Fachartikel in der Zeitschrift Documenta Ophthalmologica, online erschienen am 01.04.2026. Verfasst von Mark A Kahrhoff und 1 weiteren Autor:innen. Beteiligte Einrichtung: Department of Ophthalmology, Washington University, St. Louis, MO, USA. kmark@wustl.edu. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To evaluate variability observed in pupil light response with the EyeKinetix pupillometer in healthy non-dark-adapted individuals.

Methods

We performed objective pupillometry in 440 non-dark-adapted patients with 20/20 corrected visual acuity, normal visual fields, and no ophthalmic disease who presented for routine exam. EyeKinetix was performed as part of the routine exam screening protocol. Metrics reviewed were RAPDx amplitude scores, latency scores, and quantitative metrics of pupil dynamics measured by the EyeKinetix. Sequential retesting was analyzed for repeatability. Statistical analyses included normality testing, confidence intervals, t-tests, and Intraclass Correlation Coefficient (ICC) calculations for repeated measures.

Results

Amplitude and latency scores exhibited significant variability. The mean amplitude score was 0.0249 ± 0.247, with 5% of patients > = 2 standard deviations [> 0.51 log unit (LU)]. Latency scores showed low reliability (ICC = 0.165), whereas amplitude, constriction velocity, and release velocity demonstrated moderate to high reliability (ICC = 0.472–0.966).

Conclusion

Objective pupillometry without dark adaptation using the EyeKinetix device displays substantial variability and identifies a number of relative afferent pupil defects (RAPDs) in this cohort of healthy patients. These findings indicate that, in healthy populations, reliance on RAPDx scores alone could generate false positives. Establishing robust normative cutoffs and validating them in disease cohorts will be necessary before reliable use as a screener. Further optimization may be necessary for more clinical confidence in a primary care setting.

Schlagwörter

Stichwörter der Autor:innen

Automated pupillometryObjective pupillometryRelative afferent pupillary defect (RAPD), testing reliability, Marcus Gunn pup…Swinging flashlight test

Autor:innen und Institutionen

  • Department of Ophthalmology, Washington University, St. Louis, MO, USA. kmark@wustl.edu.
  • Complete Vision Care and Webster Eye Care, St. Louis, MO, USA. kmark@wustl.edu.
  • University of Missouri - St. Louis College of Optometry, St. Louis, MO, USA.

Zitieren

APA

Kahrhoff, M. A., & Bassi, C. J. (2026). Does EyeKinetix pupillometry clarify RAPD detection?. Documenta Ophthalmologica. https://doi.org/10.1007/s10633-026-10097-8

Vancouver

Kahrhoff MA, Bassi CJ. Does EyeKinetix pupillometry clarify RAPD detection?. Documenta Ophthalmologica. 2026. doi:10.1007/s10633-026-10097-8

BibTeX
@article{Kahrhoff2026,
  author = {Kahrhoff, Mark A and Bassi, Carl J},
  title = {Does EyeKinetix pupillometry clarify RAPD detection?},
  journal = {Documenta Ophthalmologica},
  year = {2026},
  doi = {10.1007/s10633-026-10097-8}
}
Kostenlos starten →