The impact of different reconstruction parameters on quantitative 99mTc-DPD SPECT/CT values in the assessment of cardiac transthyretin amyloidosis

Robin Schepers, Nasir Gözlügöl, Kostantinos Zeimpekis, Carola Maria Bregenzer, Christoph Gräni, Ali Afshar-Oromieh, Axel Rominger, Federico Caobelli

The International Journal of Cardiovascular Imaging · 2024

DOI: 10.1007/s10554-024-03231-7 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift The International Journal of Cardiovascular Imaging, 2024. Verfasst von Robin Schepers und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Nuclear Medicine, Bern University Hospital, Inselspital, University of Bern, Freiburgstrasse 18, Bern, 3011, Switzerland. Robin.scheper…. Hauptthemen laut MEDLINE: Amyloid Neuropathies, Familial, Predictive Value of Tests, Radiopharmaceuticals, Single Photon Emission Computed Tomography Computed Tomography. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 15 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

AIM: To assess in a phantom and in a clinical study the influence of different reconstruction parameters on quantitative SPECT/CT values in the assessment of cardiac transthyretin amyloidosis (ATTR-CA). METHOD: A hybrid SPECT/CT camera with a proprietary software for SPECT/CT-based quantification of myocardial uptake of 99mTc-DPD was used. Images were reconstructed with 6 different protocols, differing in iterations, subset and Gaussian filter. Quantitative values were tested in phantom and clinical studies across different reconstruction protocols. Values were automatically calculated both as kBq/ml and as maximum, mean and peak standardized uptake value (SUV). RESULTS: The standard parameters provided by the manufacturer (reconstruction 1) yielded higher accuracy in quantifying, with measuring 97.1% of the expected activity in the phantom. Reconstructions with higher Gaussian filter caused a systematic underestimation of quantified values of 27.2% (p < 0.01). Results were replicated in the clinical study consisting of 155 patients with suspected ATTR-CA, wherein changing the number of iterations and subsets was not associated with a statistically significant difference in quantitative values compared to reconstruction 1, while a higher Gaussian filter caused inaccurate quantification with up to 24% of difference measured. CONCLUSION: Different reconstruction parameters can impact quantitative values on 99mTc-DPD SPECT/CT. Therefore, parameters should be maintained consistently across different acquisitions and different centres.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Amyloid Neuropathies, FamilialPredictive Value of TestsRadiopharmaceuticalsSingle Photon Emission Computed Tomography Computed TomographyOrganotechnetium CompoundsPhantoms, ImagingCardiomyopathiesImage Interpretation, Computer-AssistedHumansMaleReproducibility of ResultsFemaleAgedMiddle AgedMyocardial Perfusion ImagingDiphosphonatesPrealbumin

Stichwörter der Autor:innen

99mTc-DPD SPECT/CTATTR cardiac amyloidosisQuantificationQuantitative SPECT/CTReconstruction parameterSUV

Substanzen

RadiopharmaceuticalsOrganotechnetium Compoundstechnetium Tc 99m 1,1-diphosphonopropane-2,3-dicarboxylic acidTTR protein, humanDiphosphonatesPrealbumin

Autor:innen und Institutionen

  • Department of Nuclear Medicine, Bern University Hospital, Inselspital, University of Bern, Freiburgstrasse 18, Bern, 3011, Switzerland. Robin.schepers@extern.insel.ch.
  • Department of Nuclear Medicine, Bern University Hospital, Inselspital, University of Bern, Freiburgstrasse 18, Bern, 3011, Switzerland.
  • Department of Cardiology, Bern University Hospital, Inselspital, University of Bern, Bern, Switzerland.

Zitieren

APA

Schepers, R., Gözlügöl, N., Zeimpekis, K., Bregenzer, C. M., Gräni, C., Afshar-Oromieh, A., Rominger, A., & Caobelli, F. (2024). The impact of different reconstruction parameters on quantitative 99mTc-DPD SPECT/CT values in the assessment of cardiac transthyretin amyloidosis. The International Journal of Cardiovascular Imaging. https://doi.org/10.1007/s10554-024-03231-7

Vancouver

Schepers R, Gözlügöl N, Zeimpekis K, Bregenzer CM, Gräni C, Afshar-Oromieh A, et al. The impact of different reconstruction parameters on quantitative 99mTc-DPD SPECT/CT values in the assessment of cardiac transthyretin amyloidosis. The International Journal of Cardiovascular Imaging. 2024. doi:10.1007/s10554-024-03231-7

BibTeX
@article{Schepers2024,
  author = {Schepers, Robin and Gözlügöl, Nasir and Zeimpekis, Kostantinos and Bregenzer, Carola Maria and Gräni, Christoph and Afshar-Oromieh, Ali and Rominger, Axel and Caobelli, Federico},
  title = {The impact of different reconstruction parameters on quantitative 99mTc-DPD SPECT/CT values in the assessment of cardiac transthyretin amyloidosis},
  journal = {The International Journal of Cardiovascular Imaging},
  year = {2024},
  doi = {10.1007/s10554-024-03231-7}
}
Kostenlos starten →