Augmented reality for ultrasound-guided venipuncture: a pilot study on a simulated vascular access model

Cosimo Aliani, Alberto Morelli, Simona Cristina Aiello, Fulvio Pinelli, Gianluca Villa, Stefano Romagnoli, Leonardo Bocchi

Journal of Clinical Monitoring and Computing · 15.04.2026 · 40 (4) · 917-922

DOI: 10.1007/s10877-026-01437-9 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Clinical Monitoring and Computing, online erschienen am 15.04.2026. Verfasst von Cosimo Aliani und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Information Engineering, University of Florence, Via di Santa Marta 3, Florence, 50139, Italy. Hauptthemen laut MEDLINE: Phlebotomy, Ultrasonography, Interventional, Augmented Reality. Die Arbeit nennt 11 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Università degli Studi di Firenze.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

To evaluate the feasibility of a low-cost augmented reality (AR) workflow for ultrasound-guided venipuncture and whether AR visualisation is non-inferior to conventional monitor guidance in total procedure time while improving user experience. Real-time ultrasound video from a conventional scanner was captured via an HDMI-to-USB device, pre-processed with a Python/OpenCV pipeline, and streamed over Wi-Fi to a HoloLens 2 headset using WebSocket. A turkey-thigh model with an embedded fluid-filled catheter simulated a peripheral vein. One experienced implantation nurse performed 58 venipuncture attempts with standard monitor guidance and 58 with AR (monitor blanked during AR). Total procedure time (TPT) was recorded for each attempt, and a questionnaire assessed perceived difficulty, image quality, usability, reliability, and ergonomic comfort. With all data included, overall TPT was comparable between standard and AR guidance ( $$21.32 \pm 7.90$$ s vs. $$21.64 \pm 10.30$$ s; $$p = 0.845$$ ). After excluding predefined outliers, overall TPT remained non-significantly lower with AR ( $$21.40 \pm 6.78$$ s vs. $$19.90 \pm 6.34$$ s; $$p = 0.193$$ ); on day 2 the p -value decreased to 0.103 ( $$20.45 \pm 6.32$$ s vs. $$18.30 \pm 5.12$$ s). Questionnaire responses favoured AR for gaze continuity, perceived coordination, usability, and ergonomic comfort, with acceptable image quality. AR-assisted ultrasound-guided venipuncture was feasible and did not add measurable temporal overhead for an expert operator, while improving perceived ergonomics. Larger, multi-operator clinical studies are needed to confirm performance effects and clinical impact.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

PhlebotomyUltrasonography, InterventionalAugmented RealityHumansPilot ProjectsReproducibility of ResultsFeasibility StudiesSurveys and QuestionnairesUltrasonographyUser-Computer InterfaceWorkflowComputer SimulationVascular Access DevicesAnimalsCatheterization, Peripheral

Stichwörter der Autor:innen

Augmented realityHead-mounted displayUltrasound-guided venipunctureVascular access

Autor:innen und Institutionen

  • Department of Information Engineering, University of Florence, Via di Santa Marta 3, Florence, 50139, Italy.
  • Department of Health Sciences, University of Florence, Viale Pieraccini 6, Florence, 50139, Italy.
  • Vascular Access Center, Azienda Ospedaliera Universitaria Careggi, Florence, 50139, Italy.
  • Department of Information Engineering, University of Florence, Via di Santa Marta 3, Florence, 50139, Italy. leonardo.bocchi@unifi.it.

Förderung

  • Università degli Studi di Firenze

Zitieren

APA

Aliani, C., Morelli, A., Aiello, S. C., Pinelli, F., Villa, G., Romagnoli, S., & Bocchi, L. (2026). Augmented reality for ultrasound-guided venipuncture: a pilot study on a simulated vascular access model. Journal of Clinical Monitoring and Computing, 40(4), 917-922. https://doi.org/10.1007/s10877-026-01437-9

Vancouver

Aliani C, Morelli A, Aiello SC, Pinelli F, Villa G, Romagnoli S, et al. Augmented reality for ultrasound-guided venipuncture: a pilot study on a simulated vascular access model. Journal of Clinical Monitoring and Computing. 2026;40(4):917-922. doi:10.1007/s10877-026-01437-9

BibTeX
@article{Aliani2026,
  author = {Aliani, Cosimo and Morelli, Alberto and Aiello, Simona Cristina and Pinelli, Fulvio and Villa, Gianluca and Romagnoli, Stefano and Bocchi, Leonardo},
  title = {Augmented reality for ultrasound-guided venipuncture: a pilot study on a simulated vascular access model},
  journal = {Journal of Clinical Monitoring and Computing},
  year = {2026},
  volume = {40},
  number = {4},
  pages = {917--922},
  doi = {10.1007/s10877-026-01437-9}
}
Kostenlos starten →