Oxygenator salvage with low-dose thrombolysis. Case report
Journal of Artificial Organs · 2026-06 · 29 (2)
DOI: 10.1007/s10047-026-01559-6 ↗Kurz gefasst
Fallbericht in der Zeitschrift Journal of Artificial Organs, online erschienen am 06.05.2026. Verfasst von Rodrigo Díaz und 4 weiteren Autor:innen. Beteiligte Einrichtung: Cardiovascular Critical Care Unit, ECMO Program Director, Clinica RedSalud Santiago, Avenida Bernardo O´Higgins 4850, Santiago, Chile. rodrigoa.diaz@…. Hauptthemen laut MEDLINE: Extracorporeal Membrane Oxygenation, Tissue Plasminogen Activator, Respiratory Insufficiency, Fibrinolytic Agents. Die Arbeit nennt 6 Literaturangaben.
Abstract (Originalsprache)
Membrane oxygenator exchange can be a frequent requirement during extracorporeal membrane oxygenation (ECMO). In low-income health systems, the scarcity and cost of disposables pose significant challenges to timely exchanges. We describe a 48-year-old woman placed on veno-venous ECMO for refractory hypoxemic respiratory failure secondary to pneumonia. Within 30 min, the oxygenator developed thrombosis, characterized by a transmembrane pressure gradient of 229 mmHg. A total of 10 mg of recombinant tissue plasminogen activator (rt-PA) was administered through the premembrane port. Sixty minutes after infusion, the gradient decreased to 116 mmHg, with subsequent improvement in oxygenation. No bleeding occurred. The patient was successfully weaned from ECMO and discharged from the ICU without complications. Low-dose rtPA instillation into a failing oxygenator can postpone exchange and represents a viable salvage strategy for resource-limited ECMO programs.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Cardiovascular Critical Care Unit, ECMO Program Director, Clinica RedSalud Santiago, Avenida Bernardo O´Higgins 4850, Santiago, Chile. rodrigoa.diaz@redsalud.cl.
- Cardiovascular Center, Clinica RedSalud Santiago, Santiago, Chile.
Zitieren
APA
Díaz, R., García, D., Alvarado-Neves, C., Karlezi, C. G., & Gárate, J. (2026). Oxygenator salvage with low-dose thrombolysis. Case report. Journal of Artificial Organs, 29(2). https://doi.org/10.1007/s10047-026-01559-6
Vancouver
Díaz R, García D, Alvarado-Neves C, Karlezi CG, Gárate J. Oxygenator salvage with low-dose thrombolysis. Case report. Journal of Artificial Organs. 2026;29(2). doi:10.1007/s10047-026-01559-6
BibTeX
@article{Daz2026,
author = {Díaz, Rodrigo and García, Dafna and Alvarado-Neves, Catalina and Karlezi, Carmen Gloria and Gárate, Javier},
title = {Oxygenator salvage with low-dose thrombolysis. Case report},
journal = {Journal of Artificial Organs},
year = {2026},
volume = {29},
number = {2},
doi = {10.1007/s10047-026-01559-6}
}