Fracture of the cannula tip during removal of a long-term indwelling Impella 5.5: a case report
Journal of Artificial Organs · 2026
DOI: 10.1007/s10047-026-01571-w ↗Kurz gefasst
Fallbericht in der Zeitschrift Journal of Artificial Organs, 2026. Verfasst von Ryuhei Yamamoto und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiovascular Surgery, University of Tsukuba Hospital, 2-1-1 Amakubo, Tsukuba, 305-8576, Japan. Hauptthemen laut MEDLINE: Device Removal, Heart-Assist Devices, Shock, Cardiogenic, Equipment Failure. Die Arbeit nennt 5 Literaturangaben.
Abstract (Originalsprache)
We report a rare device fracture during Impella 5.5 explantation for cardiogenic shock in a 77-year-old man with prolonged device support. Following stabilization, device removal was attempted on day 50. During withdrawal, marked resistance was encountered and the cannula tip fractured, remaining within the prosthetic graft. The tip was surgically retrieved following extension of the dissection. Post-explantation, the patient developed acute right upper extremity ischemia, likely due to vasospasm, which resolved spontaneously. This case underscores the importance of recognizing marked resistance during Impella explantation, as excessive stress concentration and mechanical shearing forces may contribute to device fracture.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Cardiovascular Surgery, University of Tsukuba Hospital, 2-1-1 Amakubo, Tsukuba, 305-8576, Japan.
- Department of Cardiovascular Surgery, University of Tsukuba Hospital, 2-1-1 Amakubo, Tsukuba, 305-8576, Japan. sakamotoh@md.tsukuba.ac.jp.
Zitieren
APA
Yamamoto, R., Iguchi, Y., Mathis, B. J., & Sakamoto, H. (2026). Fracture of the cannula tip during removal of a long-term indwelling Impella 5.5: a case report. Journal of Artificial Organs. https://doi.org/10.1007/s10047-026-01571-w
Vancouver
Yamamoto R, Iguchi Y, Mathis BJ, Sakamoto H. Fracture of the cannula tip during removal of a long-term indwelling Impella 5.5: a case report. Journal of Artificial Organs. 2026. doi:10.1007/s10047-026-01571-w
BibTeX
@article{Yamamoto2026,
author = {Yamamoto, Ryuhei and Iguchi, Yusuke and Mathis, Bryan James and Sakamoto, Hiroaki},
title = {Fracture of the cannula tip during removal of a long-term indwelling Impella 5.5: a case report},
journal = {Journal of Artificial Organs},
year = {2026},
doi = {10.1007/s10047-026-01571-w}
}