EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Advancement in continuous lactate determination in untreated human serum: multipolymer-based amperometric biosensor coupled with a low-cost 3D-printed microfluidic cell

Álvaro Jesús Sainz-Calvo, Alfonso Sierra-Padilla, Dolores Bellido-Milla, Lorena Blanco-Díaz, Juan Jesús Fernández-Alba, Carmen González-Macías, Juan José García-Guzmán, José María Palacios-Santander, Laura Cubillana-Aguilera

Microchimica Acta · 2026-01 · 193 (1)

DOI: 10.1007/s00604-025-07771-0 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Microchimica Acta, online erschienen am 27.12.2025. Verfasst von Álvaro Jesús Sainz-Calvo und 8 weiteren Autor:innen. Beteiligte Einrichtung: Institute of Research on Electron Microscopy and Materials (IMEYMAT), Department of Analytical Chemistry, Faculty of Sciences, Campus de Excelencia I…. Hauptthemen laut MEDLINE: Biosensing Techniques, Lactic Acid, Printing, Three-Dimensional, Electrochemical Techniques. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 98 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Universidad de Cadiz.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Enzyme saturation in biosensing has been established as a main hindrance in lactate monitoring due to the high concentration of this species in biological samples. In this paper, a Sonogel–carbon transducer has been used for the fabrication of an amperometric biosensor for lactate (LA) determination. The sensor was constructed by immobilizing lactate oxidase as the bioreceptor and coating the electrode surface with multipolymer layers to minimize enzyme saturation and improve operational stability. This biosensor shows adequate sensitivity (1.74 µA mM − 1 cm -2 ) in a linear working range from 0.2 mM to 20 mM, suitable for physiological lactate concentrations in biofluids. The response of the biosensor was studied in physiological medium using several potential interferents (glucose, uric acid, dopamine, ascorbic acid, KCl, NaCl, CaCl 2 and BSA), demonstrating negligible interference from these common electroactive species. A microfluidic cell was designed and fabricated using polyethylene terephthalate glycol (PETG) through a rapid and low-cost 3D printing process. Finally, the prepared microfluidic cell was applied to lactate monitoring in flow regime using the biosensor with both synthetic and real untreated human serum samples, showing excellent correlation with two standard reference techniques and confirming its reliability. These results highlight the potential of the proposed microfluidic amperometric biosensor as a robust, real-time tool for biomedical lactate monitoring. Graphical Abstract

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Biosensing TechniquesLactic AcidPrinting, Three-DimensionalElectrochemical TechniquesHumansLimit of DetectionMixed Function OxygenasesEnzymes, ImmobilizedElectrodesPolyethylene Glycols

Stichwörter der Autor:innen

3D printed microfluidic cellAmperometryEnzyme encapsulationReal-time lactate monitoringSonogel–carbon-based lactate biosensorUntreated serum analysis

Substanzen

Lactic Acidlactate 2-monooxygenaseMixed Function OxygenasesEnzymes, ImmobilizedPolyethylene Glycols

Autor:innen und Institutionen

  • Institute of Research on Electron Microscopy and Materials (IMEYMAT), Department of Analytical Chemistry, Faculty of Sciences, Campus de Excelencia Internacional del Mar (CEIMAR), University of Cadiz, Campus Universitar…
  • Departamento de Obstetricia y Ginecología, Hospital Universitario de Puerto Real, Puerto Real, Cádiz, 11510, Spain.

Förderung

  • Universidad de Cadiz

Zitieren

APA

Sainz-Calvo, Á. J., Sierra-Padilla, A., Bellido-Milla, D., Blanco-Díaz, L., Fernández-Alba, J. J., González-Macías, C., García-Guzmán, J. J., Palacios-Santander, J. M., & Cubillana-Aguilera, L. (2026). Advancement in continuous lactate determination in untreated human serum: multipolymer-based amperometric biosensor coupled with a low-cost 3D-printed microfluidic cell. Microchimica Acta, 193(1). https://doi.org/10.1007/s00604-025-07771-0

Vancouver

Sainz-Calvo ÁJ, Sierra-Padilla A, Bellido-Milla D, Blanco-Díaz L, Fernández-Alba JJ, González-Macías C, et al. Advancement in continuous lactate determination in untreated human serum: multipolymer-based amperometric biosensor coupled with a low-cost 3D-printed microfluidic cell. Microchimica Acta. 2026;193(1). doi:10.1007/s00604-025-07771-0

BibTeX
@article{SainzCalvo2026,
  author = {Sainz-Calvo, Álvaro Jesús and Sierra-Padilla, Alfonso and Bellido-Milla, Dolores and Blanco-Díaz, Lorena and Fernández-Alba, Juan Jesús and González-Macías, Carmen and García-Guzmán, Juan José and Palacios-Santander, José María and Cubillana-Aguilera, Laura},
  title = {Advancement in continuous lactate determination in untreated human serum: multipolymer-based amperometric biosensor coupled with a low-cost 3D-printed microfluidic cell},
  journal = {Microchimica Acta},
  year = {2026},
  volume = {193},
  number = {1},
  doi = {10.1007/s00604-025-07771-0}
}