Preoperative predictors of low flow limit controller configuration following HeartMate 3 implantation
Journal of Artificial Organs · 2026
DOI: 10.1007/s10047-026-01590-7 ↗Kurz gefasst
Beobachtungsstudie in der Zeitschrift Journal of Artificial Organs, 2026. Verfasst von Hiroshi Nishioka und 14 weiteren Autor:innen. Beteiligte Einrichtung: Department of Clinical Engineering, National Cerebral and Cardiovascular Center, Suita, Osaka, Japan. Hauptthemen laut MEDLINE: Heart-Assist Devices, Heart Failure. Die Arbeit nennt 11 Literaturangaben. Gefördert von: JMS Co., Ltd..
Abstract (Originalsprache)
This study aimed to identify preoperative factors associated with postoperative replacement to a controller configured with a Low Flow Limit (LFL) of 2.0 L/min following HeartMate 3 (HM3) left ventricular assist device (LVAD) implantation, with the goal of establishing proactive selection criteria to improve patient safety and quality of life (QOL). This single-center, retrospective, observational study included 127 patients who underwent primary HM3 implantation between June 2019 and March 2025. Patients were divided into those who maintained the standard LFL of 2.5 L/min controller throughout follow-up (Standard group, n = 78) and those who required postoperative replacement with a controller configured with an LFL of 2.0 L/min (2.0-LFL group, n = 49). Preoperative clinical, laboratory, echocardiographic, and hemodynamic parameters were compared using univariate and multivariate logistic regression analyses, together with receiver operating characteristic (ROC) curve analysis. 2.0-LFL group had significantly lower anthropometric indices, including height, weight, body mass index, and body surface area (BSA), as well as a shorter duration of heart failure, compared with Standard group. Although multiple clinical, echocardiographic, and hemodynamic variables were evaluated, multivariate analysis identified BSA as the sole independent predictor of postoperative controller replacement (odds ratio: 0.03; 95% confidence interval: 0.00–0.60; p = 0.022). ROC curve analysis demonstrated that a preoperative BSA cutoff value of 1.43 m^2 yielded an area under the curve of 0.74, with a specificity of 89.7% and a sensitivity of 38.8%. Patient body size, particularly BSA, appears to be the primary determinant of the need for postoperative replacement to an LFL 2.0 L/min controller after HM3 implantation. Pre-emptive selection of a 2.0 L/min controller based on a BSA threshold of 1.43 m^2 may provide a practical strategy to reduce unnecessary postoperative controller replacements while maintaining patient safety and QOL in LVAD patients.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Clinical Engineering, National Cerebral and Cardiovascular Center, Suita, Osaka, Japan.
- Department of Chemical, Environmental and Biomedical Engineering, Graduate School of Engineering, Osaka Institute of Technology, Suita, Osaka, Japan.
- Department of Transplant medicine, National Cerebral and Cardiovascular Center, 6-1 Kishibe-Shimmachi, Suita, 564-8565, Osaka, Japan. takuma.sato@ncvc.go.jp.
- Department of Cardiovascular Surgery, Osaka University Hospital, Suita, Osaka, Japan.
- Department of Transplant medicine, National Cerebral and Cardiovascular Center, 6-1 Kishibe-Shimmachi, Suita, 564-8565, Osaka, Japan.
- Department of Apparel Design, Osaka Seikei College, Osaka, Osaka, Japan.
Förderung
- JMS Co., Ltd.
Zitieren
APA
Nishioka, H., Sato, T., Komiyama, M., Iwanaga, K., Tonai, K., Kitahata, N., Takahashi, Y., Miyagawa, S., Sawada, K., Yotsuida, H., Tadokoro, N., Fukushima, S., Fujita, T., Fujisato, T., & Tsukamoto, Y. (2026). Preoperative predictors of low flow limit controller configuration following HeartMate 3 implantation. Journal of Artificial Organs. https://doi.org/10.1007/s10047-026-01590-7
Vancouver
Nishioka H, Sato T, Komiyama M, Iwanaga K, Tonai K, Kitahata N, et al. Preoperative predictors of low flow limit controller configuration following HeartMate 3 implantation. Journal of Artificial Organs. 2026. doi:10.1007/s10047-026-01590-7
BibTeX
@article{Nishioka2026,
author = {Nishioka, Hiroshi and Sato, Takuma and Komiyama, Megumi and Iwanaga, Koki and Tonai, Kohei and Kitahata, Nana and Takahashi, Yuzo and Miyagawa, Sachie and Sawada, Kazuya and Yotsuida, Hideki and Tadokoro, Naoki and Fukushima, Satsuki and Fujita, Tomoyuki and Fujisato, Toshia and Tsukamoto, Yasumasa},
title = {Preoperative predictors of low flow limit controller configuration following HeartMate 3 implantation},
journal = {Journal of Artificial Organs},
year = {2026},
doi = {10.1007/s10047-026-01590-7}
}