EUVIMEDEUVIMED entdecken
VergleichsstudieOpen Access · CC BY 4.0

Comparison between two different registration protocols for the count of sleep bruxism events in a sample of healthy individuals

Ovidiu Ionut Saracutu, Matteo Pollis, Ludovico Bulferi Bulferetti, Elifnaz Kaya, Edoardo Ferrari Cagidiaco, Marco Ferrari, Anna Colonna, Daniele Manfredini

Clinical Oral Investigations · 2025

DOI: 10.1007/s00784-025-06394-2 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Clinical Oral Investigations, 2025. Verfasst von Ovidiu Ionut Saracutu und 7 weiteren Autor:innen. Beteiligte Einrichtung: School of Dentistry, Department of Medical Biotechnologies, University of Siena, Siena, 53100, Italy. Hauptthemen laut MEDLINE: Electromyography, Sleep Bruxism. Die Arbeit wurde bisher 15-mal zitiert (Crossref) und nennt 58 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

The aim of the present research is to compare, in a group of healthy young individuals, the number of sleep bruxism (SB) events with the use of two different electromyographic (EMG) devices, which adopt two different protocols for scoring the events. Materials and methods The study was performed on 10 individuals recruited at the University of Siena, Italy. Participants were asked to perform one night of sleep-time recording wearing simultaneously, in their home environment, two different EMG devices, the Bruxoff (Bruxoff ® , Turin, Italy) and the dia-BRUXO (dia-BRUXO, Biotech-Novations, Sanremo, Italy). SB events were identified by the Bruxoff software algorithm as the EMG peaks of an amplitude that exceeds 10% of the maximum voluntary contraction (MVC) performed by the subject before the start of the registration, and that are preceded by a 20% increase in the heart rate compared to the baseline. Conversely, the dia-BRUXO software identified the EMG peaks exceeding 10% of the maximum contraction (MC) recorded during the entire registration period. Both devices were set to count the events lasting 1–5 s. After each night of recording, data was downloaded through the dedicated software of each device, which generates a report with the number of sleep bruxism events.

Results

All the recruited participants completed the registrations at the first attempt without reporting any detachment of the electrodes. The Pearson test showed a strong correlation between the number of SB events scored by the two devices ( r = 0.9613, p < 0.001).

Conclusions

Within the limitation of this study, our results suggest the existence of a strong correlation between the two different registration protocols for the registration of the number of SB events. Clinical relevance The present study suggest that clinicians can adopt easier and user-friendly protocols for the registration of sleep bruxism events, that do not necessarily require a pre-calibration of the device.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ElectromyographySleep BruxismAdultFemaleHumansMaleAlgorithmsHealthy VolunteersItalySoftware

Stichwörter der Autor:innen

ElectromyographyInstrumentally-based assessmentMaximum voluntary contractionSTABSleep bruxism

Autor:innen und Institutionen

  • School of Dentistry, Department of Medical Biotechnologies, University of Siena, Siena, 53100, Italy.
  • School of Dentistry, Department of Medical Biotechnologies, University of Siena, Siena, 53100, Italy. l.bulferibulferet@student.unisi.it.

Zitieren

APA

Saracutu, O. I., Pollis, M., Bulferi Bulferetti, L., Kaya, E., Ferrari Cagidiaco, E., Ferrari, M., Colonna, A., & Manfredini, D. (2025). Comparison between two different registration protocols for the count of sleep bruxism events in a sample of healthy individuals. Clinical Oral Investigations. https://doi.org/10.1007/s00784-025-06394-2

Vancouver

Saracutu OI, Pollis M, Bulferi Bulferetti L, Kaya E, Ferrari Cagidiaco E, Ferrari M, et al. Comparison between two different registration protocols for the count of sleep bruxism events in a sample of healthy individuals. Clinical Oral Investigations. 2025. doi:10.1007/s00784-025-06394-2

BibTeX
@article{Saracutu2025,
  author = {Saracutu, Ovidiu Ionut and Pollis, Matteo and Bulferi Bulferetti, Ludovico and Kaya, Elifnaz and Ferrari Cagidiaco, Edoardo and Ferrari, Marco and Colonna, Anna and Manfredini, Daniele},
  title = {Comparison between two different registration protocols for the count of sleep bruxism events in a sample of healthy individuals},
  journal = {Clinical Oral Investigations},
  year = {2025},
  doi = {10.1007/s00784-025-06394-2}
}
Kostenlos starten →