Mechanical Effects of Medical Device Attachment to Human Tympanic Membrane
Journal of the Association for Research in Otolaryngology · 2024
DOI: 10.1007/s10162-024-00942-5 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of the Association for Research in Otolaryngology, 2024. Verfasst von Arash Ebrahimian und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Systems Design Engineering, University of Waterloo, Waterloo, Ontario, Canada. Hauptthemen laut MEDLINE: Tympanic Membrane, Vibration, Finite Element Analysis. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 40 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: Natural Sciences and Engineering Research Council of Canada, Canada Foundation for Innovation.
Abstract (Originalsprache)
PURPOSE: Several treatment methods for hearing disorders rely on attaching medical devices to the tympanic membrane. This study aims to systematically analyze the effects of the material and geometrical properties and location of the medical devices attached to the tympanic membrane on middle-ear vibrations. METHODS: A finite-element model of the human middle ear was employed to simulate the effects of attachment of medical devices. Various types of material and geometrical properties, locations, and modeling scenarios were investigated for the medical device. RESULTS: The attachment of the device magnifies the effects of anti-resonances of the middle ear. Additionally, the variations of the material properties of the device significantly alter the middle-ear resonance frequency while changes in the umbo and stapes footplate motions are negligible at frequencies above 5 kHz. Furthermore, modeling the device as a point mass cannot accurately represent the implanted middle-ear behavior. The variations of the diameter and height of the medical device have negligible effects on the middle-ear vibrations at frequencies below 200 Hz but can have considerable impacts at higher frequencies. The effects of changing the device height were negligible at frequencies above 2 kHz. We also discuss the effects of medical device attachment on the vibration patterns of the tympanic membrane as well as the impacts of the variations of the location of the device on the stapes footplate responses. CONCLUSION: The findings of our study aid the development and optimization of new therapeutic devices, attached to the tympanic membrane, to have the least adverse effects on middle-ear vibrations.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Systems Design Engineering, University of Waterloo, Waterloo, Ontario, Canada.
- Centre for Bioengineering and Biotechnology, University of Waterloo, Waterloo, Ontario, Canada.
- Department of Systems Design Engineering, University of Waterloo, Waterloo, Ontario, Canada. nmaftoon@uwaterloo.ca.
- Centre for Bioengineering and Biotechnology, University of Waterloo, Waterloo, Ontario, Canada. nmaftoon@uwaterloo.ca.
Förderung
- Natural Sciences and Engineering Research Council of Canada
- Canada Foundation for Innovation
Zitieren
APA
Ebrahimian, A., Mohammadi, H., & Maftoon, N. (2024). Mechanical Effects of Medical Device Attachment to Human Tympanic Membrane. Journal of the Association for Research in Otolaryngology. https://doi.org/10.1007/s10162-024-00942-5
Vancouver
Ebrahimian A, Mohammadi H, Maftoon N. Mechanical Effects of Medical Device Attachment to Human Tympanic Membrane. Journal of the Association for Research in Otolaryngology. 2024. doi:10.1007/s10162-024-00942-5
BibTeX
@article{Ebrahimian2024,
author = {Ebrahimian, Arash and Mohammadi, Hossein and Maftoon, Nima},
title = {Mechanical Effects of Medical Device Attachment to Human Tympanic Membrane},
journal = {Journal of the Association for Research in Otolaryngology},
year = {2024},
doi = {10.1007/s10162-024-00942-5}
}
EUVIMED