EUVIMEDEUVIMED entdecken
Fallbericht

A case of outflow graft obstruction after EVAHEART2 implantation: role of device log analysis

Yudai Muratsuji, Ai Kawamura, Daisuke Yoshioka, Takuji Kawamura, Shin Yajima, Yusuke Misumi, Shigetaka Kusumoto, Yusuke Minematsu, Takeshi Matsumoto, Shunsuke Saito, Takayoshi Ueno, Masaki Takashina und 1 weitere

Journal of Artificial Organs · 2026-06 · 29 (2)

DOI: 10.1007/s10047-026-01558-7 ↗

Kurz gefasst

Fallbericht in der Zeitschrift Journal of Artificial Organs, online erschienen am 01.06.2026. Verfasst von Yudai Muratsuji und 12 weiteren Autor:innen. Beteiligte Einrichtung: Department of Clinical Engineering, The University of Osaka Hospital, Suita, Japan. Hauptthemen laut MEDLINE: Heart-Assist Devices, Candidiasis. Die Arbeit nennt 5 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Outflow graft obstruction (OGO) is a rare but potentially fatal complication of long-term left ventricular assist device use. This report presents a case of late-onset OGO caused by Candida albicans in a patient with an EVAHEART2. For EVAHEART2, which does not provide low-flow alarms, analysis of device log data revealing a gradual reduction in the maximum pump speeds and power consumption, as well as a gradual increase in the minimum pump speeds and power consumption, may be important for the early detection of OGO.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Heart-Assist DevicesCandidiasisHumansCandida albicans

Stichwörter der Autor:innen

Candida albicansDevice log analysisEVAHEART2Outflow graft obstruction

Autor:innen und Institutionen

  • Department of Clinical Engineering, The University of Osaka Hospital, Suita, Japan.
  • Division of Health Sciences, The University of Osaka Graduate School of Medicine, Suita, Japan.
  • Department of Cardiovascular Surgery, The University of Osaka Graduate School of Medicine, 2-2, Yamadaoka, Suita, Osaka, 565-0871, Japan. a-shibamoto@surg1.med.osaka-u.ac.jp.
  • Department of Cardiovascular Surgery, The University of Osaka Graduate School of Medicine, 2-2, Yamadaoka, Suita, Osaka, 565-0871, Japan.

Zitieren

APA

Muratsuji, Y., Kawamura, A., Yoshioka, D., Kawamura, T., Yajima, S., Misumi, Y., Kusumoto, S., Minematsu, Y., Matsumoto, T., Saito, S., Ueno, T., Takashina, M., & Miyagawa, S. (2026). A case of outflow graft obstruction after EVAHEART2 implantation: role of device log analysis. Journal of Artificial Organs, 29(2). https://doi.org/10.1007/s10047-026-01558-7

Vancouver

Muratsuji Y, Kawamura A, Yoshioka D, Kawamura T, Yajima S, Misumi Y, et al. A case of outflow graft obstruction after EVAHEART2 implantation: role of device log analysis. Journal of Artificial Organs. 2026;29(2). doi:10.1007/s10047-026-01558-7

BibTeX
@article{Muratsuji2026,
  author = {Muratsuji, Yudai and Kawamura, Ai and Yoshioka, Daisuke and Kawamura, Takuji and Yajima, Shin and Misumi, Yusuke and Kusumoto, Shigetaka and Minematsu, Yusuke and Matsumoto, Takeshi and Saito, Shunsuke and Ueno, Takayoshi and Takashina, Masaki and Miyagawa, Shigeru},
  title = {A case of outflow graft obstruction after EVAHEART2 implantation: role of device log analysis},
  journal = {Journal of Artificial Organs},
  year = {2026},
  volume = {29},
  number = {2},
  doi = {10.1007/s10047-026-01558-7}
}
Kostenlos starten →