Development of a novel phantom for evaluating jawbone SPECT targeting medication-related osteonecrosis of the jaw

Naoya Hayashi iD, Norikazu Matsutomo, Ryotaro Tokorodani, Mitsuha Fukami, Miki Nishimori, Hitomi Iwasa, Kie Nakatani, Tetsuya Yamamoto, Takuji Yamagami, Tomoaki Yamamoto

Oral Radiology · 2025

DOI: 10.1007/s11282-025-00809-2 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Oral Radiology, 2025. Verfasst von Naoya Hayashi und 9 weiteren Autor:innen. Beteiligte Einrichtung: Division of Radiology, Department of Medical Technology, Kochi Medical School Hospital, Kohasu, Oko-Cho, Nankoku, Kochi, 783-8505, Japan. jm-n.hayash…. Hauptthemen laut MEDLINE: Phantoms, Imaging, Tomography, Emission-Computed, Single-Photon, Bisphosphonate-Associated Osteonecrosis of the Jaw, Single Photon Emission Computed Tomography Computed Tomography. Die Arbeit nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

We developed a new phantom for technical evaluation of jawbone single-photon emission computed tomography (SPECT) examinations for medication-related osteonecrosis of the jaw (MRONJ). In this study, we verified the utility of the phantom by determining optimal image reconstruction parameters.

Methods

We evaluated the image quality and quantification in jawbone SPECT images obtained by different reconstruction parameters using the phantom. The phantom images were acquired using a SPECT/computed tomography (CT) system and then reconstructed using ordered-subset expectation maximization (OSEM) iterative reconstruction with resolution recovery as well as scatter and attenuation correction with various update numbers and Gaussian filter full width at half maximums (FWHMs). The percent contrast (%contrast) and absolute recovery coefficient were calculated to determine the optimal reconstruction parameters (OSEM jaw ). Nineteen patients with a clinical diagnosis of MRONJ who underwent bone SPECT/CT were enrolled for the clinical study. The performance of OSEM jaw was verified by comparison with OSEM current determined by a spherical phantom, using the correlation between the mean standardized uptake value (SUV mean ) and clinical staging and visual assessment as endpoints.

Results

In the phantom study, %contrast and absolute recovery coefficient increased with increasing update numbers. As the Gaussian filter FWHM increased, the quantitative accuracy and image sharpness decreased. The parameter determined by the phantom study (OSEM jaw ) recommended 120 updates and no filter. In the clinical study, the mean and standard deviation of SUV mean obtained from OSEM jaw were 8.9 ± 1.4 for stage 1 lesions, 12.9 ± 4.1 for stage 2 lesions, and 13.8 ± 1.4 for stage 3 lesions. For OSEM current , they were 5.4 ± 2.3 for stage 1 lesions, 8.3 ± 2.5 for stage 2 lesions, and 8.0 ± 0.9 for stage 3 lesions. The SUV mean obtained from OSEM jaw had a stronger correlation with clinical stage. Based on visual assessment, the quality of the SPECT images reconstructed by OSEM jaw (3.7 ± 0.9) was superior to that reconstructed by OSEM current (2.9 ± 1.1).

Conclusions

We developed a novel phantom and adapted it for technical evaluation. This study demonstrated the utility of the developed phantom.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Phantoms, ImagingTomography, Emission-Computed, Single-PhotonBisphosphonate-Associated Osteonecrosis of the JawSingle Photon Emission Computed Tomography Computed TomographyHumansMaleFemaleAgedMiddle AgedAged, 80 and over

Stichwörter der Autor:innen

Image reconstructionJawboneMRONJPhantomSPECTSUV

Autor:innen und Institutionen

  • Division of Radiology, Department of Medical Technology, Kochi Medical School Hospital, Kohasu, Oko-Cho, Nankoku, Kochi, 783-8505, Japan. jm-n.hayashi@kochi-u.ac.jp.
  • Department of Medical Radiological Technology, Faculty of Health Sciences, Kyorin University, B-524, 5-4-1 Shimorenjaku, Mitaka, Tokyo, 181-8612, Japan. jm-n.hayashi@kochi-u.ac.jp.
  • Department of Medical Radiological Technology, Faculty of Health Sciences, Kyorin University, B-524, 5-4-1 Shimorenjaku, Mitaka, Tokyo, 181-8612, Japan.
  • Division of Radiology, Department of Medical Technology, Kochi Medical School Hospital, Kohasu, Oko-Cho, Nankoku, Kochi, 783-8505, Japan.
  • Department of Diagnostic and Interventional Radiology, Kochi Medical School, Kochi University, Kohasu, Oko-Cho, Nankoku, Kochi, 783-8505, Japan.
  • Department of Oral and Maxillofacial Surgery, Kochi Medical School, Kochi University, Kohasu, Oko-Cho, Nankoku, Kochi, 783-8505, Japan.

ORCID-Profile: Naoya Hayashi

Zitieren

APA

Hayashi, N., Matsutomo, N., Tokorodani, R., Fukami, M., Nishimori, M., Iwasa, H., Nakatani, K., Yamamoto, T., Yamagami, T., & Yamamoto, T. (2025). Development of a novel phantom for evaluating jawbone SPECT targeting medication-related osteonecrosis of the jaw. Oral Radiology. https://doi.org/10.1007/s11282-025-00809-2

Vancouver

Hayashi N, Matsutomo N, Tokorodani R, Fukami M, Nishimori M, Iwasa H, et al. Development of a novel phantom for evaluating jawbone SPECT targeting medication-related osteonecrosis of the jaw. Oral Radiology. 2025. doi:10.1007/s11282-025-00809-2

BibTeX
@article{Hayashi2025,
  author = {Hayashi, Naoya and Matsutomo, Norikazu and Tokorodani, Ryotaro and Fukami, Mitsuha and Nishimori, Miki and Iwasa, Hitomi and Nakatani, Kie and Yamamoto, Tetsuya and Yamagami, Takuji and Yamamoto, Tomoaki},
  title = {Development of a novel phantom for evaluating jawbone SPECT targeting medication-related osteonecrosis of the jaw},
  journal = {Oral Radiology},
  year = {2025},
  doi = {10.1007/s11282-025-00809-2}
}
Kostenlos starten →