Accuracy and bias of high-frequency ultrasonography in measuring white spot lesion depth: an in-vitro comparison with micro-CT

Taylor Toothman Sulkowski, Steven Makowka, Oliver D Kripfgans, William Tanberg, Stephen Warunek, Thikriat Al-Jewair

Clinical Oral Investigations · 03.06.2026 · 30 (6)

DOI: 10.1007/s00784-026-06956-y ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Clinical Oral Investigations, online erschienen am 03.06.2026. Verfasst von Taylor Toothman Sulkowski und 5 weiteren Autor:innen. Beteiligte Einrichtung: Orthodontist, private practice, Chapel Hill, NC, USA. Hauptthemen laut MEDLINE: X-Ray Microtomography, Dental Caries. Die Arbeit nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

This in vitro study aimed to evaluate the accuracy and bias of Ultrasound (US) in measuring the depth of white spot lesions (WSLs), in comparison to micro-CT (µ-CT). Materials and methods The study included 120 bovine maxillary incisors. Artificial WSLs were created on the facial surface of each incisor. Incisors were categorized into two groups, shallow WSLs and deep WSLs and subjected to 2-day and 4-day pH cycling protocols, respectively. All samples were imaged with both µ-CT and US using second harmonic imaging at 12/24 MHz. WSL depth was measured and the average calculated. Enamel thickness was assessed on the µ-CT images using the same methodology.

Results

Mean lesion depths in the shallow WSLs group were 138.3±17.8 μm (µ-CT) and 169.2 ± 37.8 μm (US), while in the deep WSLs group, depths averaged 299.9 ± 47.7 μm (µ-CT) and 309.3 ± 75.7 μm (US). The absolute mean differences between µ-CT and US were significantly different from zero ( p < 0.001). US consistently overestimated lesion depth compared to µ-CT in both groups, with a significant difference in the shallow WSLs group ( p < 0.001).

Conclusions

US demonstrates lower accuracy than µ-CT in measuring WSL depth, consistently overestimating lesion depths, particularly in the shallow WSLs group. Clinical relevance The ability of a diagnostic method to measure WSLs of varying depths is clinically important, as treatment efficacy depends on lesion depth and the extent of enamel demineralization. US in non-invasive and demonstrates potential for clinical use; however, further validation is required before it can be reliably applied to the clinical assessment of WSL depth. Clinicians must also consider practical factors such as equipment cost and size, the need for coupling materials, and requirements for operator training and calibration.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

X-Ray MicrotomographyDental CariesAnimalsCattleIn Vitro TechniquesUltrasonographyIncisorDental EnamelReproducibility of Results

Stichwörter der Autor:innen

DemineralizationDepthMicro-CTUltrasoundWhite spot lesions

Autor:innen und Institutionen

  • Orthodontist, private practice, Chapel Hill, NC, USA.
  • Materials Testing Facility, State University of New York at Buffalo, Buffalo, NY, USA.
  • Department of Radiology and Biomedical Engineering, Medical School and College of Engineering, University of Michigan, Ann Arbor, MI, USA.
  • Department of Biostatistics, School of Public Health and Health Professions, State University of New York at Buffalo, Buffalo, NY, USA.
  • Department of Orthodontics, School of Dental Medicine, State University of New York at Buffalo, 3435 Main Street, Buffalo, NY, 14214, USA.
  • Department of Orthodontics, School of Dental Medicine, State University of New York at Buffalo, 3435 Main Street, Buffalo, NY, 14214, USA. thikriat@buffalo.edu.

Zitieren

APA

Toothman Sulkowski, T., Makowka, S., Kripfgans, O. D., Tanberg, W., Warunek, S., & Al-Jewair, T. (2026). Accuracy and bias of high-frequency ultrasonography in measuring white spot lesion depth: an in-vitro comparison with micro-CT. Clinical Oral Investigations, 30(6). https://doi.org/10.1007/s00784-026-06956-y

Vancouver

Toothman Sulkowski T, Makowka S, Kripfgans OD, Tanberg W, Warunek S, Al-Jewair T. Accuracy and bias of high-frequency ultrasonography in measuring white spot lesion depth: an in-vitro comparison with micro-CT. Clinical Oral Investigations. 2026;30(6). doi:10.1007/s00784-026-06956-y

BibTeX
@article{ToothmanSulkowski2026,
  author = {Toothman Sulkowski, Taylor and Makowka, Steven and Kripfgans, Oliver D and Tanberg, William and Warunek, Stephen and Al-Jewair, Thikriat},
  title = {Accuracy and bias of high-frequency ultrasonography in measuring white spot lesion depth: an in-vitro comparison with micro-CT},
  journal = {Clinical Oral Investigations},
  year = {2026},
  volume = {30},
  number = {6},
  doi = {10.1007/s00784-026-06956-y}
}
Kostenlos starten →