Reliable and Reproducible Protocol for Evaluating Detection Sensitivity in Bioimaging Systems Using Near-Infrared Fluorescent Dyes

Qinglai Yan, Zhaotai Gu, Nana Li, Qinyuan Zhang, Siyue Wu, Haozhi Yang, Junyu Ren

Journal of Fluorescence · 2025

DOI: 10.1007/s10895-025-04523-4 ↗

Kurz gefasst

Studienprotokoll in der Zeitschrift Journal of Fluorescence, 2025. Verfasst von Qinglai Yan und 6 weiteren Autor:innen. Beteiligte Einrichtung: Zhejiang Institute of Medical Device Supervision and Testing, Hangzhou, 310018, Zhejiang, China. jimyan820504@163.com. Hauptthemen laut MEDLINE: Fluorescent Dyes, Indocyanine Green, Optical Imaging, Infrared Rays. Die Arbeit nennt 23 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The traditional method of establishing the relationship between fluorescence brightness and dye concentration through concentration gradient, and evaluating the sensitivity of imaging detection is too repetitive and cumbersome, and is obviously affected by handwork. Therefore, the reliability and reproducibility of the data are low. This work provides a standard concentration and imaging luminance relationship curve for equipment manufacturers through concentration preparation and detection quantification by authoritative institutions based on indocyanine green dye. It only needs to detect the excitation light power and simulated fluorescence radiation of the equipment in the imaging system, and find the relationship curve to obtain the relationship between "radiation and concentration" and "signal-to-background ratio-concentration", which can be used as a quantitative basis for the camera system. All operations can achieve traceability operation, which greatly reduces the human experimental error, and provides reliable evidence for the detection sensitivity evaluation of bioimaging system based on fluorescent dyes.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Fluorescent DyesIndocyanine GreenOptical ImagingInfrared RaysHumansReproducibility of Results

Stichwörter der Autor:innen

Indocyanine greenNear-infrared fluorescence imagingQuantitative standardizationRadiometric calibrationSimulation of near-infrared fluorescence

Substanzen

Fluorescent DyesIndocyanine Green

Autor:innen und Institutionen

  • Zhejiang Institute of Medical Device Supervision and Testing, Hangzhou, 310018, Zhejiang, China. jimyan820504@163.com.
  • Guangdong OptoMedic Technologies, Inc, Foshan, 528200, Guangdong, China.
  • Zhejiang Institute of Medical Device Supervision and Testing, Hangzhou, 310018, Zhejiang, China.

Zitieren

APA

Yan, Q., Gu, Z., Li, N., Zhang, Q., Wu, S., Yang, H., & Ren, J. (2025). Reliable and Reproducible Protocol for Evaluating Detection Sensitivity in Bioimaging Systems Using Near-Infrared Fluorescent Dyes. Journal of Fluorescence. https://doi.org/10.1007/s10895-025-04523-4

Vancouver

Yan Q, Gu Z, Li N, Zhang Q, Wu S, Yang H, et al. Reliable and Reproducible Protocol for Evaluating Detection Sensitivity in Bioimaging Systems Using Near-Infrared Fluorescent Dyes. Journal of Fluorescence. 2025. doi:10.1007/s10895-025-04523-4

BibTeX
@article{Yan2025,
  author = {Yan, Qinglai and Gu, Zhaotai and Li, Nana and Zhang, Qinyuan and Wu, Siyue and Yang, Haozhi and Ren, Junyu},
  title = {Reliable and Reproducible Protocol for Evaluating Detection Sensitivity in Bioimaging Systems Using Near-Infrared Fluorescent Dyes},
  journal = {Journal of Fluorescence},
  year = {2025},
  doi = {10.1007/s10895-025-04523-4}
}
Kostenlos starten →