Increased cuff-leak pressure after abdominal and spine surgeries: a simple and novel cuff-leak test for tracheal extubation
Journal of Anesthesia · 2025
DOI: 10.1007/s00540-025-03487-w ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Anesthesia, 2025. Verfasst von Takayuki Yamada und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Anesthesiology, Pain and Palliative Care Medicine, Chiba University Hospital, 1-8-1 Inohana, Chuo-Ku, Chiba, 260-8670, Japan. tyamada_v…. Hauptthemen laut MEDLINE: Airway Extubation, Intubation, Intratracheal, Abdomen, Spine. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 28 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Kaken Pharmaceutical.
Themen: Beatmung und Weaning
Abstract (Originalsprache)
Purpose
Cuff-leak volume (CLV) tests are recommended to avoid extubation failure. We developed a novel cuff-leak pressure (CLP) test that quantitatively assesses upper airway resistance outside the tracheal tube. We hypothesized that CLP (airway pressure during apnea with the cuff deflated under a 6 l/minute oxygen flow) would increase after surgery (primary outcome) and evaluated the accuracy and reproducibility of CLP measurements by measuring the CLV (difference in expiratory tidal volume before and after deflation of a tracheal tube cuff).
Methods
CLV and CLP were measured before and after abdominal surgery ( n = 25; abdominal group) and cervical spine surgery ( n = 25; spine group) under general anesthesia and complete neuromuscular blockade.
Results
In both groups, the CLP was significantly higher after surgery (median [25%, 75% interquartile ranges]) (abdominal group: 4.0 [1.0, 8.4] cmH 2 O to 9.0 [3.4, 13.7] cmH 2 O, P = 0.007; spine group: 8.0 [3.0, 10.9] cmH 2 O to 11.0 [7.5, 13.5] cmH 2 O, P = 0.038). The cutoff values for 100% negative and positive predictive values for a positive CLV test (CLV < 110 ml or 25% of the tidal volume with the cuff inflated) were 12.0 and 17.3 cmH 2 O, respectively, with an AUC of 0.957 (95%CI 0.27–1.20). The CLP and CLV measurements were highly reproducible, as the Kendall’s coefficients of concordance were 0.898 (1st and 3rd) and 0.971 (6 consecutive breaths), respectively, although the CLV progressively increased by 29.0 [1.8, 58.8] ml for the 6 consecutive breaths ( P < 0.001).
Conclusion
Both the CLP and CLV measurements were accurate and highly reproducible to assess postoperative increase of the upper airway resistance before extubation.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Anesthesiology, Pain and Palliative Care Medicine, Chiba University Hospital, 1-8-1 Inohana, Chuo-Ku, Chiba, 260-8670, Japan. tyamada_violine@yahoo.co.jp.
- Department of Anesthesiology, Pain and Palliative Care Medicine, Chiba University Hospital, 1-8-1 Inohana, Chuo-Ku, Chiba, 260-8670, Japan.
- Department of Anesthesiology, Graduate School of Medicine, Chiba University, Chiba, Japan.
- Department of Preventive Medicine and Public Health, Keio School of Medicine, Tokyo, Japan.
Förderung
- Kaken Pharmaceutical
Zitieren
APA
Yamada, T., Ishibashi, K., Sakaguchi, Y., Kawakami, S., Nozaki-Taguchi, N., Sato, Y., & Isono, S. (2025). Increased cuff-leak pressure after abdominal and spine surgeries: a simple and novel cuff-leak test for tracheal extubation. Journal of Anesthesia. https://doi.org/10.1007/s00540-025-03487-w
Vancouver
Yamada T, Ishibashi K, Sakaguchi Y, Kawakami S, Nozaki-Taguchi N, Sato Y, et al. Increased cuff-leak pressure after abdominal and spine surgeries: a simple and novel cuff-leak test for tracheal extubation. Journal of Anesthesia. 2025. doi:10.1007/s00540-025-03487-w
BibTeX
@article{Yamada2025,
author = {Yamada, Takayuki and Ishibashi, Katsuhiko and Sakaguchi, Yuichi and Kawakami, Sadatoshi and Nozaki-Taguchi, Natsuko and Sato, Yasunori and Isono, Shiroh},
title = {Increased cuff-leak pressure after abdominal and spine surgeries: a simple and novel cuff-leak test for tracheal extubation},
journal = {Journal of Anesthesia},
year = {2025},
doi = {10.1007/s00540-025-03487-w}
}