Vendor and software based variation in left atrial strain measurements: implications for clinical practice

Vlad Danaila, Oliver Archer, Luke Stefani, Aaisha Ferkh, Shaun Khanna, Faraz Pathan, Paula Brown, Liza Thomas

The International Journal of Cardiovascular Imaging · 2025

DOI: 10.1007/s10554-025-03499-3 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift The International Journal of Cardiovascular Imaging, 2025. Verfasst von Vlad Danaila und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, Westmead Hospital, Western Sydney Local Health District, Parramatta, NSW, Australia. Hauptthemen laut MEDLINE: Atrial Function, Left, Atrial Fibrillation, Software, Image Interpretation, Computer-Assisted. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 20 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: The University of Sydney.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Left atrial strain (LAS) is a valuable echocardiographic marker of left atrial function with growing clinical utility. However, variability in LAS measurements across software vendors remains a barrier to its routine clinical use. This study aimed to compare LAS measurements obtained using a dedicated left atrial-specific measurement tool (AFI-LA (General Electric (GE)) with those derived from established LV-based strain platforms: GE EchoPAC (mid-myocardial and endocardial tracking) and TomTec-Arena (feature tracking), used for measurement of LAS. LAS was measured in 112 subjects (72 individuals in sinus rhythm and 40 paroxysmal atrial fibrillation (PAF) patients in sinus rhythm at the time of transthoracic echocardiography). Reservoir (ƐR), contractile (ƐCT), and conduit (ƐCD) phasic strain were measured using AFI-LA and compared with GE-mid, GE-endo, and TomTec strain measurements. Agreement was assessed using Bland–Altman analysis and Pearson correlation. Inter- and intra-observer reproducibility was evaluated. AFI-LA measurements showed good correlation with both platforms ( r ≥ 0.7, p < 0.001). AFI-LA consistently underestimated strain values across all phases compared to GE-mid, GE-endo, and TomTec, with the smallest bias observed against GE-mid. Proportional bias was present, particularly at higher strain values. Inter- and intra-observer reproducibility of AFI-LA measurements was high ( r > 0.85). AFI-LA provides LAS measurements that are most comparable to GE mid-myocardial strain but demonstrates systematic underestimation compared to GE endocardial and TomTec-derived values. These differences underscore the need for vendor-specific calibration before AFI-LA can be used for serial assessments or applied using existing clinical threshold values. Larger validation studies are needed to support standardization and broader clinical adoption.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Atrial Function, LeftAtrial FibrillationSoftwareImage Interpretation, Computer-AssistedHeart AtriaEchocardiographyHumansReproducibility of ResultsMaleFemalePredictive Value of TestsMiddle AgedObserver VariationAged

Stichwörter der Autor:innen

EchocardiographyIntervendor agreementLeft atrial strain

Autor:innen und Institutionen

  • Department of Cardiology, Westmead Hospital, Western Sydney Local Health District, Parramatta, NSW, Australia.
  • Westmead Clinical School, University of Sydney, Sydney, NSW, Australia.
  • Department of Cardiology, Blacktown Hospital, Western Sydney Local Health District, Parramatta, NSW, Australia.
  • Department of Cardiology, Nepean Hospital, Nepean Blue Mountains Local Health District, Parramatta, NSW, Australia.
  • Department of Cardiology, Westmead Hospital, Western Sydney Local Health District, Parramatta, NSW, Australia. liza.thomas@sydney.edu.au.
  • Westmead Clinical School, University of Sydney, Sydney, NSW, Australia. liza.thomas@sydney.edu.au.

Förderung

  • The University of Sydney

Zitieren

APA

Danaila, V., Archer, O., Stefani, L., Ferkh, A., Khanna, S., Pathan, F., Brown, P., & Thomas, L. (2025). Vendor and software based variation in left atrial strain measurements: implications for clinical practice. The International Journal of Cardiovascular Imaging. https://doi.org/10.1007/s10554-025-03499-3

Vancouver

Danaila V, Archer O, Stefani L, Ferkh A, Khanna S, Pathan F, et al. Vendor and software based variation in left atrial strain measurements: implications for clinical practice. The International Journal of Cardiovascular Imaging. 2025. doi:10.1007/s10554-025-03499-3

BibTeX
@article{Danaila2025,
  author = {Danaila, Vlad and Archer, Oliver and Stefani, Luke and Ferkh, Aaisha and Khanna, Shaun and Pathan, Faraz and Brown, Paula and Thomas, Liza},
  title = {Vendor and software based variation in left atrial strain measurements: implications for clinical practice},
  journal = {The International Journal of Cardiovascular Imaging},
  year = {2025},
  doi = {10.1007/s10554-025-03499-3}
}
Kostenlos starten →