Validierungsstudie

Reliability of spectral Doppler in handheld ultrasonographic device

Dominika Filipiak-Strzecka iD, Jarosław D Kasprzak, Piotr Lipiec

The International Journal of Cardiovascular Imaging · 2025

DOI: 10.1007/s10554-025-03372-3 ↗

Kurz gefasst

Validierungsstudie in der Zeitschrift The International Journal of Cardiovascular Imaging, 2025. Verfasst von Dominika Filipiak-Strzecka und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department and Chair of Cardiology, Bieganski Specialty Hospital, Medical University of Lodz, Kniaziewicza 1/5, Lodz, 91-347, Poland. dominika.filipi…. Hauptthemen laut MEDLINE: Mitral Valve, Ventricular Function, Left, Echocardiography, Doppler, Aortic Valve. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 29 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

To verify and validate the reliability of diastolic function parameters and valvular flow velocities acquired during the handheld ultrasound device (HUD) echocardiographic examination. Study population consisted of the consecutive patients referred for consultation due to dyspnea or impaired exercise tolerance. All patients underwent brief bedside echocardiographic screening with HUD. Within the next 24 h all patients underwent full echocardiographic examination (treated as reference). 105 patients (58 men, mean age 65 ± 14 years) were enrolled in the study. All correlations of HUD and standard echo derived measurements were high or very high. Bland-Altman plot analysis revealed the underestimation bias for mitral inflow velocities- early (E) and late (A), ), mitral annular peak early diastolic velocity (e') was not burdened with bias. The agreement between the major HFA-PEFF score parameters showed either substantial or almost perfect agreement, minor parameter- moderate agreement. The correlation of the measurements of maximum mitral velocity was high with no clinically relevant bias. Kappa coefficient showed very good agreement between HUD and reference echocardiography for detecting accelerated blood flow through the mitral valve (Kappa coefficient 0.83). The assessment of the aortic valve showed a very good correlation, with a slight underestimation bias of -0.17 m/sec (P

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Mitral ValveVentricular Function, LeftEchocardiography, DopplerAortic ValveHumansReproducibility of ResultsMaleAgedFemalePredictive Value of TestsMiddle AgedEquipment DesignBlood Flow VelocityDiastoleAged, 80 and over

Stichwörter der Autor:innen

Bedside echocardiographyHand-held ultrasonographyLeft ventricle diastolic dysfunctionSpectral Doppler

Autor:innen und Institutionen

  • Department and Chair of Cardiology, Bieganski Specialty Hospital, Medical University of Lodz, Kniaziewicza 1/5, Lodz, 91-347, Poland. dominika.filipiak@gmail.com.
  • Department and Chair of Cardiology, Bieganski Specialty Hospital, Medical University of Lodz, Kniaziewicza 1/5, Lodz, 91-347, Poland.

ORCID-Profile: Dominika Filipiak-Strzecka

Zitieren

APA

Filipiak-Strzecka, D., Kasprzak, J. D., & Lipiec, P. (2025). Reliability of spectral Doppler in handheld ultrasonographic device. The International Journal of Cardiovascular Imaging. https://doi.org/10.1007/s10554-025-03372-3

Vancouver

Filipiak-Strzecka D, Kasprzak JD, Lipiec P. Reliability of spectral Doppler in handheld ultrasonographic device. The International Journal of Cardiovascular Imaging. 2025. doi:10.1007/s10554-025-03372-3

BibTeX
@article{FilipiakStrzecka2025,
  author = {Filipiak-Strzecka, Dominika and Kasprzak, Jarosław D and Lipiec, Piotr},
  title = {Reliability of spectral Doppler in handheld ultrasonographic device},
  journal = {The International Journal of Cardiovascular Imaging},
  year = {2025},
  doi = {10.1007/s10554-025-03372-3}
}
Kostenlos starten →