Proving the automatic benchtop electrochemical station for the development of dopamine and paracetamol sensors
Microchimica Acta · 2024
DOI: 10.1007/s00604-024-06454-6 ↗Kurz gefasst
Fachartikel in der Zeitschrift Microchimica Acta, 2024. Verfasst von Marek Haššo und 6 weiteren Autor:innen. Beteiligte Einrichtung: Institute of Analytical Chemistry, Faculty of Chemical and Food Technology, Slovak University of Technology in Bratislava, Radlinského 9, Bratislava,…. Hauptthemen laut MEDLINE: Acetaminophen, Dopamine, Gold, Metal Nanoparticles. Die Arbeit wurde bisher 10-mal zitiert (Crossref) und nennt 58 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Vedecká Grantová Agentúra MŠVVaŠ SR a SAV, IUPAC, Slovak Research and Development Agency.
Abstract (Originalsprache)
The introduced work represents an implementation of the automatic benchtop electrochemical station (BES) as an effective tool for the possibilities of high-throughput preparation of modified sensor/biosensors, speeding up the development of the analytical method, and automation of the analytical procedure for the determination of paracetamol (PAR) and dopamine (DOP) as target analytes. Within the preparation of gold nanoparticles modified screen-printed carbon electrode (AuNPs-SPCE) by electrodeposition, the deposition potential E DEP , the deposition time t DEP , and the concentration of HAuCl 4 were optimized and their influence was monitored on 1 mM [Ru(NH 3 ) 6 ] 3+/2+ redox probe and 50 μM DOP. The morphology of the AuNPs-SPCE prepared at various modification conditions was observed by SEM. The analytical performance of the AuNPs-SPCE prepared at different modification conditions was evaluated by a construction of the calibration curves of DOP and PAR. SPCE and AuNPs-SPCE at modification condition providing the best sensitivity to PAR and DOP, were successfully used to determine PAR and DOP in tap water by “spike-recovery” approach. The BES yields better reproducibility of the preparation of AuNPs-SPCE (RSD = 3.0%) in comparison with the case when AuNPs-SPCE was prepared manually by highly skilled laboratory operator (RSD = 7.0%). Graphical abstract
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Institute of Analytical Chemistry, Faculty of Chemical and Food Technology, Slovak University of Technology in Bratislava, Radlinského 9, Bratislava, 812 37, Slovakia.
- Department of Chemistry and Biochemistry, Mendel University in Brno, Zemedelska 1, Brno, 613 00, Czech Republic.
- Department of Chemistry and Biochemistry, Mendel University in Brno, Zemedelska 1, Brno, 613 00, Czech Republic. zitkao@seznam.cz.
Förderung
- Vedecká Grantová Agentúra MŠVVaŠ SR a SAV
- IUPAC
- Slovak Research and Development Agency
- National Scholarship Program
- ERFD
- Mendel University in Brno
Zitieren
APA
Haššo, M., Kudr, J., Zítka, J., Šílený, J., Švec, P., Švorc, Ľ., & Zítka, O. (2024). Proving the automatic benchtop electrochemical station for the development of dopamine and paracetamol sensors. Microchimica Acta. https://doi.org/10.1007/s00604-024-06454-6
Vancouver
Haššo M, Kudr J, Zítka J, Šílený J, Švec P, Švorc Ľ, et al. Proving the automatic benchtop electrochemical station for the development of dopamine and paracetamol sensors. Microchimica Acta. 2024. doi:10.1007/s00604-024-06454-6
BibTeX
@article{Hao2024,
author = {Haššo, Marek and Kudr, Jiří and Zítka, Jan and Šílený, Jan and Švec, Pavel and Švorc, Ľubomír and Zítka, Ondřej},
title = {Proving the automatic benchtop electrochemical station for the development of dopamine and paracetamol sensors},
journal = {Microchimica Acta},
year = {2024},
doi = {10.1007/s00604-024-06454-6}
}