Deep brain stimulation in Parkinson’s disease: a comparison of accuracy and clinical outcomes of frame-based, frameless and frameless fiducial-less techniques
Neurological Sciences · 2025-06 · 46 (6) · 2675-2685
DOI: 10.1007/s10072-025-08102-0 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Neurological Sciences, online erschienen am 13.03.2025. Verfasst von Riccardo Antonio Ricciuti und 10 weiteren Autor:innen. Beteiligte Einrichtung: Department of Neurosurgery, Azienda San Camillo Forlanini, Rome, Italy. Hauptthemen laut MEDLINE: Parkinson Disease, Deep Brain Stimulation, Subthalamic Nucleus, Fiducial Markers. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 20 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Università Politecnica delle Marche.
Abstract (Originalsprache)
Background
The effectiveness of deep brain stimulation (DBS) for Parkinson’s disease (PD) depends on implantation accuracy. DBS initially employed a stereotactic frame (frame-based, FB), but technological advancements led to the development of less invasive methods based on fiducial markers (F + F) or intraoperative imaging (F-F). This study compares the accuracy and efficacy of three DBS-STN implantation techniques.
Methods
This retrospective study involved 18 patients with PD who underwent bilateral STN DBS between 2018 and 2023. Patients were divided into three groups: FB ( n = 6), F + F ( n = 7), and F-F ( n = 5). Postoperative CT and preoperative MRI fusion were used to evaluate electrode accuracy via deviations from planned targets in x, y, z axes, and calculate the radial error (RE) and vector error (VE). We analyzed Unified Parkinson’s Disease Rating Scale (UPDRS) III scores in four “on/off medication-stimulation” combinations, LEDD, and disease stage before DBS, and 3 and 12 months post-DBS.
Results
No statistically significant differences were observed between the three methods in|Δx| (FB = 1.30 ± 0.91; F + F = 1.05 ± 0.93; F-F = 1.33 ± 1.09 mm),|Δy| (FB = 0.95 ± 0.98; F + F = 1.11 ± 1.17; F-F = 1.28 ± 1.14 mm), RE (FB = 1.82 ± 0.29; F + F = 1,71 ± 0,36; F-F = 1,91 ± 1,49 mm) and VE (FB = 3,14 ± 0,35 mm; F + F = 4,92 ± 0,54 mm; F-F = 4,42 ± 1,22 mm). All groups demonstrated significant UPDRS III (> 50%) and LEDD reduction (> 40%) at 12 months, with no intergroup differences.
Conclusions
The study concludes that all three techniques provide equivalent accuracy and clinical efficacy. Centers should select DBS-STN methods based on available resources and expertise.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Neurosurgery, Azienda San Camillo Forlanini, Rome, Italy.
- Department of Neurosurgery, Belcolle Hospital, Viterbo, Italy.
- Department of Neurosurgery, Università Politecnica delle Marche, Ancona, Italy. matteomaria.ottaviani@gmail.com.
- Department of Neurosurgery, Azienda Ospedaliera di Perugia, Perugia, Italy. matteomaria.ottaviani@gmail.com.
- Department of Neurosurgery, Azienda Ospedaliera di Perugia, Perugia, Italy.
- Research Unit of Neurology, Neurophysiology, Neurobiology and Psychiatry, Department of Medicine, Università Campus Bio-Medico, Rome, Italy.
Förderung
- Università Politecnica delle Marche
Zitieren
APA
Ricciuti, R. A., Ottaviani, M. M., Mancini, F., Marano, M., Marruzzo, D., Barbieri, F., Paracino, R., De Domenico, P., Pagano, S., Di Lazzaro, V., & Dobran, M. (2025). Deep brain stimulation in Parkinson’s disease: a comparison of accuracy and clinical outcomes of frame-based, frameless and frameless fiducial-less techniques. Neurological Sciences, 46(6), 2675-2685. https://doi.org/10.1007/s10072-025-08102-0
Vancouver
Ricciuti RA, Ottaviani MM, Mancini F, Marano M, Marruzzo D, Barbieri F, et al. Deep brain stimulation in Parkinson’s disease: a comparison of accuracy and clinical outcomes of frame-based, frameless and frameless fiducial-less techniques. Neurological Sciences. 2025;46(6):2675-2685. doi:10.1007/s10072-025-08102-0
BibTeX
@article{Ricciuti2025,
author = {Ricciuti, Riccardo Antonio and Ottaviani, Matteo Maria and Mancini, Fabrizio and Marano, Massimo and Marruzzo, Daniele and Barbieri, Francesca and Paracino, Riccardo and De Domenico, Pierfrancesco and Pagano, Serena and Di Lazzaro, Vincenzo and Dobran, Mauro},
title = {Deep brain stimulation in Parkinson’s disease: a comparison of accuracy and clinical outcomes of frame-based, frameless and frameless fiducial-less techniques},
journal = {Neurological Sciences},
year = {2025},
volume = {46},
number = {6},
pages = {2675--2685},
doi = {10.1007/s10072-025-08102-0}
}
EUVIMED