Sleeve sphincter of Oddi (SO) manometry: a new method for characterizing the motility of the sphincter of Oddi
Journal of Hepato-Biliary-Pancreatic Surgery · 2008-07 · 15 (4) · 391-396
DOI: 10.1007/s00534-007-1262-1 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Journal of Hepato-Biliary-Pancreatic Surgery, online erschienen am 01.08.2008. Verfasst von Masahiko Kawamoto und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of General and Digestive Surgery, Flinders University, Flinders Medical Centre, Bedford Park, SA 5042, Australia. Die Arbeit wurde bisher 13-mal zitiert (Crossref) und nennt 29 Literaturangaben.
Abstract (Originalsprache)
BACKGROUND/PURPOSE: Perfused multilumen sphincter of Oddi (SO) manometry is accepted as the gold standard for diagnosis of SO dysfunction. However, this technique is associated with a relatively high incidence of post-procedure acute pancreatitis. In addition, triple-lumen manometry recordings may be difficult to interpret, as movement may produce artifacts. We have refined the development of a sleeve sensor for human SO manometry. This assembly aims to overcome the above limitations. In this study the accuracy of sleeve SO manometry (SOM) has been evaluated and compared with standard triple-lumen perfused SOM. METHODS: Patients undergoing SO manometric studies consented to having both standard triple-lumen and sleeve SOM. A total of 32 paired studies were performed in 29 patients. Diagnosis was made only from standard triple-lumen SOM and the patient treated accordingly. For each study, SO basal pressure, contraction, amplitude, and frequency were recorded. RESULTS: There was no statistically significant difference in the recordings of SO basal pressure, contraction, amplitude, and frequency between the two techniques. A strong correlation was demonstrated between SO basal pressure determined with the two catheters. The accuracy of sleeve SOM is comparable to standard triple-lumen SOM, with less movement artifact. One patient developed mild post-manometric pancreatitis. CONCLUSIONS: The sleeve catheter records SO pressures with comparable values to standard triple-lumen SOM. The sleeve assembly potentially can replace the use of the perfused triple-lumen catheter for the objective diagnosis of SO dysfunction.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Department of General and Digestive Surgery, Flinders University, Flinders Medical Centre, Bedford Park, SA 5042, Australia.
Zitieren
APA
Kawamoto, M., Geenen, J., Omari, T., Schloithe, A. C., Saccone, G. T., & Toouli, J. (2008). Sleeve sphincter of Oddi (SO) manometry: a new method for characterizing the motility of the sphincter of Oddi. Journal of Hepato-Biliary-Pancreatic Surgery, 15(4), 391-396. https://doi.org/10.1007/s00534-007-1262-1
Vancouver
Kawamoto M, Geenen J, Omari T, Schloithe AC, Saccone GT, Toouli J. Sleeve sphincter of Oddi (SO) manometry: a new method for characterizing the motility of the sphincter of Oddi. Journal of Hepato-Biliary-Pancreatic Surgery. 2008;15(4):391-396. doi:10.1007/s00534-007-1262-1
BibTeX
@article{Kawamoto2008,
author = {Kawamoto, Masahiko and Geenen, Joseph and Omari, Taher and Schloithe, Ann C and Saccone, Gino T P and Toouli, James},
title = {Sleeve sphincter of Oddi (SO) manometry: a new method for characterizing the motility of the sphincter of Oddi},
journal = {Journal of Hepato-Biliary-Pancreatic Surgery},
year = {2008},
volume = {15},
number = {4},
pages = {391--396},
doi = {10.1007/s00534-007-1262-1}
}
EUVIMED