Validity of a fast automated 3d spine reconstruction measurements for biplanar radiographs: SOSORT 2024 award winner

Weiying Chen iD, Mahdieh Khodaei iD, Marek Reformat iD, Edmond Lou iD

European Spine Journal · 2025

DOI: 10.1007/s00586-024-08375-7 ↗

Kurz gefasst

Validierungsstudie in der Zeitschrift European Spine Journal, 2025. Verfasst von Weiying Chen und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Electrical and Computer Engineering, University of Alberta, Donadeo ICE 11-263, 9211-116 Street NW, Edmonton, AB, T6G 1H9, Canada. Hauptthemen laut MEDLINE: Scoliosis, Imaging, Three-Dimensional, Spine, Radiography. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 14 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NSERC, Women and Children's Health Research Institute, Alberta Innovates.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To validate a fast 3D biplanar spinal radiograph reconstruction method with automatic extract curvature parameters using artificial intelligence (AI).

Methods

Three-hundred eighty paired, posteroanterior and lateral, radiographs from the EOS X-ray system of children with adolescent idiopathic scoliosis were randomly selected from the database. For the AI model development, 304 paired images were used for training; 76 pairs were employed for testing. The validation was evaluated by comparing curvature parameters, including Cobb angles (CA), apical axial vertebral rotation (AVR), kyphotic angle (T1–T12 KA), and lordotic angle (L1–L5 LA), to manual measurements from a rater with 8 years of scoliosis experience. The mean absolute differences ± standard deviation (MAD ± SD), the percentage of measurements within the clinically acceptable errors, the standard error of measurement (SEM), and the inter-method intraclass correlation coefficient ICC [2,1] were calculated. The average reconstruction speed of the 76 test images was recorded.

Results

Among the 76 test images, 134 and 128 CA were exported automatically and measured manually, respectively. The MAD ± SD for CA, AVR at apex, KA, and LA were 3.3° ± 3.5°, 1.5° ± 1.5°, 3.3° ± 2.6° and 3.5° ± 2.5°, respectively, and 98% of these measurements were within the clinical acceptance errors. The SEMs and the ICC [2,1] for the compared parameters were all less than 0.7° and > 0.94, respectively. The average time to display the 3D spine and report the measurements was 5.2 ± 1.3 s.

Conclusion

The developed AI algorithm could reconstruct a 3D scoliotic spine within 6 s, and the automatic curvature parameters were accurately and reliably extracted from the reconstructed images.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ScoliosisImaging, Three-DimensionalSpineRadiographyHumansAdolescentFemaleMaleChildArtificial IntelligenceAwards and Prizes

Stichwörter der Autor:innen

3D spinal reconstructionAdolescent idiopathic scoliosisArtificial intelligenceCurvature parameters

Autor:innen und Institutionen

  • Department of Electrical and Computer Engineering, University of Alberta, Donadeo ICE 11-263, 9211-116 Street NW, Edmonton, AB, T6G 1H9, Canada.
  • Department of Electrical and Computer Engineering, University of Alberta, Donadeo ICE 11-263, 9211-116 Street NW, Edmonton, AB, T6G 1H9, Canada. elou@ualberta.ca.

ORCID-Profile: Weiying Chen, Mahdieh Khodaei, Marek Reformat, Edmond Lou

Förderung

  • NSERC
  • Women and Children's Health Research Institute
  • Alberta Innovates

Zitieren

APA

Chen, W., Khodaei, M., Reformat, M., & Lou, E. (2025). Validity of a fast automated 3d spine reconstruction measurements for biplanar radiographs: SOSORT 2024 award winner. European Spine Journal. https://doi.org/10.1007/s00586-024-08375-7

Vancouver

Chen W, Khodaei M, Reformat M, Lou E. Validity of a fast automated 3d spine reconstruction measurements for biplanar radiographs: SOSORT 2024 award winner. European Spine Journal. 2025. doi:10.1007/s00586-024-08375-7

BibTeX
@article{Chen2025,
  author = {Chen, Weiying and Khodaei, Mahdieh and Reformat, Marek and Lou, Edmond},
  title = {Validity of a fast automated 3d spine reconstruction measurements for biplanar radiographs: SOSORT 2024 award winner},
  journal = {European Spine Journal},
  year = {2025},
  doi = {10.1007/s00586-024-08375-7}
}
Kostenlos starten →