EUVIMEDEUVIMED entdecken
VergleichsstudieOpen Access · CC BY 4.0

CT vs. MRI-based synthetic CT: accuracy of thoracic and lumbar pedicle screw placement using 3D-printed guides

Jules Cool, Peter P G Lafranca, Joëll Magré, Johannes G G Dobbe, Moyo C Kruyt, Geert J Streekstra, Tom P C Schlösser, Barend J van Royen

European Spine Journal · 2025

DOI: 10.1007/s00586-025-09367-x ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift European Spine Journal, 2025. Verfasst von Jules Cool und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopaedic Surgery and Sports Medicine, Amsterdam University Medical Centers, Amsterdam, Netherlands. j.cool@amsterdamumc.nl. Hauptthemen laut MEDLINE: Pedicle Screws, Printing, Three-Dimensional, Tomography, X-Ray Computed, Thoracic Vertebrae. Die Arbeit nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Amsterdam Movement Sciences.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

3D-printed patient-specific guides enable accurate pedicle screw placement in the treatment of paediatric spinal deformities. MRI-based synthetic CT (sCT) eliminates the radiation exposure associated with the preoperative planning CT, while still providing detailed bony morphology. It may therefore offer a radiation-free method for pedicle screw navigation. Thus, the aim of this study was to compare the accuracy of thoracic and lumbar pedicle screw placement using 3D-printed guides based on conventional CT versus sCT.

Methods

Full-spine MRI and CT images of four human cadaveric spines were obtained, followed by AI-generation of sCTs, the digital planning of pedicle screws and 3D-printed guide manufacturing. A total of 62 K-wires were inserted, 30 sCT-based and 32 CT-based. Accuracy was evaluated by superimpose analysis of planned and achieved trajectories, and by the postoperative grading of the cortical breach rate using the Gertzbein-Robbins classification, based on projected screws along the K-wires.

Results

Median translational entry-point deviations were 1.0 mm vertically and 0.5 mm horizontally for both sCT-based guides and CT-based guides. Angular deviations in the axial plane were 1.0˚ for sCT and 0.9˚ for CT. In the sagittal plane these were 1.8˚ and 2.1˚, respectively. No significant differences were observed between sCT and CT. Grading of pedicle’s cortical breach showed no severe malposition for both sCT- and CT-based guides. Conclusion sCT matched the accuracy of conventional CT in providing source data for personalized 3D-printed guides for thoracic and lumbar pedicle screw placement in terms of translational- and angular deviations, and cortical breach rates in this cadaveric study.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Pedicle ScrewsPrinting, Three-DimensionalTomography, X-Ray ComputedThoracic VertebraeLumbar VertebraeMagnetic Resonance ImagingSpinal FusionSurgery, Computer-AssistedHumansCadaverMaleFemale

Stichwörter der Autor:innen

3D-printed guidesAccuracyMRIPedicle screwsSpineSuperimpose analysis

Autor:innen und Institutionen

  • Department of Orthopaedic Surgery and Sports Medicine, Amsterdam University Medical Centers, Amsterdam, Netherlands. j.cool@amsterdamumc.nl.
  • Musculoskeletal Health - Restoration and Development, Amsterdam Movement Sciences, Amsterdam, Netherlands. j.cool@amsterdamumc.nl.
  • Department of Orthopaedic Surgery, University Medical Center Utrecht, Utrecht, Netherlands.
  • 3D lab, University Medical Center Utrecht, Utrecht, Netherlands.
  • Musculoskeletal Health - Restoration and Development, Amsterdam Movement Sciences, Amsterdam, Netherlands.
  • Biomedical Engineering and Physics, Amsterdam University Medical Centers, Amsterdam, Netherlands.

Förderung

  • Amsterdam Movement Sciences

Zitieren

APA

Cool, J., Lafranca, P. P. G., Magré, J., Dobbe, J. G. G., Kruyt, M. C., Streekstra, G. J., Schlösser, T. P. C., & van Royen, B. J. (2025). CT vs. MRI-based synthetic CT: accuracy of thoracic and lumbar pedicle screw placement using 3D-printed guides. European Spine Journal. https://doi.org/10.1007/s00586-025-09367-x

Vancouver

Cool J, Lafranca PPG, Magré J, Dobbe JGG, Kruyt MC, Streekstra GJ, et al. CT vs. MRI-based synthetic CT: accuracy of thoracic and lumbar pedicle screw placement using 3D-printed guides. European Spine Journal. 2025. doi:10.1007/s00586-025-09367-x

BibTeX
@article{Cool2025,
  author = {Cool, Jules and Lafranca, Peter P G and Magré, Joëll and Dobbe, Johannes G G and Kruyt, Moyo C and Streekstra, Geert J and Schlösser, Tom P C and van Royen, Barend J},
  title = {CT vs. MRI-based synthetic CT: accuracy of thoracic and lumbar pedicle screw placement using 3D-printed guides},
  journal = {European Spine Journal},
  year = {2025},
  doi = {10.1007/s00586-025-09367-x}
}
Kostenlos starten →