EUVIMEDEUVIMED entdecken
Multizentrische StudieOpen Access · CC BY-NC-ND 4.0

Atrial helix-fixation leadless pacemaker: real-world single-chamber implant experience

Devi G Nair, Ken W Lee, Nima Badie, Leonard Ganz, Kyungmoo Ryu, Cyrus A Hadadi

Journal of Interventional Cardiac Electrophysiology · 2025

DOI: 10.1007/s10840-025-02041-8 ↗

Kurz gefasst

Multizentrische Studie in der Zeitschrift Journal of Interventional Cardiac Electrophysiology, 2025. Verfasst von Devi G Nair und 5 weiteren Autor:innen. Beteiligte Einrichtung: St. Bernards Healthcare, Jonesboro, AR, USA. drdevignair@gmail.com. Hauptthemen laut MEDLINE: Pacemaker, Artificial, Sick Sinus Syndrome. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 20 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: Single -chamber ventricular leadless pacemakers (LPs) are well established. A dual-chamber LP system has recently become available with distinct atrial and ventricular devices. Single-chamber atrial pacing with transvenous devices is infrequent due to future upgrade concerns. This multi-center study evaluated the initial real-world use of the atrial LP by itself to treat isolated sinus node dysfunction (SND). METHODS: SND patients with normal PR interval and AV conduction to be implanted with atrial LPs after commercial US release were consecutively included. Procedural characteristics were evaluated, and electrical parameters were measured during pre-fixation mapping, post-fixation tether mode, after LP release, and before patient discharge. Acute, 30-day procedure- or device-related complications were noted. RESULTS: Aveir AR devices were implanted per standard local practice (N = 75 patients; 3 centers; 72 ± 13 years; 52% male; 92% de novo) with 100% success and placed predominantly in the right atrial appendage base (83%). The total procedure duration (from first incision to final suture) was 36 ± 33 min and the cumulative fluoroscopy duration was 7 ± 8 min. Pre-fixation mapping made repositioning unnecessary in 95% of implants. Pacing capture threshold at 0.4 ms pulse width, sensed amplitude, and impedance values of 0.6 ± 0.6 V, 2.9 ± 1.5 mV, and 329 ± 46 Ω, respectively, were measured prior to patient discharge. Capture threshold and sensed amplitude had improved significantly from LP release to patient discharge. No acute complications were observed. CONCLUSION: This initial real-world experience implanting the helix-fixation, single-chamber, atrial LP in SND patients demonstrated safe and efficient implantation, clinically acceptable electrical metrics, and no acute complications.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Pacemaker, ArtificialSick Sinus SyndromeHumansMaleFemaleAgedTreatment OutcomeHeart AtriaEquipment Design

Stichwörter der Autor:innen

Atrial pacemakerAveirHelix fixationImplant procedureLeadless pacemakerPacemaker complications

Autor:innen und Institutionen

  • St. Bernards Healthcare, Jonesboro, AR, USA. drdevignair@gmail.com.
  • Arkansas & White River Medical Center, Batesville, AR, USA. drdevignair@gmail.com.
  • Arrhythmia Research Group, Jonesboro, USA. drdevignair@gmail.com.
  • University of Arkansas Medical Sciences, Little Rock, AR, USA. drdevignair@gmail.com.
  • Health First Medical Group, Melbourne, FL, USA.
  • Abbott, Sunnyvale, CA, USA.

Zitieren

APA

Nair, D. G., Lee, K. W., Badie, N., Ganz, L., Ryu, K., & Hadadi, C. A. (2025). Atrial helix-fixation leadless pacemaker: real-world single-chamber implant experience. Journal of Interventional Cardiac Electrophysiology. https://doi.org/10.1007/s10840-025-02041-8

Vancouver

Nair DG, Lee KW, Badie N, Ganz L, Ryu K, Hadadi CA. Atrial helix-fixation leadless pacemaker: real-world single-chamber implant experience. Journal of Interventional Cardiac Electrophysiology. 2025. doi:10.1007/s10840-025-02041-8

BibTeX
@article{Nair2025,
  author = {Nair, Devi G and Lee, Ken W and Badie, Nima and Ganz, Leonard and Ryu, Kyungmoo and Hadadi, Cyrus A},
  title = {Atrial helix-fixation leadless pacemaker: real-world single-chamber implant experience},
  journal = {Journal of Interventional Cardiac Electrophysiology},
  year = {2025},
  doi = {10.1007/s10840-025-02041-8}
}
Kostenlos starten →