EUVIMEDEUVIMED entdecken
Crossref

Personalized Numerical Cardiovascular Model with Weight Growth for Evaluating Pediatric Left Ventricular Assist Devices: Derivation from an Experimental Mock Circulatory Loop

Phong Tran, Victor Tedesco, Simon Kiang, Shweta Karnik, David Nguyen, O H Frazier, Katharine H Fraser, Yaxin Wang

Annals of Biomedical Engineering · 2024

DOI: 10.1007/s10439-023-03376-x ↗

Kurz gefasst

Fachartikel in der Zeitschrift Annals of Biomedical Engineering, 2024. Verfasst von Phong Tran und 7 weiteren Autor:innen. Beteiligte Einrichtung: Innovative Device and Engineering Applications Laboratory, Texas Heart Institute, Houston, TX, USA. Hauptthemen laut MEDLINE: Heart Transplantation, Heart-Assist Devices, Heart Failure. Die Arbeit nennt 30 Literaturangaben. Gefördert von: NHLBI NIH HHS, National Heart, Lung, and Blood Institute.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Pediatric patients with heart failure have limited treatment options because of a shortage of donor hearts and compatible left ventricular assist devices (LVADs). To address this issue, our group is developing an implantable pediatric LVAD for patients weighing 5-20 kg, capable of accommodating different physiological hemodynamic conditions as patients grow. To evaluate LVAD prototypes across a wide range of conditions, we developed a numerical cardiovascular model, using data from a mock circulatory loop (MCL) and patient-specific elastance functions. The numerical MCL was validated against experimental MCL results, showing good agreement, with differences ranging from 0 to 11%. The numerical model was also tested under left heart failure conditions and showed a worst-case difference of 16%. In an MCL study with a pediatric LVAD, a pediatric dataset was obtained from the experimental MCL and used to tune the numerical MCL. Then, the numerical model simulated LVAD flow by using an HQ curve obtained from the LVAD's impeller. When the numerical MCL was validated against the experimental MCL, hemodynamic differences ranged between 0 and 9%. These findings suggest that the numerical model can replicate various physiological conditions and impeller designs, indicating its potential as a tool for developing and optimizing pediatric LVADs.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Heart TransplantationHeart-Assist DevicesHeart FailureHumansChildModels, CardiovascularTissue DonorsHemodynamics

Stichwörter der Autor:innen

Computational modelingHeart failureHemodynamicsLeft ventricular assist device

Autor:innen und Institutionen

  • Innovative Device and Engineering Applications Laboratory, Texas Heart Institute, Houston, TX, USA.
  • Bioengineering Department, Georgia Institute of Technology, Atlanta, GA, USA.
  • Department of Mechanical Engineering, University of Bath, Bath, UK.
  • Innovative Device and Engineering Applications Laboratory, Texas Heart Institute, Houston, TX, USA. ywang@texasheart.org.

Förderung

  • NHLBI NIH HHS
  • National Heart, Lung, and Blood Institute

Zitieren

APA

Tran, P., Tedesco, V., Kiang, S., Karnik, S., Nguyen, D., Frazier, O. H., Fraser, K. H., & Wang, Y. (2024). Personalized Numerical Cardiovascular Model with Weight Growth for Evaluating Pediatric Left Ventricular Assist Devices: Derivation from an Experimental Mock Circulatory Loop. Annals of Biomedical Engineering. https://doi.org/10.1007/s10439-023-03376-x

Vancouver

Tran P, Tedesco V, Kiang S, Karnik S, Nguyen D, Frazier OH, et al. Personalized Numerical Cardiovascular Model with Weight Growth for Evaluating Pediatric Left Ventricular Assist Devices: Derivation from an Experimental Mock Circulatory Loop. Annals of Biomedical Engineering. 2024. doi:10.1007/s10439-023-03376-x

BibTeX
@article{Tran2024,
  author = {Tran, Phong and Tedesco, Victor and Kiang, Simon and Karnik, Shweta and Nguyen, David and Frazier, O H and Fraser, Katharine H and Wang, Yaxin},
  title = {Personalized Numerical Cardiovascular Model with Weight Growth for Evaluating Pediatric Left Ventricular Assist Devices: Derivation from an Experimental Mock Circulatory Loop},
  journal = {Annals of Biomedical Engineering},
  year = {2024},
  doi = {10.1007/s10439-023-03376-x}
}
Kostenlos starten →