Usefulness of portrait imaging for intraoperative cone beam computed tomography in adolescent idiopathic scoliosis surgery: a phantom study
European Spine Journal · 2025
DOI: 10.1007/s00586-025-08947-1 ↗Kurz gefasst
Fachartikel in der Zeitschrift European Spine Journal, 2025. Verfasst von Masahiro Fuji und 5 weiteren Autor:innen. Beteiligte Einrichtung: Shinshu University Hospital, Matsumoto, Japan. Hauptthemen laut MEDLINE: Cone-Beam Computed Tomography, Scoliosis, Phantoms, Imaging, Spinal Fusion. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 22 Literaturangaben.
Abstract (Originalsprache)
STUDY DESIGN: Comparative study of radiation doses in intraoperative cone-beam computed tomography (CBCT) using an anthropomorphic phantom. OBJECTIVE: Compare the portrait and conventional landscape CBCT imaging methods under intraoperative conditions focusing on dose reduction rate, image quality, and imaging range. Intraoperative CBCT improves the accuracy of pedicle screw (PS) insertion in posterior spinal fusion (PSF) for adolescent idiopathic scoliosis (AIS). We recently introduced portrait imaging for intraoperative CBCT in PS insertion using surgery navigation for AIS. However, evidence is scarce on the comparative radiation dose and image quality of this orientation versus landscape imaging. METHODS: A human body phantom was equipped with radiophotoluminescent glass dosimeters both inside and on the surface for measuring radiation doses under portrait and landscape CBCT imaging methods. Radiation doses were evaluated by average dose index (ADI), surface dose (Dsurf), central dose (Dcent), and displayed dose (Ddisp). Image quality was assessed using the contrast-to-noise ratio (CNR) on phantom images. Imaging range was measured using sagittal image craniocaudal length and axial image diameter. RESULTS: The respective mean ± standard deviation radiation doses for portrait and landscape imaging were 42.9 ± 0.7 and 50.6 ± 1.1 mGy for ADI (i.e., approximately 15% less radiation dose for portrait), 53.9 ± 1.6 and 65.2 ± 1.2 mGy for Dsurf, 21.1 ± 1.4 and 21.4 ± 1.0 mGy for Dcent, and 146.6 and 145.0 mGy for Ddisp. The CNR results revealed comparable image qualities for portrait and landscape imaging of 2.23 ± 0.22 and 2.18 ± 0.19, respectively. The respective craniocaudal imaging range was 24.6 cm and 18.4 cm, while the axial imaging range was 17.6 cm and 23.7 cm. CONCLUSION: Portrait-mode CBCT allows for a wider craniocaudal imaging range and enables intraoperative CBCT imaging at lower radiation doses without compromising image quality.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Shinshu University Hospital, Matsumoto, Japan.
- Shinshu University Hospital, Matsumoto, Japan. oba1@hotmail.co.jp.
Zitieren
APA
Fuji, M., Oba, H., Ikegami, S., Uehara, M., Fujinaga, Y., & Takahashi, J. (2025). Usefulness of portrait imaging for intraoperative cone beam computed tomography in adolescent idiopathic scoliosis surgery: a phantom study. European Spine Journal. https://doi.org/10.1007/s00586-025-08947-1
Vancouver
Fuji M, Oba H, Ikegami S, Uehara M, Fujinaga Y, Takahashi J. Usefulness of portrait imaging for intraoperative cone beam computed tomography in adolescent idiopathic scoliosis surgery: a phantom study. European Spine Journal. 2025. doi:10.1007/s00586-025-08947-1
BibTeX
@article{Fuji2025,
author = {Fuji, Masahiro and Oba, Hiroki and Ikegami, Shota and Uehara, Masashi and Fujinaga, Yasunari and Takahashi, Jun},
title = {Usefulness of portrait imaging for intraoperative cone beam computed tomography in adolescent idiopathic scoliosis surgery: a phantom study},
journal = {European Spine Journal},
year = {2025},
doi = {10.1007/s00586-025-08947-1}
}
EUVIMED