Fachartikel in der Zeitschrift Neurosurgical Review, 2024. Verfasst von Konstantin Brawanski und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Neurosurgery, Medical University Innsbruck, Anichstrasse 35, Innsbruck, 6020, Austria. Hauptthemen laut MEDLINE: Ventriculoperitoneal Shunt, Feasibility Studies, Hydrocephalus, Ultrasonography. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 42 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: University of Innsbruck and Medical University of Innsbruck.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The objective of this pilot study was to assess the reliability of superb microvascular ultrasound (SMI) for the measurement of the cerebrospinal fluid (CSF) flow within VPS systems as an indirect sign for shunt dysfunction. Asymptomatic hydrocephalus patients, with a VPS system implanted between 2017 and 2021, were prospectively enrolled in the study. Using SMI, the CSF flow within the proximal and distal catheters were analysed. Before and after pumping the shunt reservoir, intraabdominal free fluid, optical nerve sheath diameter (ONSD), and papilla diameter (PD) were evaluated and correlated with the amount of valve activation. Nineteen patients were included. A flow was detectable in 100% ( N = 19) patients in the proximal and in 89.5% ( N = 17) in the distal catheter. The distal catheter tip was detectable in 27.7% ( N = 5) patients. Free intraabdominal fluid was initially detected in 21.4% ( N = 4) patients and in 57.9% ( N = 11) at the end of the examination ( P = 0.049). ONSD was significantly lower after pump activation (4.4 ± 0.9 mm versus 4.1 ± 0.8 mm, P = 0.049). Both peak velocity and flow volume per second were higher in proximal compared to distal catheters (32.2 ± 45.2 versus 5.6 ± 3.7 cm/sec, P = 0.015; 16.6 ± 9.5 ml/sec versus 5.1 ± 4.0 ml/sec, P = 0.001, respectively). No correlation was found between the number of pump activations and the changes in ONSD ( P = 0.975) or PD ( P = 0.820). SMI appears to be a very promising non-invasive diagnostic tool to assess CSF flow within the VPS systems and therefore affirm their function. Furthermore, appearance of free intraperitoneal fluid followed by repeated compression of a shunt reservoir indicates an intact functioning shunt system.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Ventriculoperitoneal ShuntFeasibility StudiesHydrocephalusUltrasonographyHumansMaleFemaleMiddle AgedAgedPilot ProjectsAdultProspective StudiesMicrovesselsAged, 80 and overReproducibility of Results

Stichwörter der Autor:innen

FlowOptical nerve sheath diameterPapilla diametersSonographySuperb microvascular ultrasoundVentriculoperitoneal shunt

Autor:innen und Institutionen

  • Department of Neurosurgery, Medical University Innsbruck, Anichstrasse 35, Innsbruck, 6020, Austria.
  • Department of Neurosurgery, Medical University Innsbruck, Anichstrasse 35, Innsbruck, 6020, Austria. ondra.petr@yahoo.com.
  • Department of Radiology, Medical University Innsbruck, Anichstrasse 35, Innsbruck, 6020, Austria.

ORCID-Profile: Ondra Petr

Förderung

  • University of Innsbruck and Medical University of Innsbruck

Zitieren

APA

Brawanski, K., Petr, O., Hernandez, C. P., Kögl, N., Thomé, C., Gizewski, E. R., Gruber, H., Verius, M., Gruber, L., & Putzer, D. (2024). Superb microvascular ultrasound is a promising non-invasive diagnostic tool to assess a ventriculoperitoneal shunt system function: a feasibility study. Neurosurgical Review. https://doi.org/10.1007/s10143-024-02665-w

Vancouver

Brawanski K, Petr O, Hernandez CP, Kögl N, Thomé C, Gizewski ER, et al. Superb microvascular ultrasound is a promising non-invasive diagnostic tool to assess a ventriculoperitoneal shunt system function: a feasibility study. Neurosurgical Review. 2024. doi:10.1007/s10143-024-02665-w

BibTeX
@article{Brawanski2024,
  author = {Brawanski, Konstantin and Petr, Ondra and Hernandez, Christian Preuss and Kögl, Nikolaus and Thomé, Claudius and Gizewski, Elke R and Gruber, Hannes and Verius, Michael and Gruber, Leonhard and Putzer, Daniel},
  title = {Superb microvascular ultrasound is a promising non-invasive diagnostic tool to assess a ventriculoperitoneal shunt system function: a feasibility study},
  journal = {Neurosurgical Review},
  year = {2024},
  doi = {10.1007/s10143-024-02665-w}
}
Kostenlos starten →