Static and dynamic characteristics of the pupil light reflex obtained in Indian eyes using a hand-held, objective pupillometer

Erum Nisha, Rahul Negi, Tithi Halder, Manasa Kalivemula, Nabeel Quadri, Ayush Kumar, Manjushree Bhate, Virender Sachdeva, Shrikant R Bharadwaj

Documenta Ophthalmologica · 30.10.2026

DOI: 10.1007/s10633-026-10160-4 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Documenta Ophthalmologica, online erschienen am 30.10.2026. Verfasst von Erum Nisha und 8 weiteren Autor:innen. Beteiligte Einrichtung: Center for Technology Innovation, Kallam Anji Reddy Campus L V Prasad Eye Institute, Road 2 Banjara Hill, Hyderabad, Telangana, 500034, India. Die Arbeit nennt 49 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To establish the baseline data of static and dynamic characteristics of the pupil light reflex in Indian eyes across a wide age range using Pupil N, a newly developed handheld, objective pupillometer and thereby compare the present outcomes with a comparable dataset in Indian eyes obtained using the NPi-200, the current standard of care in objective pupillometry.

Methods

Pupil N elicits the light reflex by delivering two 1-s-long visible light pulses of 125 lx intensity onto the eye with 4-s of interstimulus interval, all after initial 6-s of dark adaptation. Pupil diameters recorded using an infrared sensitive camera (30 fps) are processed using a commercial smartphone to derive six outcome variables: pre- and post-stimulated diameter, miosis (normalized to pre-stimulation values), constriction and dilation peak velocities and response latency. Baseline data were collected in this study in both eyes on 290 healthy participants (3–86 years of age) and the right eye data of adults were compared with those obtained using NPi-200 by Rattan et al. [ 37 ].

Results

Baseline values calculated for each age decade showed younger eyes having larger pre-and post-stimulation diameters, faster constriction and dilation peak velocities, relative to older eyes ( p < 0.001). The magnitude of miosis and response latency were similar across ages, except for larger miosis and shorter latencies in the first decade of life ( p < 0.001). Data trends were similar to Rattan et al., but the absolute values differed across outcome variables ( p < 0.001), except for pre-stimulation diameter ( p = 0.06).

Conclusions

The baseline values of pupil light reflex reported here for Pupil N may be used as future reference standards for interpreting data of individuals with neurological dysfunction. The inter-device differences reported here call for caution while using these devices interchangeably for eliciting/interpreting the pupil light reflex in clinical and research environments.

Schlagwörter

Stichwörter der Autor:innen

Constriction velocityLatencyMiosis and mydriasisNeuro critical careObjective pupillometry

Autor:innen und Institutionen

  • Center for Technology Innovation, Kallam Anji Reddy Campus L V Prasad Eye Institute, Road 2 Banjara Hill, Hyderabad, Telangana, 500034, India.
  • Child Sight Institute, Nimmagadda Prasad Children's Eye Care Centre, GMR Varalakshmi Campus, L V Prasad Eye Institute, Hanumanthawaka Junction, Visakhapatnam, Andhra Pradesh, 530040, India.
  • Child Sight Institute, Hyderabad, Telangana, India.
  • Eye and Brain Centre, L V Prasad Eye Institute, Hyderabad, Telangana, India.
  • Brien Holden Institute of Optometry and Vision Sciences, L V Prasad Eye Institute, Hyderabad, India. bharadwaj@lvpei.org.
  • Brien Holden Eye Research Centre, Hyderabad Eye Research Foundation, L V Prasad Eye Institute, Hyderabad, India. bharadwaj@lvpei.org.

Zitieren

APA

Nisha, E., Negi, R., Halder, T., Kalivemula, M., Quadri, N., Kumar, A., Bhate, M., Sachdeva, V., & Bharadwaj, S. R. (2026). Static and dynamic characteristics of the pupil light reflex obtained in Indian eyes using a hand-held, objective pupillometer. Documenta Ophthalmologica. https://doi.org/10.1007/s10633-026-10160-4

Vancouver

Nisha E, Negi R, Halder T, Kalivemula M, Quadri N, Kumar A, et al. Static and dynamic characteristics of the pupil light reflex obtained in Indian eyes using a hand-held, objective pupillometer. Documenta Ophthalmologica. 2026. doi:10.1007/s10633-026-10160-4

BibTeX
@article{Nisha2026,
  author = {Nisha, Erum and Negi, Rahul and Halder, Tithi and Kalivemula, Manasa and Quadri, Nabeel and Kumar, Ayush and Bhate, Manjushree and Sachdeva, Virender and Bharadwaj, Shrikant R},
  title = {Static and dynamic characteristics of the pupil light reflex obtained in Indian eyes using a hand-held, objective pupillometer},
  journal = {Documenta Ophthalmologica},
  year = {2026},
  doi = {10.1007/s10633-026-10160-4}
}
Kostenlos starten →