Image-Guided Adaptive Cryotherapy for Prostate Cancer Treatment

Eva Beek, Nobuhiko Hata, Kemal Tuncali, Pedro Moreira

Annals of Biomedical Engineering · 2025

DOI: 10.1007/s10439-025-03833-9 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Annals of Biomedical Engineering, 2025. Verfasst von Eva Beek und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Brigham & Women's Hospital, Harvard Medical School, 75 Francis St., Boston, 02115, MA, USA. e.x.beek@student.tue.nl. Hauptthemen laut MEDLINE: Prostatic Neoplasms, Cryosurgery, Cryotherapy, Surgery, Computer-Assisted. Die Arbeit nennt 33 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NIBIB NIH HHS, NCI NIH HHS, National Institute of Biomedical Imaging and Bioengineering.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Focal cryoablation is an effective treatment for localized and recurrent prostate cancer, offering reduced risks of side effects. However, treatment success depends on physician experience, as intraprocedural adjustments are required due to needle deflection. To determine whether an image-guided adaptive treatment strategy could reduce the required years of experience, we developed Image-guided Adaptive Cryotherapy (ImAC). We hypothesize that ImAC can be successfully implemented in MRI-guided cryotherapy to optimize needle placement. To test this hypothesis, we conducted a retrospective study comparing the performance of an experienced physician to that of ImAC.

Methods

ImAC was designed to calculate the best subsequent needle location while accounting for needle deflection and iceball formation. After development, its performance was evaluated by simulating 21 needle insertions.

Results

We found that ImAC achieved a higher median minimum ablation margin than the physician (5.0 vs 3.4 mm) while maintaining similar needle placement adjustments (± 7.5 mm).

Conclusion

These results suggest that ImAC has the potential to optimize needle placement and may reduce the years of experience required for effective cryoablation.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Prostatic NeoplasmsCryosurgeryCryotherapySurgery, Computer-AssistedMaleHumansRetrospective StudiesMagnetic Resonance Imaging

Stichwörter der Autor:innen

3D SlicerAdaptive needle placementFocal cryoablationMRI-guided cryoablationProstate cancer

Autor:innen und Institutionen

  • Department of Radiology, Brigham & Women's Hospital, Harvard Medical School, 75 Francis St., Boston, 02115, MA, USA. e.x.beek@student.tue.nl.
  • Department of Biomedical Engineering, Eindhoven University of Technology, Eindhoven, Netherlands. e.x.beek@student.tue.nl.
  • Department of Radiology, Brigham & Women's Hospital, Harvard Medical School, 75 Francis St., Boston, 02115, MA, USA. hata@bwh.harvard.edu.
  • Department of Radiology, Brigham & Women's Hospital, Harvard Medical School, 75 Francis St., Boston, 02115, MA, USA.
  • Department of Radiology, Brigham & Women's Hospital, Harvard Medical School, 75 Francis St., Boston, 02115, MA, USA. plopesdafrotamoreira@bwh.harvard.edu.

Förderung

  • NIBIB NIH HHS
  • NCI NIH HHS
  • National Institute of Biomedical Imaging and Bioengineering

Zitieren

APA

Beek, E., Hata, N., Tuncali, K., & Moreira, P. (2025). Image-Guided Adaptive Cryotherapy for Prostate Cancer Treatment. Annals of Biomedical Engineering. https://doi.org/10.1007/s10439-025-03833-9

Vancouver

Beek E, Hata N, Tuncali K, Moreira P. Image-Guided Adaptive Cryotherapy for Prostate Cancer Treatment. Annals of Biomedical Engineering. 2025. doi:10.1007/s10439-025-03833-9

BibTeX
@article{Beek2025,
  author = {Beek, Eva and Hata, Nobuhiko and Tuncali, Kemal and Moreira, Pedro},
  title = {Image-Guided Adaptive Cryotherapy for Prostate Cancer Treatment},
  journal = {Annals of Biomedical Engineering},
  year = {2025},
  doi = {10.1007/s10439-025-03833-9}
}
Kostenlos starten →