DOI: 10.1007/s00590-026-04766-2 ↗

Kurz gefasst

Validierungsstudie in der Zeitschrift European Journal of Orthopaedic Surgery & Traumatology, online erschienen am 11.05.2026. Verfasst von Austin Thomassen und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopaedic Surgery, Louisiana State University Health Shreveport, Shreveport, USA. Hauptthemen laut MEDLINE: Printing, Three-Dimensional, Bone Screws, Orthopedic Procedures, Orthopedics. Die Arbeit nennt 94 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Screw stripping is a common complication in orthopaedic surgery that can compromise fixation stability. This study evaluates the relationship between stripping torque (ST) and infill density in 3D-printed bone surrogates, establishes a conversion factor between surrogates and ASTM F-1839-08 foam, and tests the hypothesis that screw insertion angle up to 45 °C does not affect ST.

Methods

PLA surrogates with 3D Honeycomb or Gyroid infill patterns were printed at densities ranging from 5 to 27%. These samples were used for screw insertion at angles of 0°, 15°, 30°, and 45°. Partially threaded compression screws were manually inserted while continuously recording torque, performing three repetitions for each configuration ( N = 3). The maximum torque value was defined as the peak value (ST). Mean values were analyzed using linear regression to evaluate the correlation between density and ST, while multivariable linear regression was used to assess the influence of screw insertion angle.

Results

3D Honeycomb ST ranged from 0.23 Nm ± 0.02 (5%) to 3.89 Nm ± 0.14 (27%), and from 0.32 Nm ± 0.06 to 3.68 Nm ± 0.46 for Gyroid. Both patterns showed strong linear correlations between density and ST (R 2 > 0.98, p < 0.01). ST values in 3D-printed surrogates aligned with ASTM foam (0.52 Nm ± 0.08 to 2.65 Nm ± 0.55 for 10–20 PCF). No significant differences were observed among insertion angles ( p > 0.65).

Conclusion

3D-printed surrogates with 3D Honeycomb or Gyroid infills at up to 27% density, for insertion inclinations up to 45 degrees exhibit predictable, density-dependent mechanical behavior comparable to ASTM polyurethane foams supporting their use for screw insertion training. Screw insertion angles up to 45° have minimal effect on ST, potentially simplifying surrogate manufacturing. Clinical relevance These findings support the use of 3D printed surrogates across diverse training scenarios, including those with uncertain screw insertion angles, and highlight their potential to simulate diverse bone densities and pathologies, offering a more realistic and versatile solution than traditional foams. Their predictable mechanical behavior offers a safe, standardized, and accessible training method for orthopaedic residents, with potential to reduce screw stripping and fixation failure in patients.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Printing, Three-DimensionalBone ScrewsOrthopedic ProceduresOrthopedicsTorqueHumans

Stichwörter der Autor:innen

screwsstrippingsurrogatestorquetraining

Autor:innen und Institutionen

  • Department of Orthopaedic Surgery, Louisiana State University Health Shreveport, Shreveport, USA.
  • Ospedale Niguarda Ca' Granda, Milan, Italy.
  • Department of Orthopaedic Surgery, Louisiana State University Health Shreveport, Shreveport, USA. giovanni.solitro@lsuhs.edu.

Zitieren

APA

Thomassen, A., Lee, B., Granger, M., Massey, P., Chauvin, B., Addevico, F., & Solitro, G. (2026). Mechanical validation of 3D-printed orthopaedic surrogate models for screw insertion training. European Journal of Orthopaedic Surgery & Traumatology, 36(1). https://doi.org/10.1007/s00590-026-04766-2

Vancouver

Thomassen A, Lee B, Granger M, Massey P, Chauvin B, Addevico F, et al. Mechanical validation of 3D-printed orthopaedic surrogate models for screw insertion training. European Journal of Orthopaedic Surgery & Traumatology. 2026;36(1). doi:10.1007/s00590-026-04766-2

BibTeX
@article{Thomassen2026,
  author = {Thomassen, Austin and Lee, Benjamin and Granger, Mason and Massey, Patrick and Chauvin, Brad and Addevico, Francesco and Solitro, Giovanni},
  title = {Mechanical validation of 3D-printed orthopaedic surrogate models for screw insertion training},
  journal = {European Journal of Orthopaedic Surgery & Traumatology},
  year = {2026},
  volume = {36},
  number = {1},
  doi = {10.1007/s00590-026-04766-2}
}
Kostenlos starten →