Automated analysis of echocardiographic images: do different artificial intelligence tools provide concordant measurements?

Teodora Szasz, Juan I Cotella, Jeremy A Slivnick, Jia Guo, Niklas Hitschrich, Hendrik Wiebel, Marcus Schreckenberg, Federico M Asch, Victor Mor-Avi, Roberto M Lang

The International Journal of Cardiovascular Imaging · 2026

DOI: 10.1007/s10554-025-03539-y ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift The International Journal of Cardiovascular Imaging, 2026. Verfasst von Teodora Szasz und 9 weiteren Autor:innen. Beteiligte Einrichtung: Philips HealthCare, Cambridge, MA, USA. Teodora.Szasz@bsd.uchicago.edu. Hauptthemen laut MEDLINE: Image Interpretation, Computer-Assisted, Heart Diseases, Artificial Intelligence, Echocardiography, Doppler. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 21 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The use of artificial intelligence (AI) in echocardiography has been growing continuously, particularly in chamber quantification and disease detection. Although studies have validated different AI approaches for automated quantification of cardiac size and function, it is unknown whether different AI software tools provide concordant measurements. Accordingly, we aimed to: (1) evaluate the inter-software variability of the automated, AI-based measurements, and (2) compare their variability to that of conventional measurements performed by expert readers.Echocardiographic images from 116 randomly selected subjects, including 66 patients with cardiac pathology and 50 healthy volunteers were analyzed to obtain the following: (1) linear measurements: end-diastolic left ventricular internal diameter (LVIDd), basal right ventricular diameter (RVDd), and LV posterior wall thickness (LVPWd), (2) Doppler measurements (e' lateral, e' septal, mitral inflow E/A ratio), and (3) end-diastolic and end-systolic left ventricular volumes (LVEDV, LVESV), ejection fraction (LVEF), and end-systolic left atrial volume (LAESV). These measurements were performed by: (1) two fully automated AI-based software tools (Us2.AI and Philips), and (2) two expert readers using guidelines recommended methodology. Inter-software and inter-observer variability were assessed by percent absolute difference (PAD) and intraclass correlation coefficients (ICC) between corresponding measurements.AI measurements differed from each other, with the differences reaching significance for 6/10 parameters, while conventional measurements by experts significantly differed from each other for 3/10 parameters. Overall, the variability between AI tools was similar to the conventional methodology, with smaller PAD and higher ICC values noted for 6/10 and 2/10 parameters, respectively. Although different commercial AI-based tools for automated analysis of echocardiographic images do not provide identical results, the variability in the majority of these measurements was similar to or better than that of experts using the conventional methodology. Further studies are needed to ascertain these findings for other AI algorithms and larger numbers of readers and parameters.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Image Interpretation, Computer-AssistedHeart DiseasesArtificial IntelligenceEchocardiography, DopplerHumansReproducibility of ResultsObserver VariationPredictive Value of TestsMaleFemaleMiddle AgedVentricular Function, LeftAutomationAgedSoftwareCase-Control StudiesAdultStroke Volume

Stichwörter der Autor:innen

Automated measurementsEchocardiographic imagingMachine learning

Autor:innen und Institutionen

  • Philips HealthCare, Cambridge, MA, USA. Teodora.Szasz@bsd.uchicago.edu.
  • University of Chicago Medical Center, Chicago, IL, USA.
  • Philips HealthCare, Cambridge, MA, USA.
  • MedStar Health Research Institute, Washington, DC, USA.
  • University of Chicago Medical Center, Chicago, IL, USA. vmoravi@bsd.uchicago.edu.
  • , 5758 South Maryland Ave, MC 9067, Room 5512, Chicago, IL, 9067, USA.

Zitieren

APA

Szasz, T., Cotella, J. I., Slivnick, J. A., Guo, J., Hitschrich, N., Wiebel, H., Schreckenberg, M., Asch, F. M., Mor-Avi, V., & Lang, R. M. (2026). Automated analysis of echocardiographic images: do different artificial intelligence tools provide concordant measurements?. The International Journal of Cardiovascular Imaging. https://doi.org/10.1007/s10554-025-03539-y

Vancouver

Szasz T, Cotella JI, Slivnick JA, Guo J, Hitschrich N, Wiebel H, et al. Automated analysis of echocardiographic images: do different artificial intelligence tools provide concordant measurements?. The International Journal of Cardiovascular Imaging. 2026. doi:10.1007/s10554-025-03539-y

BibTeX
@article{Szasz2026,
  author = {Szasz, Teodora and Cotella, Juan I and Slivnick, Jeremy A and Guo, Jia and Hitschrich, Niklas and Wiebel, Hendrik and Schreckenberg, Marcus and Asch, Federico M and Mor-Avi, Victor and Lang, Roberto M},
  title = {Automated analysis of echocardiographic images: do different artificial intelligence tools provide concordant measurements?},
  journal = {The International Journal of Cardiovascular Imaging},
  year = {2026},
  doi = {10.1007/s10554-025-03539-y}
}
Kostenlos starten →