 68.5%, sensitivity of 35.2%, and specificity of 100% under American Academy of Sleep Medicine criteria 1 A and 1B. For AHI thresholds of ≥ 5 and ≥ 30 events/h, sensitivity was 82.1% and 33.3%, respectively, while specificity was 14.2% and 100%, respectively. BSP-AHI correlated moderately with PSG-AHI (intraclass correlation coefficient [ICC] = 0.737). BSP’s oxygen desaturation index (ODI) showed a strong correlation with PSG-ODI (ICC = 0.882). Moderate correlations were observed between BSP and PSG for non-rapid eye movement sleep duration (ICC = 0.736), rapid eye movement sleep duration (ICC = 0.664), total sleep time (ICC = 0.617), and sleep efficiency (ICC = 0.719).
Conclusions
The BSP’s high specificity but low sensitivity suggests it serves better as a confirmatory tool rather than a primary screening method. Its moderate concordance with PSG underscores its potential in settings where PSG is unavailable. However, further investigation is needed to refine its clinical applications.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Otorhinolaryngology, Faculty of Medicine Siriraj Hospital, Mahidol University, 2 Wanglang Road, Bangkok Noi, Bangkok, 10700, Thailand.
- Siriraj Sleep Center, Faculty of Medicine Siriraj Hospital, Mahidol University, Bangkok, Thailand.
- Department of Otorhinolaryngology, Faculty of Medicine Siriraj Hospital, Mahidol University, 2 Wanglang Road, Bangkok Noi, Bangkok, 10700, Thailand. mintkus32@hotmail.com.
ORCID-Profile: Wish Banhiran, Wattanachai Chotinaiwattarakul, Sarin Rungmanee, Jindapa Srikajon, Navarat Kasemsuk
Förderung
- Mahidol University
Zitieren
APA
Tirachaimongkol, V., Banhiran, W., Chotinaiwattarakul, W., Rungmanee, S., Pimolsri, C., Srikajon, J., & Kasemsuk, N. (2025). Belun Sleep Platform versus in-lab polysomnography for obstructive sleep apnea diagnosis. Sleep and Breathing. https://doi.org/10.1007/s11325-025-03433-w
Vancouver
Tirachaimongkol V, Banhiran W, Chotinaiwattarakul W, Rungmanee S, Pimolsri C, Srikajon J, et al. Belun Sleep Platform versus in-lab polysomnography for obstructive sleep apnea diagnosis. Sleep and Breathing. 2025. doi:10.1007/s11325-025-03433-w
BibTeX
@article{Tirachaimongkol2025,
author = {Tirachaimongkol, Vipada and Banhiran, Wish and Chotinaiwattarakul, Wattanachai and Rungmanee, Sarin and Pimolsri, Chawanon and Srikajon, Jindapa and Kasemsuk, Navarat},
title = {Belun Sleep Platform versus in-lab polysomnography for obstructive sleep apnea diagnosis},
journal = {Sleep and Breathing},
year = {2025},
doi = {10.1007/s11325-025-03433-w}
}
EUVIMED