Increasing the reflection efficiency of the Sedaconda ACD-S by heating and cooling the anaesthetic reflector: a bench study using a test lung

Andreas Meiser, Pierre Louis Fernando Meis, Brian O'Gara, Thomas Volk, Azzeddine Kermad

Journal of Clinical Monitoring and Computing · 2023

DOI: 10.1007/s10877-022-00967-2 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Clinical Monitoring and Computing, 2023. Verfasst von Andreas Meiser und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Anaesthesiology, Intensive Care Medicine and Pain Medicine, Saarland University Hospital Medical Center, Homburg, Saarland, Germany. Hauptthemen laut MEDLINE: Isoflurane, Anesthetics, Inhalation. Die Arbeit nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Universitätsklinikum des Saarlandes und Medizinische Fakultät der Universität des Saarlandes.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

As volatile anaesthetic gases contribute to global warming, improving the efficiency of their delivery can reduce their environmental impact. This can be achieved by rebreathing from a circle system, but also by anaesthetic reflection with an open intensive care ventilator. We investigated whether the efficiency of such a reflection system could be increased by warming the reflector during inspiration and cooling it during expiration (thermocycling).

Methods

The Sedaconda-ACD-S (Sedana Medical, Danderyd, Sweden) was connected between an intensive care ventilator and a test lung. Liquid isoflurane was infused into the device at 0.5, 1.0, 2.0 and 5.0 mL/h; ventilator settings were 500 mL tidal volume, 12 bpm, 21% oxygen. Isoflurane concentrations were measured inside the test lung after equilibration. Thermocycling was achieved by heating the breathing gas in the inspiratory hose to 37 °C via a heated humidifier without water. Breathing gas expired from the test lung was cooled to 14 °C before reaching the ACD-S. In the test lung, body temperature pressure saturated conditions prevailed. Isoflurane concentrations and reflective efficiency were compared between thermocycling and control conditions.

Results

With thermocycling higher isoflurane concentrations in the test lung were measured for all infusion rates studied. Interpolation of data showed that for achieving 0.4 (0.6) Vol% isoflurane, the infusion rate can be reduced from 1.2 to 0.7 (2.0 to 1.2) mL/h or else to 56% (58%) of control.

Conclusion

Thermocycling of the anaesthetic gas considerably increases the efficiency of the anaesthetic reflector and reduces anaesthetic consumption by almost half in a test lung model. Given that cooling can be miniaturized, this method carries a potential for further saving anaesthetics in clinical practice in the operating theatre as well as for inhaled sedation in the ICU.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

IsofluraneAnesthetics, InhalationHumansAnesthesia, InhalationHeatingLung

Stichwörter der Autor:innen

AnaConDa-SBench studyInhaled sedationIsoflurane

Substanzen

IsofluraneAnesthetics, Inhalation

Autor:innen und Institutionen

  • Department of Anaesthesiology, Intensive Care Medicine and Pain Medicine, Saarland University Hospital Medical Center, Homburg, Saarland, Germany.
  • Department of Anaesthesiology, Beth Israel Deaconess Medical Center, Boston, MA, USA.
  • Department of Anaesthesiology, Intensive Care Medicine and Pain Medicine, Saarland University Hospital Medical Center, Homburg, Saarland, Germany. azzeddine.kermad@dhzc-charite.de.
  • Department of Cardiac Anesthesiology and Intensive Care Medicine, German Heart Center Charité, Berlin, Germany. azzeddine.kermad@dhzc-charite.de.

Förderung

  • Universitätsklinikum des Saarlandes und Medizinische Fakultät der Universität des Saarlandes

Zitieren

APA

Meiser, A., Meis, P. L. F., O'Gara, B., Volk, T., & Kermad, A. (2023). Increasing the reflection efficiency of the Sedaconda ACD-S by heating and cooling the anaesthetic reflector: a bench study using a test lung. Journal of Clinical Monitoring and Computing. https://doi.org/10.1007/s10877-022-00967-2

Vancouver

Meiser A, Meis PLF, O'Gara B, Volk T, Kermad A. Increasing the reflection efficiency of the Sedaconda ACD-S by heating and cooling the anaesthetic reflector: a bench study using a test lung. Journal of Clinical Monitoring and Computing. 2023. doi:10.1007/s10877-022-00967-2

BibTeX
@article{Meiser2023,
  author = {Meiser, Andreas and Meis, Pierre Louis Fernando and O'Gara, Brian and Volk, Thomas and Kermad, Azzeddine},
  title = {Increasing the reflection efficiency of the Sedaconda ACD-S by heating and cooling the anaesthetic reflector: a bench study using a test lung},
  journal = {Journal of Clinical Monitoring and Computing},
  year = {2023},
  doi = {10.1007/s10877-022-00967-2}
}
Kostenlos starten →