Smart crutch tipsTM real-time feedback improves adherence to partial weight-bearing protocols following lower extremity fracture surgery: a pilot study
European Journal of Orthopaedic Surgery & Traumatology · 2026
DOI: 10.1007/s00590-026-04879-8 ↗Kurz gefasst
Randomisierte kontrollierte Studie in der Zeitschrift European Journal of Orthopaedic Surgery & Traumatology, 2026. Verfasst von Michael Coons und 4 weiteren Autor:innen. Beteiligte Einrichtung: New York University Langone Orthopedic Hospital, New York, USA. Hauptthemen laut MEDLINE: Crutches, Patient Compliance, Fractures, Bone. Die Arbeit nennt 30 Literaturangaben.
Abstract (Originalsprache)
Purpose
To evaluate the effectiveness of Smart Crutch Tips™ in promoting adherence to prescribed early partial weight-bearing (PWB) protocols and to assess patient experience with real-time weight-bearing feedback during home rehabilitation.
Methods
Twenty patients with lower extremity fractures prescribed partial weight-bearing (PWB) were randomized into two groups utilizing real-time feedback crutch tips (RFC). The Intervention group ( n = 10) used Smart Crutch Tips™ (ComeBack Mobility Inc., Kiev, Ukraine), which provided real-time feedback on weight-bearing compliance. The Control group ( n = 10) also used crutches equipped with Smart Crutch Tips™, but notifications were disabled, allowing passive data collection without patient feedback. Weight-bearing compliance was defined as the percentage of steps within ± 10% of the prescribed target. Secondary outcomes included patient satisfaction and device usability assessed using the System Usability Scale (SUS) throughout the home rehabilitation period.
Results
The Intervention group demonstrated significantly greater PWB compliance compared with the Control group (73.6% vs. 21.1%, p < 0.01). The Intervention group reported a mean SUS score of 84.25, indicating excellent usability. No device-related complications or adverse events were observed. Patient satisfaction, assessed through video interviews, was high, with participants in the Intervention group expressing a strong willingness to recommend the device to others with similar injuries and rating it 10 out of 10.
Conclusion
Smart Crutch Tips™ significantly improved adherence to prescribed partial weight-bearing protocols following lower extremity fracture surgery. The RFC device demonstrated high usability, excellent patient satisfaction, and no device-related complications during home rehabilitation throughout the study period.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- New York University Langone Orthopedic Hospital, New York, USA.
- New York University Langone Orthopedic Hospital, New York, USA. kenneth.egol@nyulangone.org.
Zitieren
APA
Coons, M., Cherry, F., Miller, S., Fisher, N., & Egol, K. (2026). Smart crutch tipsTM real-time feedback improves adherence to partial weight-bearing protocols following lower extremity fracture surgery: a pilot study. European Journal of Orthopaedic Surgery & Traumatology. https://doi.org/10.1007/s00590-026-04879-8
Vancouver
Coons M, Cherry F, Miller S, Fisher N, Egol K. Smart crutch tipsTM real-time feedback improves adherence to partial weight-bearing protocols following lower extremity fracture surgery: a pilot study. European Journal of Orthopaedic Surgery & Traumatology. 2026. doi:10.1007/s00590-026-04879-8
BibTeX
@article{Coons2026,
author = {Coons, Michael and Cherry, Fiona and Miller, Sophia and Fisher, Nina and Egol, Kenneth},
title = {Smart crutch tipsTM real-time feedback improves adherence to partial weight-bearing protocols following lower extremity fracture surgery: a pilot study},
journal = {European Journal of Orthopaedic Surgery & Traumatology},
year = {2026},
doi = {10.1007/s00590-026-04879-8}
}