Flap size as a confounding variable in flap perfusion measurement with the Oxygen-to-see (O2C) analysis system in microvascular head and neck reconstruction – a retrospective analysis

Mark Ooms, Philipp Winnand, Marius Heitzer, Anna Bock, Marie Sophie Katz, Johannes Bickenbach, Frank Hölzle, Ali Modabber

Clinical Oral Investigations · 2025

DOI: 10.1007/s00784-025-06468-1 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Clinical Oral Investigations, 2025. Verfasst von Mark Ooms und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Oral and Maxillofacial Surgery, University Hospital RWTH Aachen, Pauwelsstraße 30, 52074, Aachen, Germany. mooms@ukaachen.de. Hauptthemen laut MEDLINE: Plastic Surgery Procedures, Free Tissue Flaps, Head and Neck Neoplasms, Oxygen. Die Arbeit nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Universitätsklinikum RWTH Aachen.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

The Oxygen-to-see (O2C) analysis system is used for flap monitoring based on predefined threshold values for flap perfusion. However, flap size may be a confounding variable. The aim of this study was to investigate the relationship between flap size and flap perfusion in microvascular head and neck reconstruction. Materials and methods Flap perfusion values measured with the O2C analysis system between 2011 and 2020 in 252 patients undergoing microvascular head and neck reconstruction with a radial free forearm flap (RFFF) or anterolateral thigh flap (ALTF) were retrospectively analyzed. Intraoperative and postoperative flap blood flow, hemoglobin concentration, and hemoglobin oxygen saturation at 8- and 2-mm tissue depths were compared between small (≤ median flap size) and large flaps (> median flap size) for RFFFs and ALTFs separately.

Results

Intraoperative and postoperative hemoglobin concentration at a 2-mm tissue depth differed between small and large ALTFs (65.0 arbitrary units [AU] vs. 51.0 AU, p = 0.007; and 51.5 AU vs. 39.0 AU, p = 0.019). Both differences persisted in multivariable analysis ( p < 0.001 and p = 0.012). Other differences were not observed or did not persist in multivariable analysis for RFFFs and ALTFs (all p > 0.05).

Conclusions

Microvascular free flap perfusion is not related to flap size in terms of flap blood flow and hemoglobin oxygen saturation. This underscores the validity of predefined absolute threshold values in the context of flap monitoring based on perfusion measurement with the O2C analysis system. Clinical relevance Flap perfusion measurement with the Oxygen-to-see (O2C) analysis system can be used for flap monitoring of small and large flaps.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Plastic Surgery ProceduresFree Tissue FlapsHead and Neck NeoplasmsOxygenSurgical FlapsHumansRetrospective StudiesMaleFemaleMiddle AgedAgedHemoglobinsOxygen SaturationAdultRegional Blood Flow

Stichwörter der Autor:innen

Flap perfusion monitoringFlap sizeFree flapMicrovascular head and neck reconstructionOxygen-to-see analysis system

Substanzen

HemoglobinsOxygen

Autor:innen und Institutionen

  • Department of Oral and Maxillofacial Surgery, University Hospital RWTH Aachen, Pauwelsstraße 30, 52074, Aachen, Germany. mooms@ukaachen.de.
  • Department of Oral and Maxillofacial Surgery, University Hospital RWTH Aachen, Pauwelsstraße 30, 52074, Aachen, Germany.
  • Department of Intensive Care Medicine, University Hospital RWTH Aachen, Pauwelsstraße 30, 52074, Aachen, Germany.

Förderung

  • Universitätsklinikum RWTH Aachen

Zitieren

APA

Ooms, M., Winnand, P., Heitzer, M., Bock, A., Katz, M. S., Bickenbach, J., Hölzle, F., & Modabber, A. (2025). Flap size as a confounding variable in flap perfusion measurement with the Oxygen-to-see (O2C) analysis system in microvascular head and neck reconstruction – a retrospective analysis. Clinical Oral Investigations. https://doi.org/10.1007/s00784-025-06468-1

Vancouver

Ooms M, Winnand P, Heitzer M, Bock A, Katz MS, Bickenbach J, et al. Flap size as a confounding variable in flap perfusion measurement with the Oxygen-to-see (O2C) analysis system in microvascular head and neck reconstruction – a retrospective analysis. Clinical Oral Investigations. 2025. doi:10.1007/s00784-025-06468-1

BibTeX
@article{Ooms2025,
  author = {Ooms, Mark and Winnand, Philipp and Heitzer, Marius and Bock, Anna and Katz, Marie Sophie and Bickenbach, Johannes and Hölzle, Frank and Modabber, Ali},
  title = {Flap size as a confounding variable in flap perfusion measurement with the Oxygen-to-see (O2C) analysis system in microvascular head and neck reconstruction – a retrospective analysis},
  journal = {Clinical Oral Investigations},
  year = {2025},
  doi = {10.1007/s00784-025-06468-1}
}
Kostenlos starten →