Development of an Infant Air-Jet Dry Powder Aerosol Delivery System (iDP-ADS) Including a New Multifunctional Bifurcating Two-Prong Nasal Interface
Pharmaceutical Research · 2025
DOI: 10.1007/s11095-024-03814-y ↗Kurz gefasst
Fachartikel in der Zeitschrift Pharmaceutical Research, 2025. Verfasst von Sarah C Strickler und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Mechanical and Nuclear Engineering, Virginia Commonwealth University, 401 West Main Street, P.O. Box 843015, Richmond, VA, 23284-3015, …. Hauptthemen laut MEDLINE: Dry Powder Inhalers, Drug Delivery Systems. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 78 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NHLBI NIH HHS, National Heart, Lung, and Blood Institute.
Abstract (Originalsprache)
Purpose
To improve the quality of aerosol delivery to infants, the iDP-ADS was advanced to include dual-prong nose-to-lung aerosol administration with a bifurcating interface, consistently monitor lung pressures and control ventilatory parameters with a pressure monitoring and control (PMC) unit, and implement flexible nasal prongs for use across a range of subject sizes.
Methods
Four bifurcating flow pathways were integrated into the iDP-ADS and tested in vitro with a full-term infant nose-throat (NT) model for comparison to the performance of a single-prong interface. After selecting the best-performing flow pathway, flexible prong designs were evaluated in the same model and chosen for additional testing. Realistic pulmonary mechanics (PM) and age-appropriate tidal volumes were used to simulate ventilation with the PMC unit and aerosol delivery in full-term and 34-week gestational age preterm NT models.
Results
Three of the four bifurcating flow pathways matched the performance of the single-prong design (tracheal filter delivery of ~55%), and the FP4 design with co-flow was selected. A flexible prong version of FP4 produced similar performance to the rigid version. Measurements from the PMC unit demonstrated that consistent air volumes under safe operating pressures could be delivered with a PEEP between 4–6 cmH 2 O. Considering aerosol delivery, PM conditions resulted in ~4% decrease in filter deposition but high lung delivery efficiencies of ~45% and ~34% for the full-term and preterm models, respectively.
Conclusions
The best-performing interface with flexible prongs matched the lung delivery efficiency of a high-transmission single-prong interface and delivered high aerosol doses through late-preterm to full-term NT models.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Mechanical and Nuclear Engineering, Virginia Commonwealth University, 401 West Main Street, P.O. Box 843015, Richmond, VA, 23284-3015, USA.
- Department of Pharmaceutics, Virginia Commonwealth University, Richmond, VA, USA.
- Department of Mechanical and Nuclear Engineering, Virginia Commonwealth University, 401 West Main Street, P.O. Box 843015, Richmond, VA, 23284-3015, USA. pwlongest@vcu.edu.
- Department of Pharmaceutics, Virginia Commonwealth University, Richmond, VA, USA. pwlongest@vcu.edu.
Förderung
- NHLBI NIH HHS
- National Heart, Lung, and Blood Institute
Zitieren
APA
Strickler, S. C., Farkas, D. R., Momin, M. A. M., Vargas, L., Aladwani, G., Hindle, M., & Longest, W. (2025). Development of an Infant Air-Jet Dry Powder Aerosol Delivery System (iDP-ADS) Including a New Multifunctional Bifurcating Two-Prong Nasal Interface. Pharmaceutical Research. https://doi.org/10.1007/s11095-024-03814-y
Vancouver
Strickler SC, Farkas DR, Momin MAM, Vargas L, Aladwani G, Hindle M, et al. Development of an Infant Air-Jet Dry Powder Aerosol Delivery System (iDP-ADS) Including a New Multifunctional Bifurcating Two-Prong Nasal Interface. Pharmaceutical Research. 2025. doi:10.1007/s11095-024-03814-y
BibTeX
@article{Strickler2025,
author = {Strickler, Sarah C and Farkas, Dale R and Momin, Mohammad A M and Vargas, Laura and Aladwani, Ghali and Hindle, Michael and Longest, Worth},
title = {Development of an Infant Air-Jet Dry Powder Aerosol Delivery System (iDP-ADS) Including a New Multifunctional Bifurcating Two-Prong Nasal Interface},
journal = {Pharmaceutical Research},
year = {2025},
doi = {10.1007/s11095-024-03814-y}
}
EUVIMED