EUVIMEDEUVIMED entdecken
BeobachtungsstudieVergleichsstudieOpen Access · CC BY 4.0

Agreement between three noninvasive temperature monitoring devices during spinal anaesthesia for caesarean delivery: a prospective observational study

D O Vawda iD, Christopher King iD, L du Toit iD, R A Dyer iD, N J Masuku iD, D G Bishop iD

Journal of Clinical Monitoring and Computing · 2024

DOI: 10.1007/s10877-024-01154-1 ↗

Kurz gefasst

Beobachtungsstudie in der Zeitschrift Journal of Clinical Monitoring and Computing, 2024. Verfasst von D O Vawda und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Anaesthesia, School of Clinical Medicine, College of Health Sciences, Grey's Hospital, University of KwaZulu-Natal, Pietermaritzburg, 3…. Hauptthemen laut MEDLINE: Anesthesia, Spinal, Cesarean Section, Body Temperature, Monitoring, Intraoperative. Die Arbeit nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: University of KwaZulu-Natal.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Hypothermia during obstetric spinal anaesthesia is a common and important problem, yet temperature monitoring is often not performed due to the lack of a suitable, cost-effective monitor. This study aimed to compare a noninvasive core temperature monitor with two readily available peripheral temperature monitors during obstetric spinal anaesthesia. We undertook a prospective observational study including elective and emergency caesarean deliveries, to determine the agreement between affordable reusable surface temperature monitors (Welch Allyn SureTemp® Plus oral thermometer and the Braun 3-in-1 No Touch infrared thermometer) and the Dräger T-core© (using dual-sensor heat flux technology), in detecting thermoregulatory changes during obstetric spinal anaesthesia. Predetermined clinically relevant limits of agreement (LOA) were set at ± 0.5 °C. We included 166 patients in our analysis. Hypothermia (heat flux temperature < 36 °C) occurred in 67% (95% CI 49 to 78%). There was poor agreement between devices. In the Bland-Altman analysis, LOA for the heat flux monitor vs. oral thermometer were 1.8 °C (CI 1.7 to 2.0 °C; bias 0.5 °C), for heat flux monitor vs. infrared thermometer LOA were 2.3 °C (CI 2.1 to 2.4 °C; bias 0.4 °C) and for infrared vs. oral thermometer, LOA were 2.0 °C (CI 1.9 to 2.2 °C; bias 0.1 °C). Error grid analysis highlighted a large amount of clinical disagreement between methods. While monitoring of core temperature during obstetric spinal anaesthesia is clinically important, agreement between monitors was below clinically acceptable limits. Future research with gold-standard temperature monitors and exploration of causes of sensor divergence is needed.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Anesthesia, SpinalCesarean SectionBody TemperatureMonitoring, IntraoperativeThermometersHypothermiaHumansFemaleProspective StudiesPregnancyAdultAnesthesia, ObstetricalReproducibility of ResultsBody Temperature RegulationThermometryMonitoring, Physiologic

Stichwörter der Autor:innen

Caesarean deliveryHypothermiaObstetric anaesthesiaSpinal anaesthesiaTemperature monitoringThermoregulation

Autor:innen und Institutionen

  • Department of Anaesthesia, School of Clinical Medicine, College of Health Sciences, Grey's Hospital, University of KwaZulu-Natal, Pietermaritzburg, 3201, South Africa.
  • Department of Anesthesiology, Washington University School of Medicine in St Louis, MO, USA.
  • Department of Anaesthesia and Perioperative Medicine, University of Cape Town, Cape Town, South Africa.
  • Department of Anaesthesia, School of Clinical Medicine, College of Health Sciences, Grey's Hospital, University of KwaZulu-Natal, Pietermaritzburg, 3201, South Africa. davidgbishop@gmail.com.

ORCID-Profile: D O Vawda, Christopher King, L du Toit, R A Dyer, N J Masuku, D G Bishop

Förderung

  • University of KwaZulu-Natal

Zitieren

APA

Vawda, D. O., King, C., Toit, L. D., Dyer, R. A., Masuku, N. J., & Bishop, D. G. (2024). Agreement between three noninvasive temperature monitoring devices during spinal anaesthesia for caesarean delivery: a prospective observational study. Journal of Clinical Monitoring and Computing. https://doi.org/10.1007/s10877-024-01154-1

Vancouver

Vawda DO, King C, Toit LD, Dyer RA, Masuku NJ, Bishop DG. Agreement between three noninvasive temperature monitoring devices during spinal anaesthesia for caesarean delivery: a prospective observational study. Journal of Clinical Monitoring and Computing. 2024. doi:10.1007/s10877-024-01154-1

BibTeX
@article{Vawda2024,
  author = {Vawda, D O and King, Christopher and Toit, L du and Dyer, R A and Masuku, N J and Bishop, D G},
  title = {Agreement between three noninvasive temperature monitoring devices during spinal anaesthesia for caesarean delivery: a prospective observational study},
  journal = {Journal of Clinical Monitoring and Computing},
  year = {2024},
  doi = {10.1007/s10877-024-01154-1}
}
Kostenlos starten →