Fachartikel in der Zeitschrift Journal of Interventional Cardiac Electrophysiology, 2024. Verfasst von Lin-Thiri Toon und 6 weiteren Autor:innen. Beteiligte Einrichtung: Cardiac Rhythm Management Research Department, University Hospital Southampton, Southampton, SO16 6YD, UK. lin.toon@uhs.nhs.uk. Hauptthemen laut MEDLINE: Pacemaker, Artificial, Ambulatory Surgical Procedures, Prosthesis Implantation. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 11 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
MICRA implantation is not commonly done as a day-case procedure. Elective leadless pacemakers are implanted routinely in our centre.
Objective
To assess whether the day-case MICRA procedure is safe.
Methods
We retrospectively collected data from all patients undergoing elective MICRA implantation at our centre between May 2014 and Nov 2022 ( n = 81). Two patient groups were stratified: those planned to be discharged on the same day (SD, n = 52) and those planned to be observed overnight after the procedure (ON, n = 29). Patient demographics, size of the sheath used, type of MICRA device, and rate of complications were recorded. In patients with successful implants ( n = 80), device function at discharge and first routine follow-up were evaluated.
Results
There were 58% males in the SD group and 45% in the ON group. Median age was 49 years in the SD and 67 years in the ON. Among patients who were planned as a day case, 8 patients had to stay in the hospital but for < 48 h: 2 due to minor groin bleeding, 1 due to patient’s request despite fit to discharge, 4 due to the procedure carried out later in the day, and 1 for observation due to procedural complexity. MICRA implantation was successful in 80 patients. The rate of the major complications was 2% in the SD group and 7% in the ON group ( p = 0.223), and none of the co-morbidities assessed showed an association with any complications. Device parameters at the follow-up were available in 76 patients. The rate of patients with low and stable PCT at follow-up was also 98% in the SD group and 96% in the ON group.
Conclusions
Day case MICRA procedure can be performed safely in an appropriately selected patient population. Graphical abstract
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Cardiac Rhythm Management Research Department, University Hospital Southampton, Southampton, SO16 6YD, UK. lin.toon@uhs.nhs.uk.
- University of Southampton, Southampton, UK. lin.toon@uhs.nhs.uk.
- Cardiac Rhythm Management Research Department, University Hospital Southampton, Southampton, SO16 6YD, UK.
- University of Southampton, Southampton, UK.
Zitieren
APA
Toon, L. T., ElRefai, M., Abouelasaad, M., Patil, R., Paisey, J., Yue, A., & Roberts, P. (2024). Is day-case surgical procedure safe for MICRA leadless pacemaker implantation?. Journal of Interventional Cardiac Electrophysiology. https://doi.org/10.1007/s10840-024-01907-7
Vancouver
Toon LT, ElRefai M, Abouelasaad M, Patil R, Paisey J, Yue A, et al. Is day-case surgical procedure safe for MICRA leadless pacemaker implantation?. Journal of Interventional Cardiac Electrophysiology. 2024. doi:10.1007/s10840-024-01907-7
BibTeX
@article{Toon2024,
author = {Toon, Lin-Thiri and ElRefai, Mohammed and Abouelasaad, Mohamed and Patil, Roopa and Paisey, John and Yue, Arthur and Roberts, Paul},
title = {Is day-case surgical procedure safe for MICRA leadless pacemaker implantation?},
journal = {Journal of Interventional Cardiac Electrophysiology},
year = {2024},
doi = {10.1007/s10840-024-01907-7}
}
EUVIMED