Fachartikel in der Zeitschrift Annals of Biomedical Engineering, 2024. Verfasst von Shanyuanye Guan und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Mechanical Engineering, University of Melbourne, Parkville, Victoria, 3010, Australia. Hauptthemen laut MEDLINE: Arthroplasty, Replacement, Knee, Femur. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 34 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: University of Melbourne.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Tibiofemoral slip velocity is a key contributor to total knee arthroplasty (TKA) component wear, yet few studies have evaluated this quantity in vivo . The aim of the present study was to measure and compare tibiofemoral slip velocities in 3 TKA designs for a range of daily activities. Mobile biplane X-ray imaging was used to measure 6-degree-of-freedom tibiofemoral kinematics and the locations of articular contact in 75 patients implanted with a posterior-stabilized, cruciate-retaining, or medial-stabilized design while each patient performed level walking, step up, step down, sit-to-stand, and stand-to-sit. Using these data, tibiofemoral slip velocity was calculated for the duration of each activity for each TKA design. The pattern of tibiofemoral slip velocity was similar for all 3 TKA designs within each activity but markedly different across the 5 activities tested, with the magnitude of peak slip velocity being significantly higher in level walking (range: 158–211 mm/s) than in all other activities (range: 43–75 mm/s). The pattern of tibiofemoral slip velocity in both the medial and lateral compartments closely resembled the pattern of tibiofemoral (knee) flexion angular velocity, with a strong linear relationship observed between slip velocity and flexion angular velocity ( r = 0.81–0.97). Tibiofemoral slip velocity was invariant to TKA design but was significantly affected by activity type. Our measurements of slip velocity and articular contact locations for a wide range of daily activities may be used as inputs in joint simulator testing protocols and computational models developed to estimate TKA component wear.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Arthroplasty, Replacement, KneeFemurHumansFemaleMaleAgedMiddle AgedKnee ProsthesisTibiaKnee JointProsthesis DesignBiomechanical PhenomenaAged, 80 and over

Stichwörter der Autor:innen

Archard’s LawCross-shear motionRollingSlidingSlip distanceTKA wear

Autor:innen und Institutionen

  • Department of Mechanical Engineering, University of Melbourne, Parkville, Victoria, 3010, Australia.
  • University of Lyon, University Gustave Eiffel, University Claude Bernard Lyon 1, LBMC UMR T_9406, F-69622, Lyon, France.
  • Department of Mechanical Engineering, University of Melbourne, Parkville, Victoria, 3010, Australia. pandym@unimelb.edu.au.

Förderung

  • University of Melbourne

Zitieren

APA

Guan, S., Dumas, R., & Pandy, M. G. (2024). Tibiofemoral Slip Velocity in Total Knee Arthroplasty is Design-Invariant but Activity-Dependent. Annals of Biomedical Engineering. https://doi.org/10.1007/s10439-024-03490-4

Vancouver

Guan S, Dumas R, Pandy MG. Tibiofemoral Slip Velocity in Total Knee Arthroplasty is Design-Invariant but Activity-Dependent. Annals of Biomedical Engineering. 2024. doi:10.1007/s10439-024-03490-4