Biomechanical Test Designs for Fixation of Traumatic Pubic Symphysis Diastasis: A Systematic Review of Key Methodological Choices in a Field Without Consensus
Annals of Biomedical Engineering · 04.11.2026
DOI: 10.1007/s10439-026-04440-y ↗Kurz gefasst
Übersichtsarbeit in der Zeitschrift Annals of Biomedical Engineering, online erschienen am 04.11.2026. Verfasst von Suzanne van Wees und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Trauma Surgery, University Medical Center Groningen, University of Groningen, Hanzeplein 1, 9713GZ, Groningen, The Netherlands. s.van.w…. Die Arbeit nennt 122 Literaturangaben. Gefördert von: Ministerie van Economische Zaken, European Union, Samenwerkingsverband Noord-Nederland.
Abstract (Originalsprache)
Purpose
Traumatic pubic symphysis diastasis (PSD) is associated with high fixation failure rates, yet biomechanical studies use highly variable test designs, limiting comparison and clinical translation. This systematic review summarizes biomechanical test designs for PSD fixation to guide standardization and improve interpretation of fixation performance.
Methods
A systematic search of PubMed and Embase up to December 2025 identified experimental and finite element modeling (FEM) studies on PSD fixation. Methodological characteristics, including injury classification, fixation, ligament and muscle representation, loading conditions, mechanical or computational setup, specimen material and outcome measures, were extracted. Data were synthesized descriptively, and a custom quality appraisal framework assessed reporting completeness and methodological rigor.
Results
Sixty-four studies were included (57 experimental, 6 FEM, one both), with substantial methodological variation. Most modeled APC II/III injuries, using plate, external, or semi-rigid fixation methods. Around half included posterior fixation. In experimental studies, synthetic pelves were common (41%), though ligaments were represented only in cadaveric and FEM studies. Loading typically simulated single- or double-leg stance using quasi-static sacral loading (~400 N). Displacement at the pubic symphysis was the most common outcome, measured in 2D or 3D, while FEM studies often assessed implant stress.
Conclusion
We propose a baseline experimental model and practical guidance for biomechanical evaluation of PSD fixation. Key challenges remain, including ligament representation and physiological loading, for which FEM may provide valuable opportunities. Although the available evidence is limited by methodological heterogeneity and incomplete reporting, this review provides a framework to improve interpretation, comparability, and future standardization of biomechanical studies. Clinical Trial registration : PROSPERO ID: CRD420251135311
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Trauma Surgery, University Medical Center Groningen, University of Groningen, Hanzeplein 1, 9713GZ, Groningen, The Netherlands. s.van.wees@umcg.nl.
- 3D Lab, University Medical Center Groningen, University of Groningen, Hanzeplein 1, 9713GZ, Groningen, The Netherlands. s.van.wees@umcg.nl.
- Department of Trauma Surgery, University Medical Center Groningen, University of Groningen, Hanzeplein 1, 9713GZ, Groningen, The Netherlands.
- 3D Lab, University Medical Center Groningen, University of Groningen, Hanzeplein 1, 9713GZ, Groningen, The Netherlands.
- Department of Orthopedics, University Medical Center Groningen, University of Groningen, Hanzeplein 1, 9713GZ, Groningen, The Netherlands.
Förderung
- Ministerie van Economische Zaken
- European Union
- Samenwerkingsverband Noord-Nederland
Zitieren
APA
van Wees, S., IJpma, F. F. A., Buist, M. L., Assink, N., Jutte, P. C., & Meesters, A. M. L. (2026). Biomechanical Test Designs for Fixation of Traumatic Pubic Symphysis Diastasis: A Systematic Review of Key Methodological Choices in a Field Without Consensus. Annals of Biomedical Engineering. https://doi.org/10.1007/s10439-026-04440-y
Vancouver
van Wees S, IJpma FFA, Buist ML, Assink N, Jutte PC, Meesters AML. Biomechanical Test Designs for Fixation of Traumatic Pubic Symphysis Diastasis: A Systematic Review of Key Methodological Choices in a Field Without Consensus. Annals of Biomedical Engineering. 2026. doi:10.1007/s10439-026-04440-y
BibTeX
@article{vanWees2026,
author = {van Wees, Suzanne and IJpma, Frank F A and Buist, Mirka L and Assink, Nick and Jutte, Paul C and Meesters, Anne M L},
title = {Biomechanical Test Designs for Fixation of Traumatic Pubic Symphysis Diastasis: A Systematic Review of Key Methodological Choices in a Field Without Consensus},
journal = {Annals of Biomedical Engineering},
year = {2026},
doi = {10.1007/s10439-026-04440-y}
}