Fachartikel in der Zeitschrift Journal of Imaging Informatics in Medicine, 2024. Verfasst von Yongchun You und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, West China Hospital of Sichuan University, Chengdu, China. Hauptthemen laut MEDLINE: Deep Learning, Tomography, X-Ray Computed, Radiation Dosage, Algorithms. Die Arbeit wurde bisher 15-mal zitiert (Crossref) und nennt 28 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: National Key Research and Development Program of China, West China Hospital "1·3·5" Discipline of Excellence Project, West China Hospital “1·3·5” Discipline of Excellence Project.
Abstract (Originalsprache)
This study aims to investigate the maximum achievable dose reduction for applying a new deep learning-based reconstruction algorithm, namely the artificial intelligence iterative reconstruction (AIIR), in computed tomography (CT) for hepatic lesion detection. A total of 40 patients with 98 clinically confirmed hepatic lesions were retrospectively included. The mean volume CT dose index was 13.66 ± 1.73 mGy in routine-dose portal venous CT examinations, where the images were originally obtained with hybrid iterative reconstruction (HIR). Low-dose simulations were performed in projection domain for 40%-, 20%-, and 10%-dose levels, followed by reconstruction using both HIR and AIIR. Two radiologists were asked to detect hepatic lesion on each set of low-dose image in separate sessions. Qualitative metrics including lesion conspicuity, diagnostic confidence, and overall image quality were evaluated using a 5-point scale. The contrast-to-noise ratio (CNR) for lesion was also calculated for quantitative assessment. The lesion CNR on AIIR at reduced doses were significantly higher than that on routine-dose HIR (all p < 0.05). Lower qualitative image quality was observed as the radiation dose reduced, while there were no significant differences between 40%-dose AIIR and routine-dose HIR images. The lesion detection rate was 100%, 98% (96/98), and 73.5% (72/98) on 40%-, 20%-, and 10%-dose AIIR, respectively, whereas it was 98% (96/98), 73.5% (72/98), and 40% (39/98) on the corresponding low-dose HIR, respectively. AIIR outperformed HIR in simulated low-dose CT examinations of the liver. The use of AIIR allows up to 60% dose reduction for lesion detection while maintaining comparable image quality to routine-dose HIR.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiology, West China Hospital of Sichuan University, Chengdu, China.
- United Imaging Healthcare, Shanghai, China.
- Department of Radiology, West China Hospital of Sichuan University, Chengdu, China. 452766554@qq.com.
- Department of Radiology, West China Hospital of Sichuan University, Chengdu, China. HX_lizhenlin@126.com.
Förderung
- National Key Research and Development Program of China
- West China Hospital "1·3·5" Discipline of Excellence Project
- West China Hospital “1·3·5” Discipline of Excellence Project
Zitieren
APA
You, Y., Zhong, S., Zhang, G., Wen, Y., Guo, D., Li, W., & Li, Z. (2024). Exploring the Low-Dose Limit for Focal Hepatic Lesion Detection with a Deep Learning-Based CT Reconstruction Algorithm: A Simulation Study on Patient Images. Journal of Imaging Informatics in Medicine. https://doi.org/10.1007/s10278-024-01080-3
Vancouver
You Y, Zhong S, Zhang G, Wen Y, Guo D, Li W, et al. Exploring the Low-Dose Limit for Focal Hepatic Lesion Detection with a Deep Learning-Based CT Reconstruction Algorithm: A Simulation Study on Patient Images. Journal of Imaging Informatics in Medicine. 2024. doi:10.1007/s10278-024-01080-3
BibTeX
@article{You2024,
author = {You, Yongchun and Zhong, Sihua and Zhang, Guozhi and Wen, Yuting and Guo, Dian and Li, Wanjiang and Li, Zhenlin},
title = {Exploring the Low-Dose Limit for Focal Hepatic Lesion Detection with a Deep Learning-Based CT Reconstruction Algorithm: A Simulation Study on Patient Images},
journal = {Journal of Imaging Informatics in Medicine},
year = {2024},
doi = {10.1007/s10278-024-01080-3}
}
EUVIMED