Comparison of AIZE-EX and SITA-standard visual field measurements in patients with glaucoma

Kiyoshi Kano, Hiroki Nomoto, Chota Matsumoto, Shunji Kusaka, Yasuaki Kuwayama

Japanese Journal of Ophthalmology · 27.07.2026

DOI: 10.1007/s10384-026-01395-4 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Japanese Journal of Ophthalmology, online erschienen am 27.07.2026. Verfasst von Kiyoshi Kano und 4 weiteren Autor:innen. Beteiligte Einrichtung: Fukushima Eye Clinic, Osaka, Japan. Die Arbeit nennt 12 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To compare visual field test results between the AIZE-EX strategy of the imo perimeter and Humphrey field analyzer (HFA) Swedish interactive threshold algorithm (SITA)-standard in glaucoma patients. Study design Prospective cross-sectional study Methods This study included 52 eyes of 26 patients with glaucoma. Binocular testing with imo 24plus (1-2) AIZE-EX (78 test points) and HFA 30-2 SITA-Standard (76 test points) results obtained within 3 months were compared. Mean deviation (MD) and patter standard deviation (PSD) were recalculated from the 24-2 test points (54 points) and compared between devices. Agreement was assessed using cluster-robust t-test, Bland–Altman analysis, and equivalence testing using two one-sided tests (TOST); margin ±1.5 dB).

Results

MD was −7.89 ± 6.86 dB for AIZE-EX and −8.24 ± 7.13 dB for SITA-Standard (bias: +0.35 dB; 95% LoA: −4.04 to +4.73 dB; p = 0.342). PSD was 8.52 ± 4.15 and 8.20 ± 4.72 dB, respectively (bias: +0.32 dB; 95% LoA: −2.94 to +3.58 dB; p = 0.190). TOST confirmed equivalence for both MD (90% CI: −0.26 to +0.95 dB; p = 0.002) and PSD (90% CI: −0.09 to +0.72 dB; p < 0.001). Proportional bias was absent for MD (p = 0.405) but present for PSD (p = 0.012). Test time for both eyes was significantly shorter with AIZE-EX (7 min 39 s vs. 15 min 24 s; p < 0.001).

Conclusion

MD and PSD values were comparable between the AIZE-EX and SITA-Standard derived from 24-2 test points, demonstrating clinical equivalence. Test time was substantially shorter with the AIZE-EX despite a greater number of points, supporting its use in routine glaucoma follow-up.

Schlagwörter

Stichwörter der Autor:innen

AIZE-EXGlaucomaSITA-StandardVisual fieldimo

Autor:innen und Institutionen

  • Fukushima Eye Clinic, Osaka, Japan.
  • Department of Ophthalmology, Faculty of Medicine, Kindai University, 1-14-1, Minami-ku Miharadai, Sakai City, Osaka, 590-0197, Japan. nomoto@med.kindai.ac.jp.
  • Department of Ophthalmology, Faculty of Medicine, Kindai University, 1-14-1, Minami-ku Miharadai, Sakai City, Osaka, 590-0197, Japan.

Zitieren

APA

Kano, K., Nomoto, H., Matsumoto, C., Kusaka, S., & Kuwayama, Y. (2026). Comparison of AIZE-EX and SITA-standard visual field measurements in patients with glaucoma. Japanese Journal of Ophthalmology. https://doi.org/10.1007/s10384-026-01395-4

Vancouver

Kano K, Nomoto H, Matsumoto C, Kusaka S, Kuwayama Y. Comparison of AIZE-EX and SITA-standard visual field measurements in patients with glaucoma. Japanese Journal of Ophthalmology. 2026. doi:10.1007/s10384-026-01395-4

BibTeX
@article{Kano2026,
  author = {Kano, Kiyoshi and Nomoto, Hiroki and Matsumoto, Chota and Kusaka, Shunji and Kuwayama, Yasuaki},
  title = {Comparison of AIZE-EX and SITA-standard visual field measurements in patients with glaucoma},
  journal = {Japanese Journal of Ophthalmology},
  year = {2026},
  doi = {10.1007/s10384-026-01395-4}
}
Kostenlos starten →