EUVIMEDEUVIMED entdecken
ValidierungsstudieOpen Access

Validation of Signal Intensity Gradient from TOF-MRA for Wall Shear Stress by Phase-Contrast MR

Chan-Hyuk Lee iD, Sang Hyuk Lee, Hyo-Sung Kwak, Yeong-Gon Kwak, Robert S Rosenson, Young I Cho, Seul-Ki Jeong

Journal of Imaging Informatics in Medicine · 2024

DOI: 10.1007/s10278-024-00991-5 ↗

Kurz gefasst

Validierungsstudie in der Zeitschrift Journal of Imaging Informatics in Medicine, 2024. Verfasst von Chan-Hyuk Lee und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Neurology, Asan Medical Center, Seoul, Republic of Korea. Hauptthemen laut MEDLINE: Magnetic Resonance Angiography, Stress, Mechanical. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 38 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: Technology Development Program of MSS, the Korean Society of Neurosonology.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

To validate the correlation between the signal intensity gradient (SIG) from time-of-flight magnetic resonance angiography (TOF-MRA) and wall shear stress (WSS) determined by phase contrast magnetic resonance (PC-MR), we conducted both experimental and human studies. In the experimental study, we measured WSS in four tubes of different sizes with variable flow rates using PC-MR and TOF-MRA. The flow rates of water in the experimental study ranged from 0.06 to 12.75 mL/s, resulting in PC-WSS values between 0.1 and 1.6 dyne/cm2. The correlation between PC-WSS and SIG was statistically significant, showing a coefficient of 0.86 (P < 0.001, R2 = 0.75). The line fit provided the conversion equation as Y = 1.6287X - 1.1563 (Y = PC-WSS, X = SIG). For the human study, 28 subjects underwent TOF-MRA and PC-MR examinations of carotid and vertebral arteries. Arterial PC-WSS and SIG were determined in the same segment for each subject. The arterial PC-WSS ranged from 1.9 to 21.0 dyne/cm2. Both carotid and vertebral arteries showed significant correlations between PC-WSS and SIG, with coefficients of 0.85, 0.86, 0.91, and 0.81 in the right and left carotid and vertebral arteries, respectively. Our results show that SIG from TOF-MRA and SIG-WSS derived from the conversion equation provide concurrent in vivo hemodynamic information on arterial shear stress. This study was registered on ClinicalTrials.gov with the identifier NCT04585971 on October 14, 2020.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Magnetic Resonance AngiographyStress, MechanicalHumansMaleFemaleAdultCarotid ArteriesMiddle AgedVertebral Artery

Stichwörter der Autor:innen

Carotid arteryIschemic strokeSignal intensity gradientTime-of-flight magnetic resonance angiographyWall shear stress

Autor:innen und Institutionen

  • Department of Neurology, Asan Medical Center, Seoul, Republic of Korea.
  • Department of Neurology, Ulsan University Hospital, University of Ulsan College of Medicine, Ulsan, Republic of Korea.
  • Equipment Qualification Center for Nuclear Power Plants, Korea Institute of Machinery and Materials, Daejeon, Republic of Korea.
  • Department of Radiology and Research Institute of Clinical Medicine of Jeonbuk National University, Biomedical Research Institute of Jeonbuk National University Hospital, Jeonju, Republic of Korea.
  • Department of Radiotechnology, Wonkwang Health Science University, Iksan, Republic of Korea.
  • Mount Sinai Heart, Icahn School of Medicine at Mount Sinai, New York, NY, USA.

ORCID-Profile: Chan-Hyuk Lee

Förderung

  • Technology Development Program of MSS
  • the Korean Society of Neurosonology

Zitieren

APA

Lee, C. H., Lee, S. H., Kwak, H. S., Kwak, Y. G., Rosenson, R. S., Cho, Y. I., & Jeong, S. K. (2024). Validation of Signal Intensity Gradient from TOF-MRA for Wall Shear Stress by Phase-Contrast MR. Journal of Imaging Informatics in Medicine. https://doi.org/10.1007/s10278-024-00991-5

Vancouver

Lee CH, Lee SH, Kwak HS, Kwak YG, Rosenson RS, Cho YI, et al. Validation of Signal Intensity Gradient from TOF-MRA for Wall Shear Stress by Phase-Contrast MR. Journal of Imaging Informatics in Medicine. 2024. doi:10.1007/s10278-024-00991-5

BibTeX
@article{Lee2024,
  author = {Lee, Chan-Hyuk and Lee, Sang Hyuk and Kwak, Hyo-Sung and Kwak, Yeong-Gon and Rosenson, Robert S and Cho, Young I and Jeong, Seul-Ki},
  title = {Validation of Signal Intensity Gradient from TOF-MRA for Wall Shear Stress by Phase-Contrast MR},
  journal = {Journal of Imaging Informatics in Medicine},
  year = {2024},
  doi = {10.1007/s10278-024-00991-5}
}
Kostenlos starten →