EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Uncloggable ventriculoperitoneal shunt system for hydrocephalus via an integrated soft robotic device: CLEARS device

Yau C Yun, David R Santiago-Dieppa, Minghao Li, Aditya Vasan, Alexander Khalessi, James Friend

Biomedical Microdevices · 2025

DOI: 10.1007/s10544-025-00769-8 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Biomedical Microdevices, 2025. Verfasst von Yau C Yun und 5 weiteren Autor:innen. Beteiligte Einrichtung: Medically Advanced Devices Laboratory, Center for Medical Devices and Biomechanics, Department of Mechanical and Aerospace Engineering, Jacobs School…. Hauptthemen laut MEDLINE: Hydrocephalus, Ventriculoperitoneal Shunt, Robotics. Die Arbeit nennt 51 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Unaffiliated Gift Funds, Penner Endowed Chair of Engineering and Physics.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Ventriculoperitoneal (VP) shunt obstruction, often caused by protein and fat accumulation at the ventricular catheter ports, impedes cerebrospinal fluid (CSF) outflow, increases intracranial pressure (ICP), and leads to hydrocephalus. Current treatments require invasive shunt removal, reimplantation, or retrograde flush cleansing. We present a next-generation VP shunt system that actively removes blockages via external actuation. Our system, called CLogging Elimination ActuatoR Silicone (CLEARS), integrates a soft, expandable silicone tube within the catheter lumen. This soft robotic insert, capable of 900% strain, can inflate to dislodge blockages and then deflate to restore flow. To test CLEARS, we developed an ex vivo model simulating CSF flow and obstruction using a rapidly acting clogging agent. ICP upstream of the catheter was monitored to evaluate performance. When obstructed with 3 g of the clogging agent, ICP rose to 30 cmH $$_2$$ O. Upon CLEARS activation, the silicone insert expanded through catheter ports and successfully removed the clog, restoring baseline ICP ( $$\sim$$ 0 cmH $$_2$$ O) within approximately 40 s. Without the system, obstruction persisted and pressure remained elevated. Visual documentation confirmed the mechanism of action. The CLEARS system offers a promising solution to VP shunt occlusion by enabling non-invasive mechanical declogging. Our model replicates shunt obstruction and CSF dynamics, providing a testbed for device evaluation. The expandable insert maintained catheter flow and reduced ICP to normal levels after obstruction, representing a potential shift in how hydrocephalus is treated.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

HydrocephalusVentriculoperitoneal ShuntRoboticsSiliconesIntracranial PressureEquipment DesignHumansAnimals

Stichwörter der Autor:innen

CatheterCerebrospinal fluid (CSF)ObstructionShuntSoft microdevices

Substanzen

Silicones

Autor:innen und Institutionen

  • Medically Advanced Devices Laboratory, Center for Medical Devices and Biomechanics, Department of Mechanical and Aerospace Engineering, Jacobs School of Engineering and Department of Surgery, School of Medicine, Univers…
  • Department of Neurosurgery, School of Medicine, University of California San Diego, 9500 Gilman Drive, La Jolla, CA, 92093, USA.
  • Materials Science and Engineering Program, Jacobs School of Engineering, University of California San Diego, 9500 Gilman Drive, La Jolla, CA, 92093, USA.
  • Materials Science and Engineering Program, Jacobs School of Engineering, University of California San Diego, 9500 Gilman Drive, La Jolla, CA, 92093, USA. jfriend@ucsd.edu.

Förderung

  • Unaffiliated Gift Funds
  • Penner Endowed Chair of Engineering and Physics

Zitieren

APA

Yun, Y. C., Santiago-Dieppa, D. R., Li, M., Vasan, A., Khalessi, A., & Friend, J. (2025). Uncloggable ventriculoperitoneal shunt system for hydrocephalus via an integrated soft robotic device: CLEARS device. Biomedical Microdevices. https://doi.org/10.1007/s10544-025-00769-8

Vancouver

Yun YC, Santiago-Dieppa DR, Li M, Vasan A, Khalessi A, Friend J. Uncloggable ventriculoperitoneal shunt system for hydrocephalus via an integrated soft robotic device: CLEARS device. Biomedical Microdevices. 2025. doi:10.1007/s10544-025-00769-8

BibTeX
@article{Yun2025,
  author = {Yun, Yau C and Santiago-Dieppa, David R and Li, Minghao and Vasan, Aditya and Khalessi, Alexander and Friend, James},
  title = {Uncloggable ventriculoperitoneal shunt system for hydrocephalus via an integrated soft robotic device: CLEARS device},
  journal = {Biomedical Microdevices},
  year = {2025},
  doi = {10.1007/s10544-025-00769-8}
}
Kostenlos starten →