Non-contact sensor module for rapid detection of extravasation during intravenous drug administration
Biomedical Microdevices · 2024
DOI: 10.1007/s10544-024-00730-1 ↗Kurz gefasst
Fachartikel in der Zeitschrift Biomedical Microdevices, 2024. Verfasst von Muhammad Salman Al Farisi und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Information Sciences, Hiroshima City University, Hiroshima, 731-3194, Japan. alfarisi@hiroshima-cu.ac.jp. Hauptthemen laut MEDLINE: Administration, Intravenous. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 30 Literaturangaben.
Abstract (Originalsprache)
Intravenous drug administration delivers medication directly into the bloodstream, providing rapid and controlled effects, making it highly beneficial for emergencies or when immediate drug action is required. However, several risks are associated with intravenous drug administration, including infiltration and extravasation, which can lead to serious complications due to the rapid absorption of medication to the surrounding tissues. To prevent complications, here we proposed a non-contact sensor module to rapidly detect such events. The system does not interfere with the human skin, nor contaminating the flowing medication since only biocompatible materials are exposed to the liquid. The proposed sensor module was assembled as a flow channel with flow rate and pressure sensing functions. The flow rate sensing was realized using a micromachined thermal flow sensor fabricated on a thin polyimide film, while the pressure sensing was realized using a diaphragm structure and a MEMS pressure sensor. Basic characteristics of each function was evaluated and a proof of concept experiment demonstrated a rapid detection of infiltration/extravasation within a few s. Measurement of leaked fluid volume during the event was also demonstrated.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Biomedical Information Sciences, Hiroshima City University, Hiroshima, 731-3194, Japan. alfarisi@hiroshima-cu.ac.jp.
- Department of Biomedical Information Sciences, Hiroshima City University, Hiroshima, 731-3194, Japan.
- Division of Host Defense Sciences, Omics Health Sciences, Department of Integrated Health Sciences, Graduate School of Medicine, Nagoya University, Nagoya, 461-8673, Japan.
Zitieren
APA
Al Farisi, M. S., Okazaki, J., Hasegawa, Y., Matsushima, M., Kawabe, T., & Shikida, M. (2024). Non-contact sensor module for rapid detection of extravasation during intravenous drug administration. Biomedical Microdevices. https://doi.org/10.1007/s10544-024-00730-1
Vancouver
Al Farisi MS, Okazaki J, Hasegawa Y, Matsushima M, Kawabe T, Shikida M. Non-contact sensor module for rapid detection of extravasation during intravenous drug administration. Biomedical Microdevices. 2024. doi:10.1007/s10544-024-00730-1
BibTeX
@article{AlFarisi2024,
author = {Al Farisi, Muhammad Salman and Okazaki, June and Hasegawa, Yoshihiro and Matsushima, Miyoko and Kawabe, Tsutomu and Shikida, Mitsuhiro},
title = {Non-contact sensor module for rapid detection of extravasation during intravenous drug administration},
journal = {Biomedical Microdevices},
year = {2024},
doi = {10.1007/s10544-024-00730-1}
}
EUVIMED