Interocular biometric parameters comparison measured with swept-source technology
International Ophthalmology · 2022-01 · 42 (1) · 239-251
DOI: 10.1007/s10792-021-02020-8 ↗Kurz gefasst
Fachartikel in der Zeitschrift International Ophthalmology, online erschienen am 21.08.2021. Verfasst von César Albarrán-Diego und 4 weiteren Autor:innen. Beteiligte Einrichtung: Optics, Optometry and Vision Science Department, Faculty of Physics, University of Valencia, Dr Moliner 50, 46100, Burjassot, Spain. Hauptthemen laut MEDLINE: Axial Length, Eye, Tomography, Optical Coherence. Die Arbeit wurde bisher 14-mal zitiert (Crossref) und nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Universidad Complutense de Madrid.
Abstract (Originalsprache)
Purpose
In the event that any ocular parameter involved in the calculation of intraocular lens power could not be properly measured in one eye, it is important to know whether clinically relevant differences between both eyes can be expected. The aim of this work is to evaluate the symmetry of interocular biometric parameters.
Methods
This was a prospective, cross-sectional study involving 4090 subjects. Patients underwent consecutive swept-source optical biometry performed with an IOLMaster 700 (Carl Zeiss Meditec AG, Jena, Germany). The biometric parameters that were evaluated were: axial length (AL), mean anterior curvature (Rm), anterior chamber depth (ACD), crystalline lens thickness (LT), central corneal thickness (CCT) and white-to-white (WTW). The Chang–Waring chord distance (CWC-D) and the Chang–Waring chord angle (CWC-A) were also evaluated.
Results
There is an excellent correlation between both eyes for almost all the biometric parameters under study, with the exception of the CWC. Agreement for AL was better for eyes shorter than 24 mm. The linearity of the OD-vs-OS relationship can be correctly assumed for all parameters (Cusum test: p > 0.05 in all cases).
Conclusion
There are no clinically significant interocular differences for the biometric parameters under study, although for all of them, except the LT, statistically significant differences did arise. In the case of AL, moderate differences can be expected in eyes larger than 24 mm.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Optics, Optometry and Vision Science Department, Faculty of Physics, University of Valencia, Dr Moliner 50, 46100, Burjassot, Spain.
- , Miranza OA Madrid. C/Galileo 104, 28003, Madrid, Spain.
- , Miranza OA Madrid. C/Galileo 104, 28003, Madrid, Spain. nugarzon@opt.ucm.es.
- Faculty of Optics and Optometry, Complutense University of Madrid, c/Arcos de Jalón, 118, 28037, Madrid, Spain. nugarzon@opt.ucm.es.
- Faculty of Optics and Optometry, Complutense University of Madrid, c/Arcos de Jalón, 118, 28037, Madrid, Spain.
Förderung
- Universidad Complutense de Madrid
Zitieren
APA
Albarrán-Diego, C., Poyales, F., López-Artero, E., Garzón, N., & García-Montero, M. (2022). Interocular biometric parameters comparison measured with swept-source technology. International Ophthalmology, 42(1), 239-251. https://doi.org/10.1007/s10792-021-02020-8
Vancouver
Albarrán-Diego C, Poyales F, López-Artero E, Garzón N, García-Montero M. Interocular biometric parameters comparison measured with swept-source technology. International Ophthalmology. 2022;42(1):239-251. doi:10.1007/s10792-021-02020-8
BibTeX
@article{AlbarrnDiego2022,
author = {Albarrán-Diego, César and Poyales, Francisco and López-Artero, Esther and Garzón, Nuria and García-Montero, María},
title = {Interocular biometric parameters comparison measured with swept-source technology},
journal = {International Ophthalmology},
year = {2022},
volume = {42},
number = {1},
pages = {239--251},
doi = {10.1007/s10792-021-02020-8}
}
EUVIMED