Themes, Rates, and Risk of Adverse Events of the Artificial Pancreas in the United States Using MAUDE

Themes, Raten und Risiko unerwünschter Ereignisse der künstlichen Pankreas-Systeme in den USA anhand von MAUDE

Andrew M Ferguson, Alex C Lin

Annals of Biomedical Engineering · 2024-08 · 52 (8) · 2282-2286

DOI: 10.1007/s10439-024-03529-6 ↗

Auf einen Blick

Auswertungen von MAUDE-Berichten zu künstlichen Pankreas-Systemen in den USA zeigten vor allem Ereignisse durch Defekte an Gehäusematerialien mit Ausfall der Therapie.

Fragestellung

Welche Meldehäufigkeit, Themen und welches Risiko unerwünschter Ereignisse zeigen künstliche Pankreas-Systeme in den USA in MAUDE?

Methode

Auswertung von MAUDE-Berichten zu Gerätenamen von 2016 bis August 2023; thematische Analyse mit Dual-Reviewer-Prüfung und Berechnung von RR sowie adjustierten Ereignisraten je Hersteller.

Ergebnisse

Die meisten gemeldeten Ereignisse betrafen Defekte an Gehäusematerialien mit Nichtabgabe der Therapie. Adjustierte Ereignisrate und RR: Tandem Diabetes Care, Inc. 50 pro 100.000 Einheiten; RR 0.0225. Insulet 300 pro 100.000 Einheiten; RR 0.1684. Medtronic 2771.43 pro 100.000 Einheiten; RR 20.7857.

Schlussfolgerung der Autor:innen

Die Ereignisraten sind laut Autor:innen noch nicht auf einem akzeptablen Niveau; weitere Sicherheitsforschung und Entwicklung sind nötig, um die Raten zu senken.

Kernaussagen

  • Häufigstes Thema: Defekte an Gehäusematerialien mit Nichtabgabe der Therapie
  • Unterschiedliche adjustierte Ereignisraten zwischen den Herstellern
  • Neuere Medtronic-Geräte zeigten Verbesserungen der wahrscheinlichen Ereignisrate

Deutsche Zusammenfassung: KI-gestützt aus dem Original-Abstract erstellt; alle Zahlen wurden automatisch mit dem Original abgeglichen. Maßgeblich ist der Original-Abstract.

Kurz gefasst

Fachartikel in der Zeitschrift Annals of Biomedical Engineering, online erschienen am 13.05.2024. Verfasst von Andrew M Ferguson und 1 weiteren Autor:innen. Beteiligte Einrichtung: University of Cincinnati College of Medicine, Cincinnati, USA. fergusam@ucmail.uc.edu. Hauptthemen laut MEDLINE: Pancreas, Artificial. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

Auf einen BlickAbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Three manufacturers sell artificial pancreas systems in the United States for management of Type 1 Diabetes. Given the life-saving task required of an artificial pancreas there needs to be a high level of trust and safety in the devices. This evaluation sought to find the adjusted safety event reporting rate and themes along with device-associated risk in events reported utilizing the MAUDE database. We searched device names in the MAUDE database over the period from 2016 until August 2023 (the date of retrieval). Thematic analysis was performed using dual-reviewer examination with a 96% concurrence. Relative risk (RR) was calculated for injury, malfunction, and overall, for each manufacturer, as well as adjusted event rate per manufacturer. Most events reported related to defects in the manufacturing of the casing materials which resulted in non-delivery of therapy. Tandem Diabetes Care, Inc. had an adjusted event rate of 50 per 100,000 units and RR of 0.0225. Insulet had an adjusted event rate of 300 per 100,000 units and RR of 0.1684. Medtronic has an adjusted event rate of 2771.43 per 100,000 units and RR of 20.7857. The newer Medtronic devices show improvements in likely event rate. While the artificial pancreas is still in its infancy, these event rates are not at an acceptable level for a device which can precipitate death from malfunctions. Further exploration into safety events and much more research and development is needed for devices to reduce the event rates. Improved manufacturing practices, especially the casing materials, are highly recommended. The artificial pancreas holds promise for millions but must be improved before it becomes a true life-saving device that it has the potential to become.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Pancreas, ArtificialHumansUnited StatesDiabetes Mellitus, Type 1Databases, FactualEquipment Failure

Stichwörter der Autor:innen

Artificial pancreasClosed-loop dual-therapy insulinDevice safety eventsType 1 diabetes

Autor:innen und Institutionen

  • University of Cincinnati College of Medicine, Cincinnati, USA. fergusam@ucmail.uc.edu.
  • University of Cincinnati College of Pharmacy, Cincinnati, USA. fergusam@ucmail.uc.edu.
  • University of Cincinnati College of Pharmacy, Cincinnati, USA.

Zitieren

APA

Ferguson, A. M., & Lin, A. C. (2024). Themes, Rates, and Risk of Adverse Events of the Artificial Pancreas in the United States Using MAUDE. Annals of Biomedical Engineering, 52(8), 2282-2286. https://doi.org/10.1007/s10439-024-03529-6

Vancouver

Ferguson AM, Lin AC. Themes, Rates, and Risk of Adverse Events of the Artificial Pancreas in the United States Using MAUDE. Annals of Biomedical Engineering. 2024;52(8):2282-2286. doi:10.1007/s10439-024-03529-6

BibTeX
@article{Ferguson2024,
  author = {Ferguson, Andrew M and Lin, Alex C},
  title = {Themes, Rates, and Risk of Adverse Events of the Artificial Pancreas in the United States Using MAUDE},
  journal = {Annals of Biomedical Engineering},
  year = {2024},
  volume = {52},
  number = {8},
  pages = {2282--2286},
  doi = {10.1007/s10439-024-03529-6}
}
Kostenlos starten →